Table 3. Optimal kVp/mA associated with optimal SDs, EDs, and CNRs of CT images were obtained by scanning 1, 5, and 10-year-old anthropomorphic phantom (AP) heads (Please refer to Table 1).
kVp | mA | EDs (mSv) | SDs (HU) | CNRs | |
---|---|---|---|---|---|
1-year-old | 80 | 80 | 0.31 | 12.81 | 1.20 |
100 | 40 | 0.34 | 12.49 | 1.72 | |
120 | 40 | 0.61 | 10.76 | 1.95 | |
5-year-old | 80 | 120 | 0.34 | 13.09 | 1.02 |
100 | 80 | 0.52 | 11.79 | 1.50 | |
120 | 40 | 0.48 | 12.77 | 1.69 | |
10-year-old | 80 | 150 | 0.31 | 12.26 | 1.11 |
100 | 80 | 0.37 | 11.04 | 1.65 | |
120 | 40 | 0.34 | 12.30 | 1.90 |
SD = standard deviation; CNR = contrast-to-noise ratio; ED = effective dose.