Managing hardware-related metal artifacts in MRI: current and evolving techniques

Georg C Feuerriegel 1,, Reto Sutter 1
PMCID: PMC11303499  PMID: 38381196

Abstract

Magnetic resonance imaging (MRI) around metal implants has been challenging due to magnetic susceptibility differences between metal implants and adjacent tissues, resulting in image signal loss, geometric distortion, and loss of fat suppression. These artifacts can compromise the diagnostic accuracy and the evaluation of surrounding anatomical structures. As the prevalence of total joint replacements continues to increase in our aging society, there is a need for proper radiological assessment of tissues around metal implants to aid clinical decision-making in the management of post-operative complaints and complications. Various techniques for reducing metal artifacts in musculoskeletal imaging have been explored in recent years. One approach focuses on improving hardware components. High-density multi-channel radiofrequency (RF) coils, parallel imaging techniques, and gradient warping correction enable signal enhancement, image acquisition acceleration, and geometric distortion minimization. In addition, the use of susceptibility-matched implants and low-field MRI helps to reduce magnetic susceptibility differences. The second approach focuses on metal artifact reduction sequences such as view-angle tilting (VAT) and slice-encoding for metal artifact correction (SEMAC). Iterative reconstruction algorithms, deep learning approaches, and post-processing techniques are used to estimate and correct artifact-related errors in reconstructed images. This article reviews recent developments in clinically applicable metal artifact reduction techniques as well as advances in MR hardware. The review provides a better understanding of the basic principles and techniques, as well as an awareness of their limitations, allowing for a more reasoned application of these methods in clinical settings.

Keywords: Magnetic resonance imaging, Arthroplasty, Artifacts, Joint prosthesis

Introduction

The prevalence of total joint replacements is steadily increasing each year, mainly due to the aging population and the increased demands on the physical abilities of the elderly. A total of 743,327 knee and hip replacements were performed in the USA in the year 2019, and an increase of approximately 139–176% is estimated for the year 2040 creating a need for appropriate diagnostic imaging tools around metal implants [1]. There are several imaging modalities that can be used to assess and diagnose postoperative complications after joint replacement, including ultrasound, computed tomography, and radiography. However, because of its good soft tissue contrast, magnetic resonance imaging (MRI) has proved invaluable in diagnosing postoperative complications following joint replacement [2, 3]. In postoperative patients with complaints after joint replacements, common suspected abnormalities include infection and abscess formation, pseudotumour formation, nonunion, aseptic loosening, fracture, and soft tissue abnormalities. MRI around metal implants remains challenging, but because of technological advances in recent years, MRI of metal implants is a major focus in musculoskeletal imaging. Improvements in equipment hardware such as multi-channel coils, as well as the optimization of dedicated MR protocols to reduce metal artifacts, have helped to address a variety of clinical questions. In this review, the physical background of metal artifact formation and the influencing factors are discussed. Furthermore, different strategies for metal artifact reduction, including hardware- and software-based approaches, are presented and the advantages and disadvantages of these techniques are discussed. As a special focus, this review includes the recent advantages of low-field MRI and deep learning–based image reconstruction for metal artifact reduction.

Artifact formation on MRI—independent/uncontrolled factors

MRI artifacts are mainly caused by variations in the resonance frequency, which are due to differences in the magnetic susceptibility between metal implants and the surrounding tissue. The magnetic susceptibility of a substance describes how it is magnetized in a given magnetic field [4]. For example, ferromagnetic materials such as iron, cobalt, or nickel cause a greater change in the local magnetic field than paramagnetic materials such as titanium and gadolinium. The constant, homogeneous magnetic field of any MR scanner is called the B0 field and is used to polarize the spins, creating magnetization. In contrast, the B1 field is a high-frequency energy field applied perpendicularly to the B0 field to perturb the net magnetization, such as excitation or inversion pulses. Achieving a linear, homogeneous main magnetic field B0 is crucial for accurate signal encoding and image reconstruction in MRI. The sharp magnetic susceptibility transitions of metal implants to the surrounding paramagnetic soft tissue cause local inhomogeneities in the B0 field. The inhomogeneous B0 field during MRI acquisition results in three main effects: signal loss and pile-up, geometric distortion, and fat suppression failure [5]. MRI artifacts in the presence of metal implants are most severe near the implant and distant structures can be imaged with fewer limitations. It is therefore important to assess whether the region of interest will be affected by artifacts prior to imaging. In patients with, e.g., unilateral hip replacement, the contralateral hip can be assessed without dedicated sequences.

Signal loss and pile-up

Changes in the local magnetic field due to metallic implants result in rapid dephasing and incoherence of the spins within a single voxel [6, 7]. The resonance of the protons is shifted out of the bandwidth of the radiofrequency pulse, resulting in signal loss, manifested as black spots due to the absence of signal at the expected location, or signal pile-up due to increased signal at another location. Ferromagnetic implants exhibit a typical four-leaf clover pattern where the implant acts as a dipole, causing inhomogeneities with both suppression and enhancement of the local B0 magnetic field (Fig. 1).

Fig. 1.

Fig. 1

A 61-year-old patient after rotator cuff repair. The anteroposterior radiograph (A) shows four suture anchors in the humeral head made of titanium. The standard sagittal fat-saturated proton density (PD)-weighted image (B) shows the typical four-leaf clover artifact caused by the ferromagnetic metal implants, which interferes with the diagnostic evaluation of the humeral head and the insertion of the rotator cuff tendons. The axial T1-weighted Dixon MR sequence (C) demonstrates typical artifacts caused by the ferromagnetic metal implants in Dixon sequences. A dedicated short-tau inversion recovery (STIR) turbo spin echo sequence with optimized inversion pulse (STIR WARP) (D) shows significantly less artifacts with adequate fat suppression, allowing assessment of the integrity of the rotator cuff insertion. The images were acquired at 1.5 T with following sequence parameters: PD FS—echo time (TE) 58 ms, repetition time (TR) 3370 ms, receiver bandwidth 435 Hz/pixels; T1 TSE FS Dixon—TE 36 ms, TR 3620 ms, receiver bandwidth 250 Hz/pixel; STIR WARP—TE 47 ms, TR 4000 ms, receiver bandwidth 300 Hz/pixel, inversion time (IT)170 ms

Geometric distortions

The generation of an MR image relies on the precise localization of each tissue voxel through the application of position-dependent gradient fields during both slice selection and readout. Metal-induced variations in the B0 field lead to errors in the position of the selected slice, ultimately resulting in a change in the precession frequency of the affected spins [8]. As a result, spins outside the selected slice are unintentionally excited, and the selected slice is distorted by data from adjacent slices, known as through-plane or “potato chip” artifacts. Furthermore, during the readout phase, pixels become improperly aligned with the wrong positions along the frequency-encoding (readout) direction [4]. Ideally, the encoding gradient should be linear, but local field gradients can distort it so that it becomes curvilinear, with the result that the position of a voxel’s data along that axis is misregistered (Fig. 2). However, since local field gradients do not affect the phase of precession, the phase encoding direction is insensitive to in-plane artifacts [8]

Fig. 2.

Fig. 2

Signal pile-up and geometric distortion caused by a residual metal fragment after removal of orthopedic hardware from the distal humerus in a 38-year-old patient (patient 1, AD) and a 41-year-old patient (patient 2, EH). The MRI of the first patient was acquired with standard proton density turbo spin echo (PD TSE) sequences with fat saturation in axial (B) and coronal (C) orientation, as well as a standard T1-weighted TSE sequence in axial orientation (D). Note the significantly increased signal pile-up and geometric distortion in the standard sequences (asterisks). Additionally, a failure of fat suppression can be seen on the coronal PD TSE sequence around the metal fragment (C). Patient 2 was examined with dedicated metal artifact reduction sequences including an axial and sagittal short-tau inversion recovery (STIR) TSE sequence with optimized inversion pulse (STIR WARP) (F, G) and a high bandwidth (BW) T1-weighted TSE sequence. Note that only localized artifacts are seen around the metal fragment (white circle). The severity of artifacts is due to the ferromagnetic properties of the broken-off metal fragments. The images were acquired at 1.5 T with following sequence parameters: transverse PD FS—echo time (TE) 26 ms, repetition time (TR) 2902 ms, receiver bandwidth 168 Hz/pixel; coronal PD FS—TE 24 ms, TR 2341 ms, receiver bandwidth 104 Hz/pixel; T1 FSE—TE 11 ms, TR 603 ms, receiver bandwidth 107 Hz/pixel; transverse STIR WARP—TE 41 ms, TR 3000 ms, receiver bandwidth 360 Hz/pixel, inversion time (TI) 130 ms; coronal STIR WARP—TE 22 ms, TR 3000 ms, receiver bandwidth 400 Hz/pixel, IT 130 ms; T1 SE High BW—TE 11 ms, TR 400 ms, receiver bandwidth 270 Hz/pixel

Failed fat suppression

Chemical-shift-selective fat suppression relies on the distinct resonance frequencies of fat and water protons. For effective fat saturation, a short-duration RF pulse tuned to the RF of fat is applied before the MR image acquisition. This causes the fat signal to be selectively saturated (nulled) while the water signal remains unaffected [8, 9]. B0 inhomogeneities induced by metal implants shift the fat resonance peak out of the frequency range targeted by the saturation pulse, resulting in failure of fat suppression or even cause saturation of the water signal [6] (Fig. 3).

Fig. 3.

Fig. 3

A 54-year-old patient after arthrodesis of the upper ankle joint following curettage and plombage of an aneurysmatic bone cyst with titanium screws and plates. The patient suffered from recurrent ankle pain on movement. The lateral radiograph (A) shows several types of metal implants in different orientations. The high bandwidth (BW) sagittal T1-weighted fat-suppressed turbo spin echo (TSE) sequence (B, bandwidth 372 Hz/pixel) shows severe signal accumulation and geometric distortion with failure of fat saturation (asterisk). Note also that the failure of fat suppression does not allow a reliable detection of the bone marrow edema in the talus (arrow), which was most likely the source of the ankle pain, as the signal in the talus looks similar to the signal in the distal tibia. The sagittal T1-weighted Dixon sequence (C) provides adequate overall fat suppression, but significant focal artifacts remain around the implants. A sagittal short tau inversion recovery (STIR) turbo spin echo sequence (D) with optimized inversion pulse (STIR WARP) demonstrated the least artifacts and provided adequate fat suppression, visualizing the bone marrow edema of the talus head (arrow). The images were acquired at 1.5 T with following sequence parameters: T1 TSE FS High BW—echo time (TE) 10 ms, repetition time (TR) 631 ms, receiver bandwidth 372 Hz/pixel; T1 TSE FS Dixon—TE 12 ms, TR 655 ms, receiver bandwidth 248 Hz/pixel; STIR WARP—TE 48 ms, TR 4000 ms, receiver bandwidth 385 Hz/pixel, inversion time (IT) 160 ms

Metal artifact reduction in MRI—basic parameter changes

Over the past decade, several approaches have been proposed to reduce metal artifact in MRI, of which three main approaches have shown the most promise. The first approach is to improve the interaction of the metal implant with the MR hardware. The second is to adjust and adapt conventional MR sequence parameters and the third is to apply advanced metal artifact reduction techniques such as multispectral imaging, advanced reconstruction techniques, post-processing, and deep learning–based techniques.

MR hardware

Field strength

The severity of metal artifacts depends on magnetic field strength and intensifies with increasing field strength due to the linearly increasing magnetic field differences caused by the susceptibility difference between the metal implant and the surrounding soft tissue [5, 6, 10]. Consequently, imaging at a lower field strength of 0.55 T produces fewer susceptibility artifacts, increasing linearly to the most artifacts at 7.0 T. However, the use of low-field MRI has the disadvantage of a lower signal-to-noise ratio (SNR) and lower image resolution, while increased specific absorption rates (SAR) are a significant limitation at higher field strengths [5, 11, 12]. Metal artifact reduction techniques can be applied regardless of magnetic field strength, but if available, a 1.5-T scanner is preferred for imaging patients with metal implants, as it offers the best trade-off between reduced magnetic field differences and higher SNR [13].

Influence of the implant

The main reasons for artifacts in MR imaging with metal implants are due to differences in the magnetic susceptibility between metal implants and the surrounding tissue. Ideally, implants would exhibit the same magnetic susceptibility as the surrounding tissue in order to reduce both in-plane and through-plane artifacts. Signal void (dark) and pile-up (bright) artifacts from metal occur in multiple directions. In-plane refers to artifacts occurring within the imaging slice and through-plane refers to artifacts displaced from one slice to others. Titanium is now one of the most commonly used materials for metal implants which causes significantly less artifacts compared to stainless steel or cobalt-chromium [1418]. Current research is focusing on further reducing the magnetic susceptibility of metal implants by synthesizing aluminum-free titanium composite materials or carbon-fiber-reinforced polymers and biodegradable magnesium alloys [1922].

Positioning

In general, metal implants produce the least artifacts when the long axis is aligned with the magnetic field B0 [2325]. However, manual repositioning of the metal is often not possible given the body region of the metal implant, the region to be imaged, and the architecture of the MRI magnet. In these cases, swapping the phase-encoding direction and the frequency-encoding direction can help to reduce artifacts by changing the direction in which the metal artifacts propagate [26]. Therefore, due to the lack of long axis and variable orientation, imaging of complex implants may be more difficult [11].

Sequence optimization

There are several basic methods to improve image quality and reduce metal artifacts when imaging metal implants. Basic sequence optimization steps include smaller nominal voxel size encoding (thinner slices, large image matrix), high receiver and RF pulse bandwidth, shorter echo times, and changing the direction of frequency and phase encoding directions. In addition, acquisition of fast spin-echo and STIR sequences rather than single spin-echo and gradient-echo sequences further reduces image artifact. However, changing the sequence parameters often leads to a reduction in SNR and the residual metal artifacts may be substantial, which is why advanced metal artifact reduction techniques such as view angle tilting (VAT), slice encoding for metal artifact correction (SEMAC), and multiacquisition variable-resonance image combination (MAVRIC) have been developed to further improve image quality (Figs. 4 and 5).

Fig. 4.

Fig. 4

Advanced metal artifact reduction imaging of the hip after total joint arthroplasty in a 67-year-old patient (AD). The implant is made of a titanium alloy. Adequate metal artifact suppression is demonstrated with a dedicated coronal slice encoding for metal artifact correction (SEMAC) turbo spin echo sequence, allowing diagnosis of the joint effusion surrounding the head of the femoral prosthesis (A, arrows). The SEMAC sequence shows the least artifacts compared to the axial short-tau inversion recovery (STIR) turbo spin echo sequence (B) with optimized inversion pulse (STIR WARP) and a high bandwidth (BW) T1-weighted turbo spin echo sequence (C) (bandwidth 425 Hz/pixel). The conventional radiograph (D) demonstrates normal fit of the prosthesis without signs of loosening. The images were acquired at 1.5 T with following sequence parameters: STIR SEMAC WARP—echo time (TE) 36 ms, repetition time (TR) 4220 ms, receiver bandwidth 500 Hz/pixel, inversion time (IT) 160 ms; transverse STIR WARP—TE 31 ms, TR 3830 ms, receiver bandwidth 450 Hz/pixel, IT 150 ms; T1 TSE High BW—TE 9 ms, TR 600 ms, receiver bandwidth 425 Hz/pixel

Fig. 5.

Fig. 5

A 61-year-old patient with shoulder arthroplasty using implants made of titanium alloys. The radiograph (A) shows the large and complex implant with signs of loosening between the cement and the bone of the proximal humerus. The coronal short tau inversion recovery (STIR) turbo spin echo sequence (B) combined with compressed sensing (CS) and slice encoding for metal artifact correction (SEMAC) demonstrate only little artifacts within an acceptable acquisition time of 6:31 min and allows assessment of the fluid and increased signal around the implant indicating low-grade inflammation (outline arrows). After administration of intravenous contrast, the axial short-tau inversion recovery (STIR) turbo spin echo sequence (C) with optimized inversion pulse (STIR WARP) and a high bandwidth (BW) T1-weighted turbo spin echo sequence (D) (bandwidth 480 Hz/pixel) show a corresponding onion-like enhancement of the surrounding tissue (C, white arrows). Joint aspiration proved a low-grade infect with Cutibacterium acnes. The images were acquired at 1.5 T with following sequence parameters: STIR SEMAC—echo time (TE) 44 ms, repetition time (TR) 4500 ms, receiver bandwidth 780 Hz/pixel, inversion time (IT) 145 ms; STIR WARP—TE 33 ms, TR 4250 ms, receiver bandwidth 390 Hz/pixel, TI 160 ms; T1 TSE High BW—TE 10 ms, TR 648 ms, receiver bandwidth 480 Hz/pixels

Choosing the right sequence

In general, fast spin-echo and turbo spin-echo sequences are preferable to gradient-echo sequences due to the rapid intravoxel dephasing of spins in gradient-echo sequences resulting in signal loss and increased artifacts [27, 28]. Spin-echo sequences exhibit a refocusing pulse applied at half of the time interval between the initial 90° flip and the echo which is able to correct in part for focal signal inhomogeneities such as metal artifacts. In contrast, gradient-echo sequences do not have a refocusing pulse. When acquiring 3-dimensional (3D) TSE sequences, it is important to remember that the slice-encoding direction is also phase-encoded and therefore non-selective or slice-selective excitation must be chosen [5, 29]. In order to reduce the potential for spatial mis-selection and through-plane distortion, non-selective excitation should be used, as it does not require a slice selection encoding gradient [30]. However, any spins outside the excitation bandwidth of the RF pulse will not be excited and these regions will appear dark. Therefore, sufficiently high RF bandwidths are also recommended for 3D non-selective sequences [31].

Receiver bandwidth

Increasing the receiver bandwidth is one of the most effective and simple measures to reduce in-plane artifacts. A higher bandwidth corresponds to a stronger gradient, which reduces signal mis-registration in the in-plane frequency encoding direction [32, 33]. Common receiver bandwidths for imaging of the knee joint with and without metal implants are illustrated in Tables 1 and 2. However, higher bandwidths require strong gradients, which is often a limitation. In addition, the receiver bandwidth is inversely related to the SNR: When the bandwidth increases, the SNR decreases. This can be compensated for by increasing the number of excitations, which prolongs the scan time and may also lead to an increase in the total RF energy applied to the patient being examined, which can be another limitation, particularly at higher field strengths [34, 35].

Table 1.

Sample protocol at 1.5-T MRI for patients with knee prostheses

Knee pulse sequences Transverse STIR Coronal STIR Sagittal IM Sagittal STIR Coronal T1
MARS technique SEMAC CS SEMAC CS SEMAC CS
Echo time (ms) 37 39 35 39 8.3
Repetition time (ms) 5220 4000 4500 4020 600
Acquisition matrix (mm) 752 × 736 320 × 320 384 × 384 320 × 320 832 × 832
Slice thickness (mm) 3.5 4 3.5 3.5 3
Flip angle (°) 135 180 180 180 135
FOV (mm) 183 × 180 200 × 200 199 × 199 200 × 200 200 × 200
Bandwidth (Hz/pixel (kHz)) 412 539 521 539 523
Inversion time (ms) 160 160 160
Averages 2 1 1 1 2
Turbo factor 9 11 19 11 4
Number of slices 46 30 32 32 32
Phase-encoding steps 263 224 307 224 333
Slice-encoding steps 15 12 12
Direction R >  > L R >  > L A >  > P H >  > F H >  > F
Acquisition time (min) 2:33 5:18 3:47 4:27 2:10

MARS metal artifact reduction sequences, STIR short tau inversion recovery, CS compressed sensing, IM intermediate weighted

Table 2.

Sample protocol of the knee at 3-T MRI for patients without metal implants

Knee pulse sequences Coronal STIR Sagittal IM Sagittal IM Transverse IM Coronal T1
Echo time (ms) 40 31 32 41 10
Repetition time (ms) 5000 3250 2810 4390 550
Acquisition matrix (mm) 384 × 269 448 × 314 512 × 358 418 × 314 512 × 410
Slice thickness (mm) 3 3 3 2.5 3
Flip angle (°) 160 135 135 180 135
FOV (mm) 159 × 159 160 × 160 199 × 199 149 × 149 159 × 159
Bandwidth (Hz/pixel (kHz)) 181 180 181 180 244
Inversion time (ms) 210
Averages 2 1 1 1 1
Turbo factor 9 7 7 7 3
Number of slices 28 31 32 40 28
Phase-encoding steps 269 314 358 314 410
Acquisition time (min) 1:45 1:18 1:18 1:19 0:58

STIR short tau inversion recovery, IM intermediate weighted

Matrix sizes and slice thickness

Other simple sequence adjustments include increasing the matrix size and decreasing the slice thickness. Increasing the matrix size results in a smaller encoded voxel size and therefore reduces in-plane distortion and intra-voxel dephasing [6, 24, 36]. It is important to note that matrix increase is most efficient when performed in the frequency-encoding direction. For smaller slice thicknesses, stronger slice selection gradients are required, which reduce through-plane artifacts. In addition, the spread of frequency differences within the smaller voxel may be reduced [7]. However, a slice thickness of less than 3 mm is associated with increased scan time, which can be compensated for by increasing the interslice spacing, but at the expense of tissue coverage. Thinner slices are also associated with decreased SNR and increased SAR; therefore, a slice thickness of 3 to 4 mm is considered adequate in clinical routine.

Frequency and phase encoding directions

Another simple adjustment to reduce metal implant artifacts is to change the direction of the phase and frequency encoding. In-plane mis-registrations usually occur along the frequency encoding direction, and changing the direction is often a simple fix to shift metal artifacts away from the region of interest [7, 37].

Fat suppression techniques

Several fat suppression techniques are available in clinical routine, the most common being chemical shift–based techniques such as spectral (selective) fat saturation and Dixon imaging, and inversion-based techniques such as short-tau inversion recovery (STIR) [38, 39]. Dixon methods such as IDEAL (Iterative Decomposition of Water and Fat with Echo Asymmetry and Least Squares Estimation) offer better fat suppression in the presence of metallic implants than standard fast spin echo sequences and, unlike STIR imaging, allow the use of contrast-enhanced MR imaging, providing better diagnostic performance in the assessment of post-operative infection [4042]. However, compared to STIR sequences, chemical shift–based techniques are more likely to fail due to their strong dependence on B0 and B1 homogeneities. In general, STIR sequences are less dependent on B0 and B1 inhomogeneities, making them the fat suppression sequences of choice for imaging near metal implants [5]. STIR is based on the different T1 relaxation times and recovery of longitudinal magnetization after application of a 180° inversion pulse up to the application of a 90° excitation pulse which reverses the intravoxel dephasing. This inversion time (TI) can also be used to modulate fat saturation [43]. As a limitation, STIR sequences generally offer a lower SNR compared to chemical shift–based fat suppression at equivalent acquisition times and are also limited in the use of post contrast imaging because enhancing tissues may have similarly short T1 relaxation times as fat and may also be suppressed [44].

Evolving techniques

View angle tilting

As an approach to correct in-plane distortions, VAT has been proposed by Cho et al. [45]. In this approach, a superimposed slice selection gradient is applied simultaneously with the readout gradient, causing a tilting of the readout encoding dimension towards the slice selection dimension. Consequently, all excited spins in the targeted slice process at the same frequency, eliminating any off-resonance induced shift along the readout direction. Although VAT does not increase scan time, it introduces some blurring, which can be diminished by reducing slice thickness [46]. As VAT depends on the ratio of slice gradient to readout gradient, the best artifact reduction is achieved at lower bandwidths. However, it should be noted that VAT does not correct for through-plane artifacts, which is considered to be its main limitation [47].

Slice encoding for metal artifact correction

Compared to conventional artifact reduction techniques, SEMAC has been shown to be an efficient technique for imaging near metal implants (Figs. 4 and 5). Although based on a 2D sequence where individual slices are excited with an RF pulse, additional phase encoding steps in the slice selection dimension are applied to resolve a larger 3D slab around each slice position. The additional information about the distortion of the slices can then be used to correct for through-plane distortion of the acquired slice and of the adjacent slices [4850]. The number of additional phase encoding steps can be chosen individually for each scan, depending on the size and complexity of the metal implant and the area of field inhomogeneity. However, increasing the number of steps also increases the scan time, which ultimately leads to a trade-off between the degree of artifact reduction and acquisition times. In literature, it has been proposed that 11 slice-encoding steps (for T2-weighted images) to 19 slice-encoding steps (for STIR and T1-weighted images) might be sufficient for artifact reduction around hip prostheses [5, 51]. The SEMAC technique addresses through-plane artifacts, but it is often implemented with VAT to address in-plane artifacts as well.

Multiacquisition variable-resonance image combination

MAVRIC has been developed to address metal-induced in-plane and through-plane distortions caused by the field inhomogeneities surrounding the implant [52, 53]. It uses a 3D acquisition technique with a spatially non-selective excitation which—in contrast to gradient-based slice selection—performs phase encoding along two dimensions with reduced distortion. In general, spin precession around metal implants occurs over a wide range of frequencies due to B0 inhomogeneities; the bandwidth of a single non-selective RF pulse is not wide enough to cover the full range of off-resonance frequencies and the lack of excitation would result in signal loss [54]. To overcome this problem, MAVRIC acquires spectral bins, which are essentially 3D slabs with slightly different resonance frequencies. The acquired 3D slabs are combined into a composite image using a sum-of-squares or maximum intensity projection scheme. On the down side, MAVRIC lacks slice selectivity which requires time-consuming 3D imaging.

Hybrid approaches and future developments

Several hybrid approaches have been proposed to overcome some of the limitations of the techniques themselves [55, 56]. One variant combines the slab selectivity of SEMAC with the smooth bin combination and higher SNR implemented by MAVRIC. The so-called MAVRIC SL technique combines the slice-direction phase encoding of SEMAC with the increased spectral coverage and has demonstrated improved image quality and reduced image artifacts when imaging patients with total hip and shoulder prosthesis (Table 3).

Table 3.

Sample protocol at 1.5-T MRI for patients with shoulder prostheses

Shoulder pulse sequences Transverse T1 Transverse STIR Coronal STIR Sagittal IM
MARS technique SEMAC CS SEMAC CS
Echo time (ms) 6.6 38 44 35
Repetition time (ms) 685 4000 5200 5000
Acquisition matrix (mm) 384 × 384 320 × 320 256 × 256 256 × 256
Slice thickness (mm) 4 4 4 3.5
Flip angle (°) 132 135 140 180
FOV (mm) 180 × 180 180 × 180 200 × 200 200 × 200
Bandwidth (Hz/pixel (kHz)) 401 402 539 521
Inversion time (ms) 160 160
Averages 1 1 1 1
Turbo factor 3 12 11 19
Number of slices 30 36 20 25
Phase-encoding steps 619 299 302 261
Slice-encoding steps - 12 12 -
Direction A >  > P A >  > P A >  > P R >  > L
Acquisition time (min) 2:47 2:26 5:45 4:12

MARS metal artifact reduction sequences, STIR short tau inversion recovery, CS compressed sensing, IM intermediate weighted

In order to reduce the scan time of SEMAC, a combination with off-resonance suppression (ORS) was proposed [57]. In this technique, separate RF bandwidths and gradient strengths of the excitation and refocusing pulses are applied. Therefore, the range of spins which contribute to the image are limited and fewer phase-encoding steps are necessary for SEMAC. This approach limits back-folding artifacts and enables flexibility of scan orientation with the disadvantages of signal voids and loss of homogeneity.

As described above, the large static gradients in the B0 field in close proximity to metal implants result in increased intravoxel dephasing and rapid T2* decay. Ultra-short echo time allows imaging of tissues with very short T2 and allows the signal to be acquired before it has dephased [58]. A hybrid technique combining UTE and MAVRIC has been proposed for imaging of hip prosthesis, attempting to combine the advantages of multispectral imaging of MAVRIC and a non-selective 3D UTE [59]. Further research has focused on the development of an externally calibrated parallel imaging technique for three-dimensional multispectral imaging (3D-MSI) using broadband UTE [60]. The technique allows for a significant reduction in scan time while maintaining similar metal artifact reduction as conventional MAVRIC acquisitions (Table 4).

Table 4.

Sample protocol at 1.5-T MRI for patients with minor surgical implants of the shoulder such as anchors or screws

Shoulder pulse sequences Coronal IM Coronal STIR Sagittal T1 Sagittal STIR Transversal IM Transversal STIR
Echo time (ms) 29 35 10 35 29 35
Repetition time (ms) 3210 4000 548 4000 3200 4000
Acquisition matrix (mm) 768 × 768 640 × 640 768 × 768 640 × 640 768 × 768 640 × 640
Slice thickness (mm) 3 3 4 4 3 3
Flip angle (°) 135 135 135 135 135 135
FOV (mm) 159 × 159 160 × 160 160 × 160 160 × 160 159 × 159 160 × 160
Bandwidth (Hz/pixel (kHz)) 449 411 407 411 407 401
Inversion time (ms) 160 160 160
Averages 2 2 1 2 2 2
Turbo factor 14 9 3 12 12 11
Number of slices 27 27 29 23 30 28
Phase-encoding steps 307 224 288 224 307 224
Phase encoding direction F >  > H F >  > H F >  > H H >  > F A >  > P A >  > P
Acquisition time (min:s) 1:47 2:54 1:27 2:14 1:58 3:06

STIR short tau inversion recovery, IM intermediate weighted

One drawback of advanced and hybrid metal artifact reduction techniques is the often long scan time. Two main approaches have been investigated to overcome this problem (Figs. 5 and 6). Compressed sensing (CS) has been used to accelerate SEMAC acquisition and demonstrated feasibility for imaging of total hip prosthesis and other anatomic regions, and for the differentiation between normal postoperative MRI findings and abnormal findings after total hip prosthesis [6164]. The combination of CS and SEMAC has been shown to reduce acquisition times to 5–6 min for hip prostheses, allowing it to be incorporated easily into routine imaging protocols. At the same time, image quality is improved compared to standard SEMAC pulse sequences because of the increased number of slice-encoding steps. In addition, its high accuracy in detecting periprosthetic pathologies, such as infections, makes it a reliable sequence for assessing periprosthetic complications of large implants [50, 65, 66]. For MAVRIC imaging near metal implants, a short spectral calibration scan of about 35 s can be used for optimizing the number of spectral bins to minimize susceptibility effects and reduce scan time [17, 67]. For the combined MAVRIC-SL sequence, this calibration scan allowed to achieve scan times of 5–6 min for conventional acquisitions of hip prostheses compared to more than 8 min for MAVRIC imaging alone, and with an additional decrease of the repetition time an isotropic MAVRIC-SL was possible in 7:16 min with improved SNR [56] (Table 5).

Fig. 6.

Fig. 6

1.5-T MRI of the right hip of a 57-year-old patient 1.5 years after total hip arthroplasty. The radiograph shows a normal position of the implant made of titanium alloy (A). However, the coronal short-tau inversion recovery (STIR) turbo spin echo sequence (B) combined with compressed sensing (CS) and slice encoding for metal artifact correction (SEMAC) shows severe joint inflammation with increased synovial fluid, extensive bone marrow edema (arrowhead and outline arrow), and strongly hyperintense signal of the surrounding soft tissues (arrow); this imaging pattern is highly suspicious of a periprosthetic joint infection. The axial short-tau inversion recovery (STIR) turbo spin echo sequence (C) with optimized inversion pulse (STIR WARP) shows abscess collections within the anterior surgical access route. Joint aspiration proved a low-grade infect with Cutibacterium avidum. The images were acquired at 1.5 T with following sequence parameters: STIR SEMAC CS—echo time (TE) 36 ms, repetition time (TR) 4220 ms, receiver bandwidth 500 Hz/pixel, inversion time (IT) 160 ms; STIR WARP—TE 31 ms, TR 4000 ms, receiver bandwidth 450 Hz/pixel, IT 150 ms

Table 5.

Sample protocol at 1.5-T MRI for patients with hip prostheses

Hip pulse sequences Transverse STIR Coronal STIR Transverse T1 Coronal T2 Sagittal T1
MARS technique SEMAC CS
Echo time (ms) 38 37 16 54 11
Repetition time (ms) 5040 5000 500 4480 593
Acquisition matrix (mm) 383 × 384 256 × 256 512 × 512 512 × 512 320 × 320
Slice thickness (mm) 7 3.5 6 4 4
Flip angle (°) 135 180 135 150 170
FOV (mm) 180 × 180 280 × 280 200 × 200 220 × 220 200 × 200
Bandwidth (Hz/pixel (kHz)) 411 539 406 413 407
Inversion time (ms) 170 145
Averages 3 1 1 1 1
Turbo factor 9 13 3 18 3
Number of slices 30 28 33 24 34
Phase-encoding steps 419 359 619 571 576
Slice-encoding steps 12
Direction A >  > P R >  > L R >  > L R >  > L H >  > F
Acquisition time (min) 1:37 4:42 1:59 2:46 1:25

MARS metal artifact reduction sequences, STIR short tau inversion recovery, CS compressed sensing, IM intermediate weighted

Recently, the use of deep learning to reconstruct under-sampled MR data with the aim of accelerating image acquisition, reducing image noise, and improving image quality has attracted attention [6870]. However, clinical applications of deep learning reconstruction for metal artifact reduction in MRI are still sparse and have yet to prove applicability in clinical routine [71].

Summary

Advances in MRI around metal implants have significantly improved the visualization and diagnosis of implant-related abnormalities that cannot be assessed with other modalities. There are a variety of measures that can be taken to improve image quality around metal implants, but it is important to consider the material properties of the prosthesis, its geometry, and the clinical questions that need to be answered first, in order to use the correct sequences. Although metal artifacts are reduced at lower field strengths, today 1.5-T MRI is the preferred field strength for prosthesis imaging. For smaller, non-complex metal implants, basic metal artifact reduction techniques (increased receive bandwidth, reduced slice thickness, multi-echo spin-echo sequences, and STIR for fat suppression instead of frequency selective fat saturation) may be sufficient to obtain diagnostic images. However, advanced techniques such as CS-SEMAC and MAVRIC-SL should be used for imaging of large and complex metal implants such as joint prostheses, enabling a strong artifact reduction and reduced scan times. Finally, future developments in prosthesis composites and the application of evolving imaging techniques such as deep learning applications for metal artifact reduction may further improve imaging around orthopedic hardware.

Acknowledgements

The authors gratefully acknowledge the thorough editing of the manuscript by Sabine Schrimpf, PhD (Department of Radiology, Balgrist University Hospital, Zurich, Switzerland).

Funding

Open access funding provided by University of Zurich

Data Availability

The data that support the findings of this study are not openly available due to reasons of sensitivity and are available from the corresponding author upon reasonable request.

Declarations

Conflict of interest

The author’s institution has an academic research collaboration with Siemens Healthineers. However, Siemens employees were not involved in drafting or editing of this manuscript.

Footnotes

Key points

• Advances in metal artifact reduction techniques such as bandwidth optimization, view angle tilting, or slice-encoding for metal artifact correction have substantially improved image quality of MRI around metal implants.

• STIR imaging offers better fat suppression in the presence of metal implants than chemical shift–based methods with the disadvantage of lower SNR at comparable scan times.

• Slice encoding for metal artifact correction sequences combined with compressed sensing allows the precise imaging of periprosthetic complications.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Shichman I, Roof M, Askew N, Nherera L, Rozell JC, Seyler TM, et al. Projections and epidemiology of primary hip and knee arthroplasty in medicare patients to 2040–2060. JBJS Open Access. 2023;8(1):e22.00112. [DOI] [PMC free article] [PubMed]
  • 2.Zubrod CJ, Barrett MF. Magnetic resonance imaging of tendon and ligament injuries. Clin Tech Equine Pract. 2007;6(3):217–29. 10.1053/j.ctep.2007.08.008 [DOI] [Google Scholar]
  • 3.Umans H, Cerezal L, Linklater J, Fritz J. Postoperative MRI of the ankle and foot. Magn Reson Imaging Clin N Am. 2022;30(4):733–55. 10.1016/j.mric.2022.05.006 [DOI] [PubMed] [Google Scholar]
  • 4.Lee EM, Ibrahim EH, Dudek N, Lu JC, Kalia V, Runge M, et al. Improving MR image quality in patients with metallic implants. Radiographics. 2021;41(4):E126-e137. 10.1148/rg.2021200092 [DOI] [PubMed] [Google Scholar]
  • 5.Germann C, Nanz D, Sutter R. Magnetic resonance imaging around metal at 1.5 Tesla: techniques from basic to advanced and clinical impact. Invest Radiol. 2021;56(11):734–48. 10.1097/RLI.0000000000000798 [DOI] [PubMed] [Google Scholar]
  • 6.Jungmann PM, Agten CA, Pfirrmann CW, Sutter R. Advances in MRI around metal. J Magn Reson Imaging: JMRI. 2017;46(4):972–91. 10.1002/jmri.25708 [DOI] [PubMed] [Google Scholar]
  • 7.Khodarahmi I, Isaac A, Fishman EK, Dalili D, Fritz J. Metal about the hip and artifact reduction techniques: from basic concepts to advanced imaging. Semin Musculoskelet Radiol. 2019;23(3):e68–81. 10.1055/s-0039-1687898 [DOI] [PubMed] [Google Scholar]
  • 8.Hargreaves BA, Worters PW, Pauly KB, Pauly JM, Koch KM, Gold GE. Metal-induced artifacts in MRI. Am J Roentgenol. 2011;197(3):547–55. 10.2214/AJR.11.7364 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Buckwalter KA. Optimizing imaging techniques in the postoperative patient. Semin Musculoskelet Radiol. 2007;11(3):261–72. 10.1055/s-2008-1038315 [DOI] [PubMed] [Google Scholar]
  • 10.Koch KM, Hargreaves BA, Pauly KB, Chen W, Gold GE, King KF. Magnetic resonance imaging near metal implants. J Magn Reson Imaging: JMRI. 2010;32(4):773–87. 10.1002/jmri.22313 [DOI] [PubMed] [Google Scholar]
  • 11.Guermazi A, Miaux Y, Zaim S, Peterfy CG, White D, Genant HK. Metallic artefacts in MR imaging: effects of main field orientation and strength. Clin Radiol. 2003;58(4):322–8. 10.1016/S0009-9260(02)00540-8 [DOI] [PubMed] [Google Scholar]
  • 12.Germann C, Falkowski AL, von Deuster C, Nanz D, Sutter R. Basic and advanced metal-artifact reduction techniques at ultra-high field 7-T magnetic resonance imaging-phantom study investigating feasibility and efficacy. Invest Radiol. 2022;57(6):387–98. 10.1097/RLI.0000000000000850 [DOI] [PubMed] [Google Scholar]
  • 13.Khodarahmi I, Keerthivasan MB, Brinkmann IM, Grodzki D, Fritz J. Modern low-field MRI of the musculoskeletal system: practice considerations, opportunities, and challenges. Invest Radiol. 2023;58(1):76–87. 10.1097/RLI.0000000000000912 [DOI] [PubMed] [Google Scholar]
  • 14.Koff MF, Shah P, Koch KM, Potter HG. Quantifying image distortion of orthopedic materials in magnetic resonance imaging. J Magn Reson Imaging: JMRI. 2013;38(3):610–8. 10.1002/jmri.23991 [DOI] [PubMed] [Google Scholar]
  • 15.Månsson S, Müller GM, Wellman F, Nittka M, Lundin B. Phantom based qualitative and quantitative evaluation of artifacts in MR images of metallic hip prostheses. Phys Med. 2015;31(2):173–8. 10.1016/j.ejmp.2014.12.001 [DOI] [PubMed] [Google Scholar]
  • 16.Ahmad FU, Sidani C, Fourzali R, Wang MY. Postoperative magnetic resonance imaging artifact with cobalt-chromium versus titanium spinal instrumentation: presented at the 2013 Joint Spine Section Meeting. Clin Artic J Neurosurg Spine. 2013;19(5):629–36. 10.3171/2013.7.SPINE1359 [DOI] [PubMed] [Google Scholar]
  • 17.Filli L, Jud L, Luechinger R, Nanz D, Andreisek G, Runge VM, et al. Material-dependent implant artifact reduction using SEMAC-VAT and MAVRIC: a prospective MRI phantom study. Invest Radiol. 2017;52(6):381–7. 10.1097/RLI.0000000000000351 [DOI] [PubMed] [Google Scholar]
  • 18.Kim SC, Lee HJ, Son SG, Seok HK, Lee KS, Shin SY, et al. Aluminum-free low-modulus Ti-C composites that exhibit reduced image artifacts during MRI. Acta Biomater. 2015;12:322–31. 10.1016/j.actbio.2014.10.016 [DOI] [PubMed] [Google Scholar]
  • 19.Jin Z, Guan Y, Yu G, Sun Y. Magnetic resonance imaging of postoperative fracture healing process without metal artifact: a preliminary report of a novel animal model. Biomed Res Int. 2016;2016:1429892. 10.1155/2016/1429892 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Zimel MN, Hwang S, Riedel ER, Healey JH. Carbon fiber intramedullary nails reduce artifact in postoperative advanced imaging. Skelet Radiol. 2015;44(9):1317–25. 10.1007/s00256-015-2158-9 [DOI] [PubMed] [Google Scholar]
  • 21.Ernstberger T, Buchhorn G, Heidrich G. Artifacts in spine magnetic resonance imaging due to different intervertebral test spacers: an in vitro evaluation of magnesium versus titanium and carbon-fiber-reinforced polymers as biomaterials. Neuroradiology. 2009;51(8):525–9. 10.1007/s00234-009-0537-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Filli L, Luechinger R, Frauenfelder T, Beck S, Guggenberger R, Farshad-Amacker N, et al. Metal-induced artifacts in computed tomography and magnetic resonance imaging: comparison of a biodegradable magnesium alloy versus titanium and stainless steel controls. Skelet Radiol. 2015;44(6):849–56. 10.1007/s00256-014-2057-5 [DOI] [PubMed] [Google Scholar]
  • 23.Ganapathi M, Joseph G, Savage R, Jones AR, Timms B, Lyons K. MRI susceptibility artefacts related to scaphoid screws: the effect of screw type, screw orientation and imaging parameters. J Hand Surg Br. 2002;27(2):165–70. 10.1054/JHSB.2001.0717 [DOI] [PubMed] [Google Scholar]
  • 24.Lee MJ, Kim S, Lee SA, Song HT, Huh YM, Kim DH, et al. Overcoming artifacts from metallic orthopedic implants at high-field-strength MR imaging and multi-detector CT. Radiographics. 2007;27(3):791–803. 10.1148/rg.273065087 [DOI] [PubMed] [Google Scholar]
  • 25.Suh JS, Jeong EK, Shin KH, Cho JH, Na JB, Kim DH, et al. Minimizing artifacts caused by metallic implants at MR imaging: experimental and clinical studies. AJR Am J Roentgenol. 1998;171(5):1207–13. 10.2214/ajr.171.5.9798849 [DOI] [PubMed] [Google Scholar]
  • 26.Sperling JW, Potter HG, Craig EV, Flatow E, Warren RF. Magnetic resonance imaging of painful shoulder arthroplasty. J Shoulder Elbow Surg. 2002;11(4):315–21. 10.1067/mse.2002.124426 [DOI] [PubMed] [Google Scholar]
  • 27.Eustace S, Jara H, Goldberg R, Fenlon H, Mason M, Melhem ER, et al. A comparison of conventional spin-echo and turbo spin-echo imaging of soft tissues adjacent to orthopedic hardware. AJR Am J Roentgenol. 1998;170(2):455–8. 10.2214/ajr.170.2.9456963 [DOI] [PubMed] [Google Scholar]
  • 28.Ariyanayagam T, Malcolm PN, Toms AP. Advances in metal artifact reduction techniques for periprosthetic soft tissue imaging. Semin Musculoskelet Radiol. 2015;19(4):328–34. 10.1055/s-0035-1563734 [DOI] [PubMed] [Google Scholar]
  • 29.Chang EY, Bae WC, Chung CB. Imaging the knee in the setting of metal hardware. Magn Reson Imaging Clin N Am. 2014;22(4):765–86. 10.1016/j.mric.2014.07.009 [DOI] [PubMed] [Google Scholar]
  • 30.Naraghi AM, White LM. Magnetic resonance imaging of joint replacements. Semin Musculoskelet Radiol. 2006;10(1):98–106. 10.1055/s-2006-934220 [DOI] [PubMed] [Google Scholar]
  • 31.Sofka CM, Potter HG. MR imaging of joint arthroplasty. Semin Musculoskelet Radiol. 2002;6(1):79–85. 10.1055/s-2002-23166 [DOI] [PubMed] [Google Scholar]
  • 32.Ulbrich EJ, Sutter R, Aguiar RF, Nittka M, Pfirrmann CW. STIR sequence with increased receiver bandwidth of the inversion pulse for reduction of metallic artifacts. AJR Am J Roentgenol. 2012;199(6):W735-742. 10.2214/AJR.11.8233 [DOI] [PubMed] [Google Scholar]
  • 33.Kumar NM, de Cesar NC, Schon LC, Fritz J. Metal artifact reduction magnetic resonance imaging around arthroplasty implants: the negative effect of long echo trains on the implant-related artifact. Invest Radiol. 2017;52(5):310–6. 10.1097/RLI.0000000000000350 [DOI] [PubMed] [Google Scholar]
  • 34.Toms AP, Smith-Bateman C, Malcolm PN, Cahir J, Graves M. Optimization of metal artefact reduction (MAR) sequences for MRI of total hip prostheses. Clin Radiol. 2010;65(6):447–52. 10.1016/j.crad.2009.12.014 [DOI] [PubMed] [Google Scholar]
  • 35.Lu W, Pauly KB, Gold GE, Pauly JM, Hargreaves BA. SEMAC: slice encoding for metal artifact correction in MRI. Magn Reson Med. 2009;62(1):66–76. 10.1002/mrm.21967 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Talbot BS, Weinberg EP. MR imaging with metal-suppression sequences for evaluation of total joint arthroplasty. Radiographics. 2016;36(1):209–25. 10.1148/rg.2016150075 [DOI] [PubMed] [Google Scholar]
  • 37.Khodarahmi I, Fishman EK, Fritz J. Dedicated CT and MRI Techniques for the evaluation of the postoperative knee. Semin Musculoskelet Radiol. 2018;22(4):444–56. 10.1055/s-0038-1653955 [DOI] [PubMed] [Google Scholar]
  • 38.Del Grande F, Santini F, Herzka DA, Aro MR, Dean CW, Gold GE, et al. Fat-suppression techniques for 3-T MR imaging of the musculoskeletal system. Radiographics. 2014;34(1):217–33. 10.1148/rg.341135130 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Haase A, Frahm J, Hänicke W, Matthaei D. 1H NMR chemical shift selective (CHESS) imaging. Phys Med Biol. 1985;30(4):341–4. 10.1088/0031-9155/30/4/008 [DOI] [PubMed] [Google Scholar]
  • 40.Molière S, Dillenseger JP, Ehlinger M, Kremer S, Bierry G. Comparative study of fat-suppression techniques for hip arthroplasty MR imaging. Skelet Radiol. 2017;46(9):1209–17. 10.1007/s00256-017-2670-1 [DOI] [PubMed] [Google Scholar]
  • 41.Park HJ, Lee SY, Choi SH, Hong HP, Choi YJ, Kim MS. Reduced metallic artefacts in 3 T knee MRI using fast spin-echo multi-point Dixon compared to fast spin-echo T2-weighted sequences. Clin Radiol. 2017;72(11):996.e991-996.e996. 10.1016/j.crad.2017.07.012 [DOI] [PubMed] [Google Scholar]
  • 42.Lee JB, Cha JG, Lee MH, Lee YK, Lee EH, Jeon CH. Usefulness of IDEAL T2-weighted FSE and SPGR imaging in reducing metallic artifacts in the postoperative ankles with metallic hardware. Skelet Radiol. 2013;42(2):239–47. 10.1007/s00256-012-1449-7 [DOI] [PubMed] [Google Scholar]
  • 43.Bydder GM, Pennock JM, Steiner RE, Khenia S, Payne JA, Young IR. The short TI inversion recovery sequence–an approach to MR imaging of the abdomen. Magn Reson Imaging. 1985;3(3):251–4. 10.1016/0730-725X(85)90354-6 [DOI] [PubMed] [Google Scholar]
  • 44.Fischer T, Baz YE, Waelti S, Wildermuth S, Leschka S, Güsewell S, et al. Short tau inversion recovery (STIR) after intravenous contrast agent administration obscures bone marrow edema-like signal on forefoot MRI. Skelet Radiol. 2022;51(3):573–9. 10.1007/s00256-021-03852-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Cho ZH, Kim DJ, Kim YK. Total inhomogeneity correction including chemical shifts and susceptibility by view angle tilting. Med Phys. 1988;15(1):7–11. 10.1118/1.596162 [DOI] [PubMed] [Google Scholar]
  • 46.Butts K, Pauly JM, Gold GE. Reduction of blurring in view angle tilting MRI. Magn Reson Med. 2005;53(2):418–24. 10.1002/mrm.20375 [DOI] [PubMed] [Google Scholar]
  • 47.Ai T, Padua A, Goerner F, Nittka M, Gugala Z, Jadhav S, et al. SEMAC-VAT and MSVAT-SPACE sequence strategies for metal artifact reduction in 1.5T magnetic resonance imaging. Invest Radiol. 2012;47(5):267–76. 10.1097/RLI.0b013e318240a919 [DOI] [PubMed] [Google Scholar]
  • 48.Sutter R, Hodek R, Fucentese SF, Nittka M, Pfirrmann CW. Total knee arthroplasty MRI featuring slice-encoding for metal artifact correction: reduction of artifacts for STIR and proton density-weighted sequences. AJR Am J Roentgenol. 2013;201(6):1315–24. 10.2214/AJR.13.10531 [DOI] [PubMed] [Google Scholar]
  • 49.Sutter R, Ulbrich EJ, Jellus V, Nittka M, Pfirrmann CW. Reduction of metal artifacts in patients with total hip arthroplasty with slice-encoding metal artifact correction and view-angle tilting MR imaging. Radiology. 2012;265(1):204–14. 10.1148/radiol.12112408 [DOI] [PubMed] [Google Scholar]
  • 50.Jungmann PM, Bensler S, Zingg P, Fritz B, Pfirrmann CW, Sutter R. Improved visualization of juxtaprosthetic tissue using metal artifact reduction magnetic resonance imaging: experimental and clinical optimization of compressed sensing SEMAC. Invest Radiol. 2019;54(1):23–31. 10.1097/RLI.0000000000000504 [DOI] [PubMed] [Google Scholar]
  • 51.Lee YH, Lim D, Kim E, Kim S, Song HT, Suh JS. Usefulness of slice encoding for metal artifact correction (SEMAC) for reducing metallic artifacts in 3-T MRI. Magn Reson Imaging. 2013;31(5):703–6. 10.1016/j.mri.2012.11.004 [DOI] [PubMed] [Google Scholar]
  • 52.Hayter CL, Koff MF, Shah P, Koch KM, Miller TT, Potter HG. MRI after arthroplasty: comparison of MAVRIC and conventional fast spin-echo techniques. AJR Am J Roentgenol. 2011;197(3):W405-411. 10.2214/AJR.11.6659 [DOI] [PubMed] [Google Scholar]
  • 53.Koch KM, Lorbiecki JE, Hinks RS, King KF. A multispectral three-dimensional acquisition technique for imaging near metal implants. Magn Reson Med. 2009;61(2):381–90. 10.1002/mrm.21856 [DOI] [PubMed] [Google Scholar]
  • 54.Hargreaves BA, Taviani V, Litwiller DV, Yoon D. 2D multi-spectral imaging for fast MRI near metal. Magn Reson Med. 2018;79(2):968–73. 10.1002/mrm.26724 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 55.Koch KM, Brau AC, Chen W, Gold GE, Hargreaves BA, Koff M, et al. Imaging near metal with a MAVRIC-SEMAC hybrid. Magn Reson Med. 2011;65(1):71–82. 10.1002/mrm.22523 [DOI] [PubMed] [Google Scholar]
  • 56.Zochowski KC, Miranda MA, Cheung J, Argentieri EC, Lin B, Kaushik SS, et al. MRI of hip arthroplasties: comparison of isotropic Multiacquisition Variable-Resonance Image Combination Selective (MAVRIC SL) acquisitions with a conventional MAVRIC SL acquisition. AJR Am J Roentgenol. 2019;213(6):W277-w286. 10.2214/AJR.19.21606 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 57.den Harder JC, van Yperen GH, Blume UA, Bos C. Off-resonance suppression for multispectral MR imaging near metallic implants. Magn Reson Med. 2015;73(1):233–43. 10.1002/mrm.25126 [DOI] [PubMed] [Google Scholar]
  • 58.Chang EY, Du J, Chung CB. UTE imaging in the musculoskeletal system. J Magn Reson Imaging. 2015;41(4):870–83. 10.1002/jmri.24713 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 59.Koch KM, King KF, Carl M, Hargreaves BA. Imaging near metal: the impact of extreme static local field gradients on frequency encoding processes. Magn Reson Med. 2014;71(6):2024–34. 10.1002/mrm.24862 [DOI] [PubMed] [Google Scholar]
  • 60.Wiens CN, Artz NS, Jang H, McMillan AB, Reeder SB. Externally calibrated parallel imaging for 3D multispectral imaging near metallic implants using broadband ultrashort echo time imaging. Magn Reson Med. 2017;77(6):2303–9. 10.1002/mrm.26327 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 61.Khodarahmi I, Khanuja HS, Stern SE, Carrino JA, Fritz J. Compressed sensing SEMAC MRI of hip, knee, and ankle arthroplasty implants: a 1.5-T and 3-T intrapatient performance comparison for diagnosing periprosthetic abnormalities. AJR Am J Roentgenol. 2023;221(5):661–72. 10.2214/AJR.23.29380 [DOI] [PubMed] [Google Scholar]
  • 62.Fritz J, Ahlawat S, Demehri S, Thawait GK, Raithel E, Gilson WD, et al. Compressed sensing SEMAC: 8-fold accelerated high resolution metal artifact reduction MRI of cobalt-chromium knee arthroplasty implants. Invest Radiol. 2016;51(10):666–76. 10.1097/RLI.0000000000000317 [DOI] [PubMed] [Google Scholar]
  • 63.Filli L, Jungmann PM, Zingg PO, Rüdiger HA, Galley J, Sutter R, et al. MRI with state-of-the-art metal artifact reduction after total hip arthroplasty: periprosthetic findings in asymptomatic and symptomatic patients. Eur Radiol. 2020;30(4):2241–52. 10.1007/s00330-019-06554-5 [DOI] [PubMed] [Google Scholar]
  • 64.Germann C, Filli L, Jungmann PM, Graf DN, Fritz J, Pfirrmann CWA, et al. Prospective and longitudinal evolution of postoperative periprosthetic findings on metal artifact-reduced MR imaging in asymptomatic patients after uncemented total hip arthroplasty. Skelet Radiol. 2021;50(6):1177–88. 10.1007/s00256-020-03666-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 65.Fritz J, Fritz B, Thawait GK, Raithel E, Gilson WD, Nittka M, et al. Advanced metal artifact reduction MRI of metal-on-metal hip resurfacing arthroplasty implants: compressed sensing acceleration enables the time-neutral use of SEMAC. Skelet Radiol. 2016;45(10):1345–56. 10.1007/s00256-016-2437-0 [DOI] [PubMed] [Google Scholar]
  • 66.Galley J, Sutter R, Stern C, Filli L, Rahm S, Pfirrmann CWA. Diagnosis of periprosthetic hip joint infection using MRI with metal artifact reduction at 1.5 T. Radiology. 2020;296(1):98–108. 10.1148/radiol.2020191901 [DOI] [PubMed] [Google Scholar]
  • 67.Kaushik SS, Marszalkowski C, Koch KM. External calibration of the spectral coverage for three-dimensional multispectral MRI. Magn Reson Med. 2016;76(5):1494–503. 10.1002/mrm.26065 [DOI] [PubMed] [Google Scholar]
  • 68.Feuerriegel GC, Weiss K, Kronthaler S, Leonhardt Y, Neumann J, Wurm M, et al. Evaluation of a deep learning-based reconstruction method for denoising and image enhancement of shoulder MRI in patients with shoulder pain. Eur Radiol. 2023;33(7):4875–84. 10.1007/s00330-023-09472-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 69.Almansour H, Herrmann J, Gassenmaier S, Afat S, Jacoby J, Koerzdoerfer G, et al. Deep learning reconstruction for accelerated spine MRI: prospective analysis of interchangeability. Radiology. 2023;306(3):e212922. [DOI] [PubMed]
  • 70.Sneag DB, Abel F, Potter HG, Fritz J, Koff MF, Chung CB, et al. MRI advancements in musculoskeletal clinical and research practice. Radiology. 2023;308(2):e230531. 10.1148/radiol.230531 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 71.Arabi H, Zaidi H. Truncation compensation and metallic dental implant artefact reduction in PET/MRI attenuation correction using deep learning-based object completion. Phys Med Biol. 2020;65(19):195002. 10.1088/1361-6560/abb02c [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The data that support the findings of this study are not openly available due to reasons of sensitivity and are available from the corresponding author upon reasonable request.


Articles from Skeletal Radiology are provided here courtesy of Springer

RESOURCES