. 2024 Jul 31;11:1411752. doi: 10.3389/fcvm.2024.1411752

Table 2.

4D measures of the great vessels of healthy controls.

4Dprosp 4Dretro Bias (%) p ICC p
%-Cardiac cycle acquired (range) 85 ± 6
[69–95]
96 ± 1
[93–99]
−11 ± 6 <0.01
Aorta
Thoracic aorta
 Pulse wave velocity (m/s) 5.0 ± 0.8 5.0 ± 0.8 1 ± 11 0.72 0.86 <0.01
Ascending aortic plane
 Total volume (ml) 73 ± 13 73 ± 11 0 ± 9 0.13 0.91 <0.01
 Cardiac output (L/min) 4.5 ± 1.2 4.6 ± 1.1 −2 ± 12 0.10 0.93 <0.01
 Max flow rate (ml/s) 333 ± 86 328 ± 64 1 ± 10 0.72 0.94 <0.01
 Max velocity (cm/s) 98 ± 21 95 ± 16 4 ± 13 0.89 0.88 <0.01
 Max pressure gradient (mmHg) 4.1 ± 1.8 3.7 ± 1.2 10 ± 27 0.69 0.87 <0.01
Main pulmonary artery
Total volume (ml) 75 ± 15 78 ± 14 −4 ± 10 0.02 0.92 <0.01
Cardiac output (L/min) 4.7 ± 1.4 4.9 ± 1.3 −5 ± 12 0.04 0.92 <0.01
Max flow rate (ml/s) 314 ± 86 329 ± 78 −5 ± 11 0.02 0.93 <0.01
Max velocity (cm/s) 80 ± 14 80 ± 12 1 ± 12 0.701 0.81 <0.01
Max pressure gradient (mmHg) 2.6 ± 1.0 2.6 ± 0.7 4 ± 26 0.60 0.78 <0.01

Mean ± SD are shown for vascular measurements of healthy controls (n = 32) from both 4D sequences which were prospectively (4Dprosp) and retrospectively gated (4Dretro). A percent-bias is reported using the 4Dretro sequence as reference. ICC, intraclass correlation.