Table 1.

Stent characteristics of three bioresorbable scaffolds.

Characteristics AMS-1 Absorb GT 1 MOTIV
Scaffold material Magnesiµm alloy PLLA Tyrocore
Drug coating None Everolimus+PDLLA Sirolimus+Tyrocore
Strut thickness 3.0 mm, 165 µm 2.5 mm, 150 µm 2.5 mm, 95 µm
3.5 mm, 165 µm 3.0 mm, 150 µm 3.0 mm, 105 µm
3.5 mm, 150 µm 3.5 mm, 115 µm
Crossing profile 1.5 mm 1.44 mm 1.3 mm
Delivery Single-step inflation Muti-step inflation Single-step inflation
Radial strength 0.17 N/mm 0.14 N/mm 0.22 N/mm
Recoil <8% 2.3% 2.0%
Max expansion over nominal 0.6 mm 0.5 mm 0.75 mm, 2.5–3.0 mm
0.5 mm, 3.5 mm
Resorption profile At least 4 months Loss of mechanical support in 18 months Vessel uncaged in 12 months
Resorption in 36 months Resorption in 48 months

AMS, absorbable metal stent; PLLA, poly L-lactic acid; PDLLA, poly D, L-lactic acid.