Skip to main content
. 2024 Jul 9;11(7):305. doi: 10.3390/vetsci11070305

Table 3.

Sensitivity results of the multiplex real-time PCR.

Pathogens Concentration Total Positive Detection Rate 90% Detection Rate
PEDV 1000 copies/μL 15 15 100% >90%
100 copies/μL 15 15 100% >90%
10 copies/μL 15 9 60% <90%
TGEV 1000 copies/μL 15 15 100% >90%
100 copies/μL 15 15 100% >90%
10 copies/μL 15 0 0% <90%
PDCoV 1000 copies/μL 15 15 100% >90%
100 copies/μL 15 15 100% >90%
10 copies/μL 15 3 20% <90%
PoRVA 1000 copies/μL 15 15 100% >90%
100 copies/μL 15 15 100% >90%
10 copies/μL 15 6 40% <90%

Negative was determined by Cq ≥ 35 and positive was validated by Cq < 35.