Skip to main content
. 2024 Jul 22;29(3):329–339. doi: 10.5603/rpor.101092

Table 1.

Monte Carlo calculated flatness of the off-axis dose profile for monoenergetic beams of 6, 10, and 12 MeV energy, and for applicators of diameters ranging from 5 to 10 cm with 1 cm steps

Beam energy [MeV] Field diameter [cm]
5 6 7 8 9 10
6 2.7 1.4 0.9 1.7 2.0 1.6
10 2.7 1.7 1.4 2.2 2.4 1.0
12 0.6 1.9 2.7 2.3 1.0 3.0