Skip to main content
. 2023 Oct 19;3(1):101109. doi: 10.1016/j.jscai.2023.101109

Table 3.

Physician and staff radiation doses.

Dosimeter location Novel system (μSv) Control (μSv) P
Primary operator
 Head-level median dose 0.0 [0.0, 0.0] 2.1 [0.7, 3.3] <.001
 Head-level min dose 0.0 0.3
 Head-level max dose 0.3 8.9
 Waist-level median dose 0.1 [0.0, 0.2] 0.0 [0.0, 0.0] <.001
 Waist-level min dose 0.0 0.0
 Waist-level max dose 0.8 0.0
Scrub technologist
 Head-level median dose 0.0 [0.0, 0.0] 0.3 [0.1, 0.4] <.001
 Head-level min dose 0.0 0.0
 Head-level max dose 0.1 1.1
 Waist-level 0.0 [0.0, 0.0] 0.0 [0.0, 0.0] .31
 Waist-level min dose 0.0 0.0
 Waist-level max dose 0.1 0.8
Circulating nurse
 Head-level median dose 0.0 [0.0, 0.0] 0.1 [0.0, 0.2] <.001
 Head-level min dose 0.0 0.0
 Head-level max dose 0.0 0.5

Values in brackets are [25th percentile, 75th percentile].

max, maximum; min, minimum.