Preoperative and postoperative intraocular pressure and medications

Case Preop IOP (meds) Goal IOP Laser settings POM 1
IOP (meds)
% IOP reduction from preop
POM 3
IOP (meds)
% IOP reduction from preop
POM 6
IOP (meds)
% IOP reduction from preop
POM 12 IOP (meds)
% IOP reduction from preop
POM 18
IOP (meds)
% IOP reduction from preop
POM 24 IOP (meds)
% IOP reduction from preop
1 42 (5) < 21

2000 mW

90 s

21 (4)

50.0%

13 (4)

69.0%

14 (4)

66.7%

47 (5)*

0.0%

2 36 (6) < 18

1500 mW

180 s

36 (6)*

0.0%

3 14 (2) < 12

1500 mW

180 s

20 (2)

0.0%

4 28 (2) < 18

2000 mW

180 s

11 (2)

60.7%

20 (2)

28.6%

14 (2)

50.0%

5 41 (4) < 15

2000 mW

90 s

26 (4)

36.6%

12 (4)

70.7%

12 (4)*

70.7%

9 (4)

78.0%

14 (4)

65.8%

12 (4)

70.7%

6 26 (4) < 15

2000 mW

90 s

9 (4)

65.4%

8 (4)

69.2%

38 (4)

0.0%

8 (4)

69.2%

14 (4)

46.1%

10 (4)

61.5%

7 33 (4) < 21

2000 mW

90 s

8 (0)

75.7%

11 (0)

66.7%

14 (0)

57.6%

10 (0)

69.7%

6 (0)

81.8%

12 (0)

63.6%

IOP intraocular pressure, preop preoperative, POM postoperative month, meds medications, mpCPC micropulse transscleral cyclophotocoagulation

*Underwent additional intervention. Postoperative IOP after additional surgery other than mpCPC was not included in analysis