Spatial Resolution of 22Na Point Source Measured with Filtered Backprojection

Axial position Radial off-center position (cm) Full width at half maximum (mm)
Radial Tangential Axial
½ AFOV 1 1.63 1.68 2.55
10 2.10 2.06 2.97
20 2.17 2.87 2.82
⅛ AFOV 1 1.76 1.47 2.91
10 1.99 2.08 2.80
20 2.20 2.78 3.05