Table 2.

Scan parameters for TWIST, gated CEMRA, and bT1RESS.

TWIST Gated CEMRA bT1RESS
Acquired voxel size (mm) 2.29 × 1.38 × 2.54 1.35 × 1.15 × 2.92 2.19 × 2.19 × 2.20
Reconstructed voxel size (mm) 1.38 × 1.38 × 1.80 1.15 × 1.15 × 1.80 1.09 × 1.09 × 1.10
Slice oversampling (%) 0 8.3 50
Field of view (mm) 420 450 385
# slices (acquired/reconstructed) 56 48 64
Partial Fourier (phase, slice) 6/8, 6/8 7/8, 6/8 7/8, 7/8
Asymmetric echo Yes Yes No
Parallel acceleration GRAPPA 2 GRAPPA 2 GRAPPA 2 × 2
TR/TE/TI (ms) 2.40/0.85/none 2.75/0.94/none 2.90/1.45/320
Excitation flip angle (deg) 25 40 70
Magnetization preparation No No Composite saturation, fat suppression
Electrocardiographic gating No Yes Yes
Shot duration (ms) n/a 286 186
Trigger delay (ms) n/a 250 300
Bandwidth (Hz/pixel) 650 590 1495
Readout 3D spoiled gradient-echo 3D spoiled
gradient-echo
3D bSSFP
Image subtraction Yes No No
Other Central region A 33%, sampling density B 50%, 25 measurements Contrast agent timing obtained using prior TWIST or fluoro triggering Local shimming to heart and aorta

TWIST time-resolved angiography with interleaved stochastic trajectories, CEMRA contrast-enhanced magnetic resonance angiography, bT1RESS balanced T1 relaxation-enhanced steady-state, GRAPPA generalized autocalibrating partially parallel acquisition, 3D three-dimensional, bSSFP balanced steady-state free precession, TR repetition time, TE echo time, TI time interval.