FIGURE 2.

FIGURE 2

Schematics depicting the spot positioning and centering calibration measurement. The leftmost columns show the five different couch positions for performing the irradiation on the DailyQA3 device. D0 is the dose measurement by aligning the ion chamber to the imaging isocenter and the diode to the SI and LR axes. The other four dose measurements are based on ±1.5mm SI and LR couch movements. Based on the couch movement, the position of the detectors relative to the spots are different as can be seen in the illustrations. With D0, D1.5mmSI, and D1.5mmSI measurements, the offset along the SI (or Y direction) direction can be determined based on the dose asymmetry along the S(0,y) curve. Similar idea is also applied to the LR offset (or X‐directions) with D0, D1.5mmLR, and D1.5mmLR measurements and S(x,0) curve.