TABLE 2.

Median and interquartile range (IQR) of IDENTIFY™ reported offsets before beam‐on at non‐zero couch angles.

Couch angle (°) Vertical (IQR) mm Longitudinal (IQR) (mm) Lateral (IQR) (mm)
270 −0.01 (0.47) −0.03 (0.33) 0.16 (0.32)
225 −0.06 (0.17) 0.21 (0.38) −0.16 (0.18)
135 −0.08 (0.16) 0.19 (0.38) −0.41 (0.19)
90 −0.14 (0.20) 0.05 (0.54) −0.25 (0.24)
0° (end of treatment) −0.05 (0.18) −0.03 (0.35) −0.01 (0.15)