Skip to main content
. 2024 Jul 24;14(15):1590. doi: 10.3390/diagnostics14151590

Table 1.

Activity recovery coefficient as a function of syringe diameter and post-reconstruction Gaussian filters. Recovery is calculated as the ratio between the mean of a central 6 mm diameter VOI and the reference solution.

Diameter of Syringe/FWHM 0 mm 2 mm 4 mm 6 mm
26.5 mm 1.00 1.01 1.01 1.02
19.1 mm 1.01 1.01 1.00 0.97
14.6 mm 0.91 0.90 0.86 0.78
12.0 mm 0.83 0.80 0.73 0.63
8.7 mm 0.63 0.60 0.50 0.41