Skip to main content
. Author manuscript; available in PMC: 2025 Jul 8.
Published in final edited form as: Phys Med Biol. 2024 Jul 8;69(14):10.1088/1361-6560/ad5d4a. doi: 10.1088/1361-6560/ad5d4a

Table 3 –

Summary of multivariate regression fit (intercept and slopes) for iodine and gadolinium datasets for each CNR measure. Associated p-values listed in parentheses.

Iodine CNRlo CNRhi Covar-CNRlo Covar-CNRhi
Intercept −2.040 (0,00) 2.214 (0.21) 1.479 (0.00) −7.633 (0.05)
Tube Voltage 0.002 (0.04) 0.026 (0.04) 0.004 (0.00) 0.100 (0.00)
Tube Current 0.011 (0.00) 0.014 (0.00) 0.003 (0.00) 0.040 (0.00)
Energy Threshold 0.001 (0.21) −0.110 (0.00) −0.038 (0.00) −0.131 (0.00)
Material Concentration 0.692 (0.00) 0.501 (0.00) 0.264 (0.00) 0.759 (0.00)
Image Type −0.131 (0.00) 0.746 (0.06) −0.022 (0.37) 2.095 (0.02)
Gadolinium CNRlo CNRhi Covar-CNRlo Covar-CNRhi
Intercept −3.531 (0.00) 3.028 (0.38) 2.258 (0.00) −9.333 (0.10)
Tube Voltage 0.009 (0.00) 0.026 (0.28) 0.011 (0.00) 0.090 (0.03)
Tube Current 0.016 (0.00) 0.047 (0.00) 0.007 (0.00) 0.082 (0.00)
Energy Threshold 0.008 (0.00) −0.183 (0.00) −0.066 (0.00) −0.171 (0.00)
Material Concentration 0.886 (0.00) 0.973 (0.00) 0.399 (0.00) 1.367 (0.00)
Image Type −0.431 (0.00) 1.886 (0.02) −0.022 (0.53) 2.537 (0.05)