Table 3 –
Summary of multivariate regression fit (intercept and slopes) for iodine and gadolinium datasets for each CNR measure. Associated p-values listed in parentheses.
| Iodine | CNRlo | CNRhi | Covar-CNRlo | Covar-CNRhi |
|---|---|---|---|---|
| Intercept | −2.040 (0,00) | 2.214 (0.21) | 1.479 (0.00) | −7.633 (0.05) |
| Tube Voltage | 0.002 (0.04) | 0.026 (0.04) | 0.004 (0.00) | 0.100 (0.00) |
| Tube Current | 0.011 (0.00) | 0.014 (0.00) | 0.003 (0.00) | 0.040 (0.00) |
| Energy Threshold | 0.001 (0.21) | −0.110 (0.00) | −0.038 (0.00) | −0.131 (0.00) |
| Material Concentration | 0.692 (0.00) | 0.501 (0.00) | 0.264 (0.00) | 0.759 (0.00) |
| Image Type | −0.131 (0.00) | 0.746 (0.06) | −0.022 (0.37) | 2.095 (0.02) |
| Gadolinium | CNRlo | CNRhi | Covar-CNRlo | Covar-CNRhi |
| Intercept | −3.531 (0.00) | 3.028 (0.38) | 2.258 (0.00) | −9.333 (0.10) |
| Tube Voltage | 0.009 (0.00) | 0.026 (0.28) | 0.011 (0.00) | 0.090 (0.03) |
| Tube Current | 0.016 (0.00) | 0.047 (0.00) | 0.007 (0.00) | 0.082 (0.00) |
| Energy Threshold | 0.008 (0.00) | −0.183 (0.00) | −0.066 (0.00) | −0.171 (0.00) |
| Material Concentration | 0.886 (0.00) | 0.973 (0.00) | 0.399 (0.00) | 1.367 (0.00) |
| Image Type | −0.431 (0.00) | 1.886 (0.02) | −0.022 (0.53) | 2.537 (0.05) |