Table 2.

Rhythm detection performance of AED shock advisory systems during CPR in clinical studies: Comparison of AWC vs. cprINSIGHT algorithms (results are computed from data in original article).

This study De Graaf et al. (2021)19
AED used for data collection DEFIGARD Touch7 DGT7
(SCHILLER Médical)
Lifepack CR2 (STRYKER)
Name of the algorithm (study type) AWC (clinical) cprINSIGHT (clinical)
Strategy of the embedded algorithm To improve the treatment of refibrillation (administer an early shock in shockable patients) To provide continuous CPR at 2-minutes for non-shockable patients
Number of patients 285 365
Number of analysis strips
(validation dataset)
2412 726*
Duration of analysis
during CC (Step 1)
16 s 30 s
Duration of analysis
during reconfirmation (Step 2)
5 s 10 s
Confirmation rate** 27.7%
(557/2011)
29.6%
(215/726)
Performance of analysis during clean ECG: Standard analysis
 VF (Sensitivity) 100% (283/283)
 NSR (Specificity) 100% (16/16) NA
 ONR (Specificity) 96.8% (150/155)
 ASYS (Specificity) 97.6% (82/84)
Performance of regular analysis in presence of chest compressions (Step 1)
 VF (Sensitivity) 92.9% (380/409) 81.8% (90/110)
 NSR (Specificity) NA (0/0) NA (0/0)
 ONR (Specificity) 95.2% (885/930) 78.2% (169/216)
 ASYS (Specificity) 86.5% (463/535) 60.3% (240/398)
Performance of reconfirmation analysis in absence of chest compressions (Step 2)
 VF (Sensitivity) 98.4% (374/380) 100% (15/15)
 NSR (Specificity) NA (0/0) NA (0/0)
 ONR (Specificity) 93.3% (42/45) 97.8% (46/47)
 ASYS (Specificity) 100% (72/72) 97.4% (148/152)
Performance of analyses sequence (Step 1 + Step 2)
 VF (Sensitivity) 91.4% (374/409) 95.4% (105/110)
 NSR (Specificity) NA (0/0) NA (0/0)
 ONR (Specificity) 99.7% (927/930) 99.5% (215/216)
 ASYS (Specificity) 100% (535/535) 97.5% (388/398)
 All rhythms (Accuracy)*** 98.0% (1836/1874) 97.8% (708/724)
Global shock advisory system performance: Standard analysis + AWC (Step 1 + Step 2)
 VF (Sensitivity) 94.9% (657/692)
 NSR (Specificity) 100% (16/16) NA
 ONR (Specificity) 99.3% (1077/1085)
 ASYS (Specificity) 99.7% (617/619)

NA: Not applicable or not reported; VF: Ventricular fibrillation; NSR: Normal sinus rhythm; ONR: Other non-shockable rhythm; ASYS: Asystole.

*

The number of strips represents all cases analyzed in Step 1 plus two pulseless ventricular tachycardias for which performance was not disclosed.

**

Confirmation rate is estimated as the percentage of cases needing a confirmation analysis (Step 2) among all analyses during CC (Step 1).

***

Accuracy related to performance of analyses sequence (Step 1 + Step 2) is the sum of true positives and true negatives divided by the total number of analysis strips.