TABLE 6.

Number of cervical autocontours (RPA) and manual contours (GT) that met clinical dose criteria.

Structure Metric, criteria RPA GT Number of data points
Bag_Bowel V40 Gy < 30% 33 34 36
Bladder V45 Gy < 50% 25 25 39
Femoral heads Max dose < 50 Gy 34 34 34
Femoral heads V40 Gy < 15% 34 34 34
Femoral heads V45 Gy < 60% 34 34 34
Left Femur_Head Max dose < 50 Gy 27 27 27
Left Femur_Head V40 Gy < 15% 27 27 27
Left Femur_Head V45 Gy < 60% 27 27 27
Right Femur_Head Max dose < 50 Gy 27 27 27
Right Femur_Head V40 Gy < 15% 27 27 27
Right Femur_Head V45 Gy < 60% 27 27 27
Left kidney Mean dose < 18 Gy 15 15 15
Left kidney V15 Gy < 50% 15 15 15
Left kidney V20 Gy < 33% 15 15 15
Right kidney Mean dose < 18 Gy 15 15 15
Right kidney V15 Gy < 50% 15 15 15
Right kidney V20 Gy < 33% 15 15 15
Liver V35 Gy < 50% 5 5 5
Rectum V45 Gy < 80% 34 34 39
SpinalCord Max dose < 45 Gy 17 17 17

Abbreviations: GT, Ground Truth; RPA, Radiation Planning Assistant.