Skip to main content
. 2024 Aug 7;14:18318. doi: 10.1038/s41598-024-68862-1

Table 3.

Performance of the reference network on different SBP ranges on the MIMIC dataset. Constraining the data range can result in significant (but artificial) accuracy improvements.

SBP Range (mmHg) Bias (mmHg) SD (mmHg)
65–200 −3.45 15.8
75–165 −4.59 14.0
75–150 −4.42 12.4