TABLE 2.

C. auris surveillance assay cartridge (CaurisSurV cartridge) performance against the sample sets from two different sitese

Culture positives PCRa positives Culture negatives PCR positives Combined PCR Positivesb PCR negativesc
UCLA (N = 31) WW (N = 10) UCLA (N = 10) WW UCLA + WW (N = 51) UCLA (N = 31) WW (N = 3)
CaurisSurV Cartridge Positives 30 10 6 0 46 0 0
Negatives 1 0 3 0 4 29 3
Invalidsd 0 0 1 0 1 2 0
Sensitivity (%) 96.8 100.0 66.7 NA NA
Combined sensitivity (%) 97.5 92
Specificity (%) 100.0 100.0
a

BDMax PCR assay.

b

Irrespective of culture positivity.

c

Culture not done.

d

Invalids were excluded from sensitivity/specificity calculations.

e

WW, Wadsworth Center, NY; NA, not applicable.