Table 3.

The intra- and inter-operator reproducibility of loading the Mitra devices with blood and the intra- and inter-day reproducibility of extraction, digestion, and LC-MS/MS measurement.

Protein Peptide Operator reproducibility
Workflow reproducibility
%CV intra %CV inter %CV total %CV intra %CV inter %CV total
n = 6 o = 5 n = 5 d = 5
apoA-I THLAPYSDELR 5.2 8.2 9.7 6.0 0.0 6.0
apoB FPEVDVLTK 3.8 0.0 3.8
GFEPTLEALFGK 8.1 9.7 12.6
apoC-I TPDVSSALDK 4.9 7.8 9.2 6.6 2.9 7.2
apoC-III GWVTDGFSSLK 4.7 6.2 7.8 6.2 5.8 8.5
apoE SELEEQLTPVAEETR 6.5 6.9 9.5 10.4 0.0 10.4
AATVGSLAGQPLQER 4.2 8.0 9.0 6.1 3.2 6.9
LGPLVEQGR 5.7 8.4 10.2 4.6 2.7 5.4
HSA DDNPNLPR 3.6 2.4 4.3
LVNEVTEFAK 4.5 8.8 9.9 3.2 1.0 3.4

Within-operator and within-day%CV are expressed as the average%CV of n replicates over o operators or d days, respectively. The total%CV (%CVintra+inter) is calculated as (CVintra2 + CVinter2)1/2, where CVinter is calculated as the%CV of the individual or daily mean (B) corrected for%CVintra (CVinter = (B2 − CVintra2/n)1/2.