Skip to main content
. 2024 Jul;63(4):368–376. doi: 10.30802/AALAS-JAALAS-23-000122

Table 3.

Comparison of observed bioluminescence for treatment groups in needle driver tips

Mean RLU Number of needle driver tips with bioluminescence detected
Saline 5,611 4/21
Multienzyme detergent 133 1/21
Neutral pH detergent 0 0/21
Chlorhexidine scrub 70 2/21