TABLE 9.

Comparison of Acuros BV‐calculated and RadCalc‐calculated point doses for completed interstitial cervical plans. All patients below had two implants and two separate plans, each with a distinct calculation point.

Patient Acuros BV dose point implant 1 (cGy) RadCalc dose point implant 1 (cGy) Percent difference (%) Acuros BV dose point implant 2 (cGy) RadCalc dose point implant 2 (cGy) Percent difference (%)
6 537.09 535.43 −0.3% 546.38 545.04 −0.2%
7 567.02 581.52 2.6% 549.86 560.23 1.9%
8 512.20 526.22 2.7% 534.89 549.18 2.7%
9 594.79 605.36 1.8% 321.83 326.49 1.4%
10 515.25 531.50 3.2% 545.50 562.17 3.1%