Equipment and supplies recommended for axillary artery access

Key attributes Example
Ultrasound Small profile probe
Vascular linear array transducer
Minimum 6-cm depth
L25x 6-13 MHz Transducer for Edge (Sonosite)
Micropuncture kit 21-gauge, echogenic needle
0.018ʺ guidewire
Micropuncture Access Kit (Cook)
Sheath for secondary access (femoral) 7F-8F
90-110 cm length preferred for dry closure
Side port for angiography
Braided construction
Hydrophilic coating
Flexor Shuttle Sheath (Cook)
Pinnacle Destination Sheath (Terumo)
Vascular closure device Suture-mediated closure
Pre-close capability, maintain access
Perclose Proglide (Abbott)
Dry closure balloon Compliant balloon
6-12 mm diameter sized 1.0-1.1:1 to vessel
20-40 mm length
Long shaft 120 cm
0.018ʺ or 0.035ʺ wire compatible
7F sheath compatible
Multiple products suitable
Covered stents for bailout Self-expanding covered stents preferred
Oversize by 10-20% relative to vessel
120 cm catheter length
2.5, 5, 7.5 cm lengths
Viabahn Endoprosthesis (W.L. Gore)
5 mm and 6 mm diameter (6F sheath, 0.018ʺ guidewire)
7 mm and 8 mm diameter (7F sheath, 0.018ʺ guidewire)
5 mm and 6 mm diameter (7F sheath, 0.035ʺ guidewire)
7 mm to 10 mm diameter (8F sheath, 0.035ʺ guidewire)