Table 1: TEER Device Selection Based on Valve Morphology[3639].

Device Mitral Leaflet Considerations
MitraClip G4 NT Recommended minimum 6 mm leaflet insertion grasped in each leaflet; more suitable for restricted leaflets Favours a smaller valve area (4 mm grasping width)
MitraClip G4 NTW Recommended minimum 6 mm leaflet insertion grasped in each leaflet; more suitable for restricted leaflets Favours broad jet (6 mm grasping width)
MitraClip G4 XT Recommended minimum 9 mm leaflet insertion grasped in each leaflet
MitraClip G4 XTW Recommended minimum 9 mm leaflet insertion grasped in each leaflet
Favours broad jet (6 mm grasping width)
PASCAL 10 mm long clasp, 10 mm central spacer proficient in variable valve leaflet sizes and jets

TEER = transcatheter edge-to-edge repair.