Skip to main content
. 2024 Jan 9;33(10):2377–2390. doi: 10.1007/s10068-023-01491-y

Table 1.

Summary of eight polycyclic aromatic hydrocarbons (PAH8) and validation results of two different methods (alkali digestion and ultrasound-assisted extraction) with linearity, limit of detection (LOD), limit of quantification (LOQ), and expanded uncertainty

Compounds Abbreviation Quantitative ion (m/z) Qualita-tive ion (m/z) TEFa Alkali digestion extraction method Ultrasound assisted extraction method
R2 LODb (µg/kg) LOQc (µg/kg) Expanded uncertaintyd
(%)
R2 LOD
(µg/kg)
LOQ
(µg/kg)
Expanded uncertainty
(%)
Benz[a]anthracene B[a]A 228 226, 229 0.1 0.9998 0.096 0.292 4.022 0.9991 0.079 0.241 4.463
Chrysene Chry 228 226, 229 0.01 0.9996 0.082 0.248 6.145 0.9992 0.074 0.225 6.196
Benzo[b]fluoranthene B[b]F 252 250, 253 0.1 0.9997 0.091 0.277 5.649 0.9999 0.083 0.252 5.603
Benzo[k]fluoranthene B[k]F 252 250, 253 0.1 0.9999 0.120 0.365 6.485 0.9995 0.105 0.317 7.061
Benzo[a]pyrene B[a]P 252 250, 253 1 0.9994 0.070 0.212 4.511 0.9998 0.115 0.348 4.495
Indeno[1,2,3-cd]pyrene I[c,d]P 276 274, 277 0.1 0.9999 0.129 0.391 4.784 0.9997 0.108 0.327 4.995
Dibenz[a,h]anthracene D[a,h]A 278 276, 279 5 0.9999 0.175 0.531 4.447 0.9998 0.097 0.294 4.615
Benzo[g,h,i]perylene B[g,h,i]P 276 274, 277 0.01 0.9983 0.105 0.318 8.608 0.9984 0.104 0.315 8.529

aToxic equivalency factors (TEFs)

bMethod limit of detection, 3.3σ/s (σ = standard deviation of response; s = slope of regression equation)

cMethod limit of quantification, 10σ/s

dExpanded uncertainties in percentage at 10 μg/kg (confidence level about 95%, k = 2). If nominal concentration level is < 100 μg/kg, the typical expanded uncertainty range should be within 44% (Codex, 2011)