Standard parameters of the chest ULD-CT protocol

Parameter ULD-CT
Tube voltage 100 kV
Tube current modulation range 10–740 mA
Noise Index 85
CTDIvol 0.41
Scan time 0.5 s
Pitch 0.5
Collimation 128 × 0.625 mm
Kernel Lung, mediastinum
ASIR-V 50%

ULD-CT: Ultralow-dose computed tomography

CTDIvol: Volumetric Computed Tomography Dose Index