Table 3.

Difference and correlation between mean gradient measurements obtained by different modalities before and after transcatheter aortic valve replacement.

Modality Absolute mean differencea P value Pearson correlation
Pre-TAVR mean gradient
 OpSens vs cath 2.2 ​± ​3.5 1.00 0.96
 OpSens vs TEE 3.0 ​± ​3.2 .30 0.96
 OpSens vs TTE 7.8 ​± ​7.5 .32 0.70
 Cath vs TEE 3.5 ​± ​3.0 .33 0.96
 Cath vs TTE 6.7 ​± ​5.1 .20 0.84
 TEE vs TTE 5.8 ​± ​6.0 .44 0.81
Post-TAVR mean gradient
 OpSens vs cath 1.2 ​± ​1.3 .16 0.89
 OpSens vs TEE 3.1 ​± ​1.9 <.0001 0.61
 OpSens vs TTE 3.2 ​± ​1.7 <.0001 0.71
 Cath vs TEE 3.9 ​± ​2.0 <.0001 0.55
 Cath vs TTE 3.6 ​± ​2.2 <.0001 0.75
 TEE vs TTE 1.8 ​± ​1.7 1.00 0.63

Cath, catheterization using 2-pigtail catheters; OpSens, OpSens OptoWire III; TAVR, transcatheter aortic valve replacement; TEE, transesophageal echocardiogram; TTE, transthoracic echocardiogram.

a

Values are given as mm Hg ​± ​standard deviation.