Skip to main content
. 2022 Sep 17;1(6):100449. doi: 10.1016/j.jscai.2022.100449

Table 2.

Hemodynamic impact of ModulHeart support

No support ModulHeart support P value
Proximal systolic pressure, mm Hg 109 ± 25 110 ± 24 .93
Distal systolic pressure, mm Hg 109 ± 25 116 ± 24 .47
Proximal diastolic pressure, mm Hg 52 ± 18 54 ± 13 .81
Distal diastolic pressure, mm Hg 52 ± 18 58 ± 12 .40
Pump pressure gradient
 Systolic, mm Hg 0 6 ± 3 <.0001
 Diastolic, mm Hg 0 4 ± 2 <.0001
 Mean, mm Hg 0 5 ± 2 <.0001
PAP systolic, mm Hg 48 ± 10 42 ± 11 .24
PAP diastolic, mm Hg 12 ± 6 7 ± 5 .05
PAP mean, mm Hg 24 ± 5 19 ± 5 .03
Central venous pressure, mm Hg 10 ± 4 6 ± 3 .05
LVEDP, mm Hg 9 ± 5 2 ± 1 .006
Cardiac index, L/min/m2 2.4 ± 0.3 3.1 ± 1 .47

Values are mean ± SD.

LVEDP, left ventricle end-diastolic pressure; PAP, pulmonary artery pressure.