Device Deployment Shape Anchoring Mechanism Implantation Approach Valve Position Key Measurements CE Mark
Tendyne Self-expandable D-shaped outer stent Apical tether Transapical Intra-annular SL and IC distances 2020
Intrepid Self-expandable Circular Radial force, leaflet engagement by outer stent cleats Transapical Intra-annular Perimeter, SL and IC distances No
TIARA Self-expandable D-shaped 3 ventricular anchoring tabs Transapical Intra-annular Perimeter, IC distances No
Evoque Self-expandable Circular Circumferential anchors for both ventricular and atrium Transapical
Transseptal
Supra-annular Maximum diameter No
Sapien M3 Self-expandable Circular Nitinol docking Transseptal Intra-annular Perimeter, IC distances No
HighLife Self-expandable Circular Valve-in-ring Transapical
(Transseptal implantation of ring)
Intra-annular Maximum diameter No
AltaValve Self-expandable Circular Atrial anchoring, suspended valve Transapical
Transseptal
Supra-annular No