Table 3. Optimal kVp/mA associated with optimal SDs, EDs, and CNRs of CT images were obtained by scanning 1, 5, and 10-year-old anthropomorphic phantom (AP) heads (Please refer to Table 1).

kVp mA EDs (mSv) SDs (HU) CNRs
1-year-old 80 80 0.31 12.81 1.20
100 40 0.34 12.49 1.72
120 40 0.61 10.76 1.95
5-year-old 80 120 0.34 13.09 1.02
100 80 0.52 11.79 1.50
120 40 0.48 12.77 1.69
10-year-old 80 150 0.31 12.26 1.11
100 80 0.37 11.04 1.65
120 40 0.34 12.30 1.90

SD = standard deviation; CNR = contrast-to-noise ratio; ED = effective dose.