. 2024 Jul;69(7):782–789. doi: 10.4187/respcare.11658

Improving Breath Detection From Pulsed-Flow Oxygen Sources Using a New Nasal Interface

Cole D Christianson 1,✉, Efrem Violato 2, Mozhgan Sabz 3, Hossein Rouhani 4, Thomas Waring 5, Andrew R Martin 6
PMCID: PMC11285500  PMID: 38503467

Abstract

BACKGROUND:

Patients with COPD and other lung diseases are treated with long-term oxygen therapy (LTOT). Portable oxygen sources are required to administer LTOT while maintaining patient autonomy. Existing portable oxygen equipment has limitations that can hinder patient mobility. A novel nasal interface is presented in this study, aiming to enhance breath detection and triggering efficiency of portable pulsed-flow oxygen devices, thereby improving patient mobility and independence.

METHOD:

To examine the effectiveness of the new interface, 8 respiratory therapists participated in trials using different oxygen sources (tank with oxygen-conserving device, SimplyGo Mini portable oxygen concentrator [POC], and OxyGo NEXT POC) and breathing types (nasal and oral) while using either the new nasal interface or a standard cannula. Each trial was video recorded so participant breaths could be retroactively matched with a pulse/no-pulse response, and triggering success rates were calculated by dividing the number of oxygen pulses by the number of breaths in each trial. After each trial, volunteers were asked to rate their perceived breathing resistance.

RESULTS:

Nasal breathing consistently resulted in higher triggering success rates compared to oral breathing for pulsed-flow oxygen devices. POCs exhibited higher triggering success rates than did the oxygen tanks with conserving device. However, there were no significant differences in triggering success rates between the two POC models. The new nasal interface demonstrated improved triggering success rates compared to the standard cannula. Whereas the new nasal interface was associated with a slight increase in perceived breathing resistance during nasal breathing trials, participants reported manageable resistance levels when using the interface.

CONCLUSIONS:

This study demonstrates that the new nasal interface can improve triggering success rates of pulsed-flow oxygen devices during both nasal and oral breathing scenarios. Further research involving patient trials is recommended to understand the clinical implications of improved pulse triggering.

Keywords: oxygen therapy, portable oxygen concentrators, breath detection, oxygen-conserving devices, COPD

Introduction

Long-term oxygen therapy (LTOT) is a treatment for hypoxemic patients with COPD as well as other chronic lung diseases, including interstitial lung disease and pulmonary hypertension. In 2020, over 37 million people were living with COPD worldwide, whereas by 2050 it is projected that over 65 million people will live with COPD.1 An estimated 3.3 million North Americans receive LTOT, 60–65% of whom are diagnosed with COPD, contributing over $49 billion to national health care expenditures in the United States.2-4

It has been demonstrated that survival is improved when LTOT is administered for 15 h/d or more.5,6 Portable oxygen sources are required to administer LTOT over extended periods while maintaining patient autonomy. Unfortunately, limitations associated with currently available portable oxygen equipment can restrict patient mobility, daily activities, and independence.7-10 The most common portable oxygen sources include gas cylinders and portable oxygen concentrators (POCs). Whereas portable liquid oxygen systems offer greater capacity than gaseous systems, they are expensive to service and maintain. Over recent years, cost pressures have eliminated liquid oxygen systems in many North American regions.11 According to a survey that included responses from 836 LTOT patients, most consider using gas cylinders as strenuous or extremely strenuous due to the tank’s weight and lack of maneuverability.4 Gas cylinders are increasingly being replaced by POCs because POCs can be made much smaller, with some weighing < 5 lbs. Notably, 47% of survey respondents described using a POC to achieve increased mobility and independence despite not meeting their oxygen needs.4 This is concerning because receiving sufficient oxygen levels from LTOT is necessary to prevent end-organ damage and other serious health complications related to hypoxemia.12 On the other hand, lack of physical activity can contribute to increases in chronic lung disease exacerbations and morbidity, poor mental health, and quality of life.13-16 LTOT patients frequently must choose between healthy oxygen levels and mobility/independence due to a lack of portable oxygen sources that are both reliable and physically manageable.

Unreliable breath detection is a well-acknowledged pain point for those who use portable oxygen sources.17-21 The smallest and most efficient portable oxygen sources use pulsed delivery systems; however, their performance is variable.18,22 Pulsed-flow oxygen sources rely on pressure-triggering to detect inhalation. During inhalation, room air entrainment creates a small drop in pressure monitored through the cannula supply tubing, referred to herein as the signal pressure. When the signal pressure exceeds the triggering threshold of a pulsed-flow oxygen device, a pulse of oxygen is released. In contrast, continuous-flow sources send oxygen at all times, even during exhalation, and therefore are much less efficient than pulsed-flow sources. Signal pressures are highly sensitive to the variable fit of conventional nasal prongs within the nostrils of individual patients.17 As a result, there is significant variability in triggering efficiency. In particular, patients may not be able to trigger oxygen delivery during sleep, oral breathing, or other circumstances characterized by low nasal inhalation flows.17-21 Breath detection issues during oral breathing are especially common and may discourage oxygen users from living active lifestyles due to the tendency toward oral breathing during exercise.23

The reported LTOT patient preference is a single source of portable oxygen that can be used in all breathing scenarios.2 Some POCs have been marketed as single-source, 24-h-a-day solutions. However, nighttime hypoxemic events are reported frequently during the use of POCs compared with larger, stationary, continuous-flow oxygen sources.24 As a result, it is common for patients to switch from a portable to a stationary system before sleeping. The need to switch between different oxygen sources daily adds to the cost of providing LTOT and may be strenuous for some LTOT patients.

The present work describes a new nasal interface being developed to ensure consistent pulsed oxygen delivery to patients during a variety of breathing scenarios. This nasal interface provides control over patient signal pressures, thereby improving triggering efficiency for pulsed-flow oxygen sources. However, imposed breathing resistance increases with signal pressure. Past studies related to the new nasal interface were completed in vitro using nasal airway replicas and simulated breathing with a single portable oxygen source (SimplyGo Mini POC).20,21 In the present study, the performance of the new nasal interface is compared to that of a standard cannula when used by healthy volunteers while connected to multiple oxygen sources during nasal and oral breathing. The research questions answered include:

  1. Does the new nasal interface change triggering success rate and perceived breathing resistance?

  2. Does the type of oxygen source influence triggering success rates?

  3. Do triggering success rates change during nasal vs oral breathing?

QUICK LOOK.

Current knowledge

Many patients are dissatisfied with the reliability and portability of current portable oxygen sources. Unreliable breath detection limits pulsed-flow oxygen delivery. Mouth breathing and cannula misplacement contribute to this issue. Additionally the interface and the type of pulse dose device impacts patient satisfaction.

What this paper contributes to our knowledge

The new device was tested on healthy volunteers when using multiple different oxygen sources while breathing through their nose or mouth. Triggering was improved with nasal versus oral breathing in all subjects. POCs had a higher trigger success rate than cylinders with an oxygen-conserving device. The use of a nasal pillows interface markedly improved trigger success rate during oral breathing.

Methods

Prototypes

The new nasal interface uses nasal pillows to provide a seal around the user’s nostrils, drawing inhaled air through the air entrainment body (Fig. 1). On the bottom face of the air entrainment body, there are openings that can be blocked in different combinations to control the pressure drop across the cannula tubing (signal pressure). The prototype used in this study has two settings (Fig. 2). Setting 2 of the new nasal interface (both holes open) is designed to mimic the signal pressures generated using a standard nasal cannula. Setting 2 is, therefore, designed for use by most patients during periods of rest. Setting 1 of the new nasal interface (one hole open) is designed to amplify inhalation signals during sleep and oral breathing. Users simply slide the panel from one side to the other to change settings.

Fig. 1.

Fig. 1.

3D-printed prototype with small nasal pillows.

Fig. 2.

Fig. 2.

Computer-animated design of new nasal interface and device setting depictions.

Small, medium, or large Avalon Aire Nasal Pillows (AA-02P, Avalon Aire, West Chester, Pennsylvania) were used to match each study participant’s nostril size. Prototypes were attached to 6-ft-long standard cannula tubing (Hudson RCI RUS 1103, Teleflex, Wayne, Pennsylvania) using Loctite 4902 FL (Henkel, Düsseldorf, Germany), which meets ISO 10993 biocompatibility standards. The base of the prototypes was 3D printed using clear resin (FLBMCL01, BioMed Clear Resin, FormLabs, Somerville, Massachusetts), which is approved for biocompatibility evaluation of breathing gas pathways in health care applications (ISO-18562).

Testing Methods

Tests took place at the Northern Alberta Institute of Technology Center for Advanced Medical Simulation. Eight respiratory therapists (RTs) or RT students volunteered. The age range of participants was 22–47 y old, with 0–23 y of RT work experience. Seven volunteers were female, and one was male. Five participants indicated they are very comfortable setting up and instructing supplemental oxygen use for patients; the remaining 3 indicated they are somewhat comfortable. RT or RT students were recruited to effectively anticipate and communicate any potential issue with the new device. No other screening criteria were implemented. The Northern Alberta Institute of Technology Research Ethics Board granted ethics approval (application number 2022–05) prior to testing.

Each volunteer was asked to breathe at different flows through their nose or mouth while connected to an oxygen tank (G-10501294, VitalAire, Bagneux, France), SimplyGo Mini POC (1113601, Phillips Respironics, Murrysville, Pennsylvania), or OxyGo NEXT POC (1400-5000, OxyGo, Orlando, Florida) via standard nasal cannula or the new nasal interface. An EasyPulse5+6 oxygen-conserving device (OCD) was used with the tank (505627, Precision Medical, Northampton, Pennsylvania). A respiratory belt (Respiration Sensor, TTLP2HRVSYS, Bio-Medical Instruments, Clinton Township, Michigan) was strapped around each participant’s chest to monitor flows. The respiratory belt was calibrated by matching spirometer readings (MicroLoop Spirometer, ML3535-S, Micro Medical, Rochester, Kent, United Kingdom) to the expansion/contraction of each user’s chest. More specifically, 3 peak inspiratory flow readings from the spirometer were taken for each volunteer. Each peak flow measurement was matched with a simultaneous respiratory belt signal, in mV, to calculate a conversion factor between the raw data signal and flow with units of L×s/min×mV. Subsequent raw data signals from the respiratory belt were then multiplied by the average conversion factor calculated for each user to provide flow readings.

At the start of each trial, participants were asked to breathe nasally at their usual resting rates. Participants were asked to alter their nasal breathing to higher or lower flows later in each trial. This process was then repeated for oral breathing. Each participant had different baseline flow levels and responded differently to the high/low flow request. As a result, a large variation of flows (∼5–120 L/min) was tested.

Every trial was video recorded with a close-up view of each oxygen source to track when the source sent a pulse of oxygen. Each breath was retroactively matched with a pulse/no-pulse response, and the number of pulses was divided by the number of breaths to calculate triggering success rates. Approximately 30 s were allowed per trial. The number of breaths included for each triggering success rate calculation varied slightly between trials, as individual breathing rates differed. On average, 9 breaths were included in each calculation, with an SD of 2 breaths. After every trial, participants were asked to rate the breathing resistance on a scale of 1–10, one being no resistance and 10 being extreme resistance. Table 1 depicts the breathing resistance scoring metric provided to participants.

Table 1.

Resistance Score Description Chart

graphic file with name DE-RESC240041T001.jpg

Due to time constraints, complete triggering success rate data could not be collected for the new interface setting 2. As a result, triggering success rates for only the standard cannula and interface setting 1 are presented. New nasal interface setting 2 was treated as a lower priority because this setting is designed to perform similarly to a standard cannula. However, breathing resistance scores are still presented for interface setting 2. Lastly, one of the new nasal interface prototypes had an air leak due to improper assembly. As a result, data from only 7 of 8 participants were included in the statistical analysis.

Statistical Methods

Twelve repeated measurements and 84 total triggering success rate samples were included to account for every oxygen source (tank/OCD combination, OxyGo NEXT, or SimplyGo Mini) and breathing type (nasal or oral) combination, with participants using either a standard cannula or nasal interface setting 1. Additionally, participants were asked to rate breathing resistances for each cannula (standard, interface setting 1, or interface setting 2) and breathing type. As a result, 6 repeated measurements and 42 total breathing resistance scores were collected. This sample size was justified a priori using G*Power software (version 3.1.9.7, Heinrich Heine University, Düsseldorf, Germany)25,26 to estimate statistical power for a repeated measures analysis of variance (within factors). Cohen F was set to 0.25 to reflect a medium effect size, and the correlation among repeated measurements was set to 0.5.27 Type-1 and type-2 error probabilities were both set to 0.05, resulting in a desired sample size of 28 (5 participants) assuming sphericity. Extra participants were recruited allowing for dropouts, technical issues, non-sphericity, and/or other violated assumptions.

All statistical tests were completed using SPSS Statistical software (version 28.1.1, IBM, Armonk, New York). A P value < .05 was used to indicate significance for all tests. Data were first tested for normality and sphericity using a Shapiro-Wilk or Mauchly test as appropriate. The collected data were not normally distributed, so Friedman tests were used to test for significant differences.28 Furthermore, individual Wilcoxon matched-pairs signed-rank tests were performed to isolate differences.29 Triggering success rate and breathing resistance scores are continuous and ordinal data types, respectively, both of which are compatible with Friedman and Wilcoxon matched-pairs tests.28,29

Results

Shapiro-Wilk tests revealed that triggering success rate and breathing resistance score data were not normally distributed (P < .001). Triggering success rate data were spherical (P = .08), however, breathing resistance scores were not (P = .01). Consequently, Friedman and Wilcoxon tests were conducted, and results for these tests are presented in the following sections.

Triggering Success Rates

The number of well-timed oxygen pulses was divided by the total number of breaths to calculate the triggering success rate for each trial. Examples of triggering success rate results for an individual participant when using an oxygen tank with OCD are shown in Figure 3. In Figure 3A, the gray curve represents inhalation and exhalation flows when the participant was nasal breathing. The participant started breathing at their standard flow levels (∼30–35 L/min), and when they were asked to lower their flow, measurements dropped to about 20 L/min. The participant was then asked to start breathing orally at their standard flow levels. When asked to increase their oral breathing flow, the participant reached almost 120 L/min. Each nasal and oral breathing trial for a given cannula type and oxygen source was split by a participant break, where participants were given a chance to catch their breath and/or ask questions.

Fig. 3.

Fig. 3.

Example triggering success rate results while using an oxygen tank with (A) a standard nasal cannula and (B) new nasal interface setting 1. The gray and black curves represent inhalation and exhalation flows when the participant was nasal and oral breathing, respectively. Breaths during the subject break were not included in the data analysis.

Friedman test results showed significant differences between oxygen source, breathing type, and nasal interface groups (P < .001). One-tailed Wilcoxon test results indicated that triggering success rates were higher when nasal breathing versus oral breathing (P < .001), when using new nasal interface setting 1 versus a standard cannula (P < .001), and when using an OxyGo NEXT or SimplyGo Mini versus a tank/OCD (P < .001). However, significant differences were not found between OxyGo NEXT and SimplyGo Mini triggering success rates (P = .89) (Fig. 4).

Fig. 4.

Fig. 4.

Triggering success rate 95% CI based on Wilcoxon test groupings. Triggering success rate samples were split based on breathing type (top row), based on cannula type (second row), based on oxygen source (third and fourth rows), and based on cannula type, collected during oral breathing trials only (fifth row). P values are provided for each comparison. OCD = oxygen-conserving device; POC = portable oxygen concentrator.

The average triggering success rate across all nasal breathing trials was 92% based on 387 total breaths and 50% for oral breathing trials based on 340 breaths. The average triggering success rate was 63% across all standard cannula trials based on 392 breaths and was 79% for nasal interface setting 1 based on 335 total breaths. The average triggering success rate across all tank/OCD trials was 56% based on 262 total breaths and was 78% for POC trials based on 465 (226 SimplyGo Mini and 239 OxyGo NEXT) breaths (Fig. 4).

Wilcoxon tests also showed a significant difference (P < .001) between performance of the standard cannula and new interface setting 1 after isolating oral breathing trials. The averaging triggering success rate for standard cannula oral breathing trials was 36% based on 182 breaths and was 64% for interface setting 1 oral breathing trials based on 158 breaths.

Breathing Resistance Scores

Friedman test results showed significant differences between breathing resistance scores among the breathing type/nasal interface groups (P = .003). A one-tailed Wilcoxon test revealed that breathing resistance scores were higher during nasal breathing versus oral breathing (P = .006). However, no significant differences were found between each cannula group based on 2-tailed Wilcoxon test results. The P values for standard cannula versus interface setting 2, standard cannula versus interface setting 1, and interface setting 2 versus 1 were .56, .059, and .059, respectively. Breathing resistance scores for each breathing type and cannula combination are provided in Figure 5.

Fig. 5.

Fig. 5.

Breathing resistance scores. Error bars represent the interquartile range. The interquartile range limits overlap for all oral breathing trials.

Discussion

The difference in triggering success rates between nasal and oral breathing trials (92% vs 50%) indicates that breath detection during oral breathing is a serious issue when using pulsed-flow oxygen devices. POCs were triggered more often than tanks (78% vs 56%). According to Dakkak et al,4 POCs are already the patient-preferred device for reasons such as portability and lighter weight. Improved triggering efficiency may be another advantage for some POCs over some oxygen-conserving regulators.

With a 16% increase across all trials, the new nasal interface significantly improved triggering success rates. The new nasal interface increased triggering success rates from 36% to 64% during oral breathing trials. Whereas this is a substantial improvement, further adjustments could be made to the interface design to ideally approach a 100% triggering rate.

Improved triggering success rate comes at the expense of increased imposed breathing resistance.20,21 Interface setting 1 received a mean perceived resistance score of 3 during nasal breathing trials, indicating minor resistance. In contrast, interface setting 2 and the standard cannula received mean scores between 1–2, correlating to no resistance. No significant differences were found between perceived resistance scores for the studied cannula types, and, as designed, interface setting 2 scored similarly to the standard cannula. Whereas these are encouraging results, interface setting 1 received a score of 9 from one participant, indicating major to extreme resistance. Further improvement to the prototype design could minimize resistance in order to enhance patient comfort.

One limitation of this study is that the precise timing of each oxygen pulse was not recorded. Each breath was given a pulse/no-pulse response, but the exact time at which the pulses were released within the breath could not be tracked. The timing at which the pulse is released can influence FIO2.20,21 Another limitation is that healthy participants were recruited instead of patients. The perceived resistance of healthy people may vary significantly from patients with a pulmonary disorder, and triggering success rates may differ as well. Furthermore, all recruited participants had at least some degree of familiarity with oxygen equipment, which may have added a bias to the study in that participants who knew how the devices were intended to function prior to debriefing may have altered their breathing patterns. Lastly, devices were used for a relatively short amount of time, which precluded assessment of how participants would tolerate the device over longer periods of use. Long-term comfort, for example, when used overnight, remains to be assessed. Patient trials should be conducted in the future to address the limitations of this study.

Conclusions

A new nasal interface was designed to improve triggering success rates of pulsed-flow oxygen sources. Eight RTs were recruited to test the device and asked to breathe through their nose or mouth while connected to an oxygen tank with OCD, SimplyGo Mini POC, or OxyGo NEXT POC via a standard nasal cannula or the new nasal interface. The results indicate that oral breathing posed a serious issue for pulsed-flow oxygen sources in terms of triggering success rate. Both POCs triggered more frequently than the oxygen tank/OCD combination. The new nasal interface improved triggering efficiency compared to a standard nasal cannula. Improved triggering efficiency comes at the expense of a slight increase in perceived breathing resistance.

Footnotes

Mr Christianson and Dr Martin are co-founders and owners of Pulmvita, which has licensed intellectual property from the University of Alberta for the purpose of developing the new nasal interface. The remaining authors have disclosed no conflicts of interest.

Mr Christianson presented a version of this paper at AARC Congress 2023 held November 5–8, 2023, in Nashville, Tennessee.

This study was supported by a Mitacs Accelerate Entrepreneur grant (IT29414).

See the Related Editorial on Page 905

REFERENCES

  • 1.Tellez D, Gondalia R, Barrett M, Benjafield A, Nunez CM, Malhotra A. An estimate of the Americas’ prevalence of chronic obstructive pulmonary disease in 2050. Abstract presented at: American Thoracic Society 2021 International Conference, held virtually May 14–19, 2021. [Google Scholar]
  • 2.Inogen. Annual Report 2021. Available at: https://investor.inogen.com/financial-information/annual-reports. Accessed October 15, 2023. [Google Scholar]
  • 3.Ford ES, Murphy LB, Khavjou O, Giles WH, Holt JB, Croft JB. Total and state-specific medical and absenteeism costs of COPD among adults age 18 years in the United States for 2010 and projections through 2020. Chest 2015;147(1):31-45. [DOI] [PubMed] [Google Scholar]
  • 4.Dakkak J, Tang W, Smith JT, Balasubramanian A, Mattson M, Ainechi A, et al. Burden and unmet needs with portable oxygen in patients on long-term oxygen therapy. Ann Am Thorac Soc 2021;18(9):1498-1505. [DOI] [PubMed] [Google Scholar]
  • 5.Long-term domiciliary oxygen therapy in chronic hypoxic cor pulmonale complicating chronic bronchitis and emphysema. Report of the Medical Research Council Working Party. Lancet 1981;1(8222):681-686. [PubMed] [Google Scholar]
  • 6.Nocturnal Oxygen Therapy Trial Group. Continuous or nocturnal oxygen therapy in hypoxemic chronic obstructive lung disease: a clinical trial. Ann Intern Med 1980;93(3):391-398. [DOI] [PubMed] [Google Scholar]
  • 7.AlMutairi HJ, Mussa CC, Lambert CT, Vines DL, Strickland SL. Perspectives from COPD subjects on portable long-term oxygen therapy devices. Respir Care 2018;63(11):1321-1330. [DOI] [PubMed] [Google Scholar]
  • 8.Arnold E, Bruton A, Donovan-Hall M, Fenwick A, Dibb B, Walker E. Ambulatory oxygen: why do COPD patients not use their portable systems as prescribed? A qualitative study. BMC Pulm Med 2011;11:9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Graney BA, Wamboldt FS, Baird S, Churney T, Fier K, Korn M, et al. Informal caregivers experience of supplemental oxygen in pulmonary fibrosis. Health Qual Life Outcomes 2017;15(1):133. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Jacobs SS, Lindell KO, Collins EG, Garvey CM, Hernandez C, McLaughlin S, et al. Patient perceptions of the adequacy of supplemental oxygen therapy. Results of the American Thoracic Society Nursing Assembly Oxygen Working Group survey. Ann Am Thorac Soc 2018;15(1):24-32. [DOI] [PubMed] [Google Scholar]
  • 11.Hess MW. Oxygen therapy in COPD. Respir Care 2023;68(7):998-1012. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Koczulla AR, Schneeberger T, Jarosch I, Kenn K, Gloeckl R. Long-term oxygen therapy. Dtsch Arztebl Int 2018;115(51–52):871-877. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Lampinen P, Heikkinen E. Reduced mobility and physical activity as predictors of depressive symptoms among community-dwelling older adults: an 8-year follow-up study. Aging Clin Exp Res 2003;15(3):205-211. [DOI] [PubMed] [Google Scholar]
  • 14.Pitta F, Troosters T, Probst VS, Spruit MA, Decramer M, Gosselink R. Quantifying physical activity in daily life with questionnaires and motion sensors in COPD. Eur Respir J 2006;27(5):1040-1055. [DOI] [PubMed] [Google Scholar]
  • 15.Moy ML, Teylan M, Weston NA, Gagnon DR, Danilack VA, Garshick E. Daily step count is associated with plasma C-reactive protein and IL-6 in a US cohort with COPD. Chest 2014;145(3):542-550. [DOI] [PubMed] [Google Scholar]
  • 16.Eaton T, Garrett JE, Young P, Fergusson W, Kolbe J, Rudkin S, et al. Ambulatory oxygen improves quality of life of COPD patients: a randomized controlled study. Eur Respir J 2002;20(2):306-312. [DOI] [PubMed] [Google Scholar]
  • 17.Chen JZ, Katz IM, Pichelin M, Zhu K, Caillibotte G, Noga ML, et al. Comparison of pulsed versus continuous oxygen delivery using realistic adult nasal airway replicas. Int J Chron Obstruct Pulmon Dis 2017;12:2559-2571. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Chen JZ, Katz IM, Pichelin M, Zhu K, Caillibotte G, Finlay WH, et al. In vitro–in silico comparison of pulsed oxygen delivery from portable oxygen concentrators versus continuous flow oxygen delivery. Respir Care 2019;64(2):117-129. [DOI] [PubMed] [Google Scholar]
  • 19.Martin DC. Contemporary portable oxygen concentrators and diverse breathing behaviors – a bench comparison. BMC Pulm Med 2019;19(1):217. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Christianson C. Improving portable oxygen concentrator performance through the development of a new nasal interface. doi: 10.7939/R3-84KY-H719. [DOI]
  • 21.Christianson CD, Pillay K, Chen JZ, Finlay WH, Martin AR. In vitro evaluation of a nasal interface used to improve delivery from a portable oxygen concentrator. J Med Devices 2022;16(1):011101. [Google Scholar]
  • 22.Branson RD, King A, Giordano SP. Home oxygen therapy devices: providing the prescription. Respir Care 2019;64(2):230-232. [DOI] [PubMed] [Google Scholar]
  • 23.Bitos K, Furian M, Mayer L, Schneider SR, Buenzli S, Mademilov MZ, et al. Effect of high-flow oxygen on exercise performance in COPD patients. Randomized trial. Front Med (Lausanne) 2020;7:595450. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Yáñez AM, Prat JP, Álvarez-Sala JL, Calle M, Díaz Lobato S, García Gonzalez JL, et al. Oxygenation with a single portable pulse-dose oxygen-conserving device and combined stationary and portable oxygen delivery devices in subjects with COPD. Respir Care 2015;60(3):382-387. [DOI] [PubMed] [Google Scholar]
  • 25.Faul F, Erdfelder E, Lang A-G, Buchner A. G*Power 3: a flexible statistical power analysis program for the social, behavioral, and biomedical sciences. Behav Res Methods 2007;39(2):175-191. [DOI] [PubMed] [Google Scholar]
  • 26.Faul F, Erdfelder E, Buchner A, Lang A-G. Statistical power analyses using G*Power 3.1: tests for correlation and regression analyses. Behav Res Methods 2009;41(4):1149-1160. [DOI] [PubMed] [Google Scholar]
  • 27.Sigma Plus Statistiek. SPSS tutorials: effect size in statistics - The ultimate guide. Available at: https://www.spss-tutorials.com/effect-size/#anova. Accessed October 15, 2023.
  • 28.Laerd Statistics. Friedman test in SPSS Statistics - How to run the procedure, understand the output using a relevant example. Available at: https://statistics.laerd.com/spss-tutorials/friedman-test-using-spss-statistics.php. Accessed October 15, 2023. [Google Scholar]
  • 29.Laerd Statistics. Wilcoxon signed-rank test in SPSS Statistics - procedure, output, and interpretation of output using a relevant example. Available at: https://statistics.laerd.com/spss-tutorials/wilcoxon-signed-rank-test-using-spss-statistics.php. Accessed October 15, 2023. [Google Scholar]

Articles from Respiratory Care are provided here courtesy of SAGE Publications