Table 1.
The Regression parameters and validation results for determining IND and MOM by the proposed methods
| Parameter | IPC method | Spectroscopic methods | ||||
|---|---|---|---|---|---|---|
| First derivative (1D) | Ratio derivative (RD) | |||||
| MOM | IND | MOM | IND | MOM | IND | |
| Wavelength (nm) | 259 | 259 | 275.67 | 251.62 | 280.64 | 310.49 |
| Linearity range (µg mL−1) | 0.1–10 | 0.1–10 | 1.1–32 | 1–30 | 1.1–32 | 1–30 |
| Slope | 62.347 | 55.128 | − 0.007 | 0.01 | 0.065 | 0.083 |
| Intercept | 4.120 | 36.436 | − 0.016 | 0.013 | 0.024 | − 0.017 |
| correlation coefficients (r) | 0.9997 | 0.9998 | 0.9996 | 0.9998 | 0.9998 | 0.9999 |
| Accuracy (%R) | 99.97 | 99.91 | 100.02 | 99.69 | 100.28 | 100.11 |
| Precision (%RSD) | ||||||
| Repeatabilitya | 0.531 | 0.832 | 0.366 | 0.774 | 0.682 | 0.755 |
| Intermediate precisionb | 0.732 | 0.965 | 0.348 | 0.675 | 0.985 | 0.857 |
| LOD (μg/mL) | 0.026 | 0.031 | 0.290 | 0.314 | 0.202 | 0.227 |
| LOQ (μg/mL) | 0.079 | 0.094 | 0.879 | 0.951 | 0.612 | 0.688 |
| Sy/x | 0.965 | 1.012 | 9.1 × 10–4 | 1.2 × 10–5 | 5.6 × 10–4 | 8.12 × 10–4 |
| Sa | 0.5 | 0.52 | 2.0 × 10–5 | 5.00 × 10–4 | 1.0 × 10–5 | 0.01 |
| Sb | 0.099 | 0.103 | 4.0 × 10–5 | 5.00 × 10–5 | 2.2 × 10–5 | 3.4 × 10–5 |
| % RSDc | 1.334 | 0.677 | 0.737 | 0.629 | 0.556 | 0.97 |
| % Errord | 0.471 | 0.239 | 0.301 | 0.257 | 0.227 | 0.395 |
| Robustnesse (%RSD) | 0.317 | 0.586 | 0.432 | 0.602 | 0.552 | 0.499 |
aThe intraday (n = 3), average of three responses of three different concentrations repeated three times on the same day
bThe interday (n = 3), average of three responses of different three concentrations repeated three times at three successive days
cPercentage relative standard deviation(SD × 100/mean %R)
dPercentage relative error. = % RSD/
eAverage of three determinations