Table 1.
1st | 2nd | 3rd | Mean | |
---|---|---|---|---|
Scenario 1 | ||||
Below the detection limit. | ||||
Scenario 2 | ||||
150 J | ||||
Maximum voltage across rescuer, V | 97.7 | 60.1 | 61.9 | 73.2 |
Maximum current through rescuer, mA | ||||
Total energy, J | 0.04 | 0.01 | 0.02 | 0.02 |
Proportion of energy*, % | 0.03 | 0.01 | 0.01 | 0.02 |
200 J | ||||
Maximum voltage across rescuer, V | 98.6 | 94.1 | 96.8 | 96.5 |
Maximum current through rescuer, mA | ||||
Total energy, J | 0.04 | 0.04 | 0.03 | 0.04 |
Proportion of energy*, % | 0.02 | 0.02 | 0.02 | 0.02 |
360 J | ||||
Maximum voltage across rescuer, V | 155.9 | 164.0 | 150.5 | 156.8 |
Maximum current through rescuer, mA | ||||
Total energy, J | 0.08 | 0.08 | 0.06 | 0.07 |
Proportion of energy*, % | 0.02 | 0.02 | 0.02 | 0.02 |
Scenario 3 | ||||
150 J | ||||
Maximum voltage across rescuer, V | 222.2 | 258.0 | 232.0 | 237.4 |
Maximum current through rescuer, mA | ||||
Total energy, J | 0.16 | 0.19 | 0.14 | 0.16 |
Proportion of energy*, % | 0.11 | 0.13 | 0.09 | 0.11 |
200 J | ||||
Maximum voltage across rescuer, V | 318.0 | 232.0 | 265.1 | 271.7 |
Maximum current through rescuer, mA | ||||
Total energy, J | 0.32 | 0.19 | 0.24 | 0.25 |
Proportion of energy*, % | 0.16 | 0.10 | 0.12 | 0.13 |
360 J | ||||
Maximum voltage across rescuer, V | 318.9 | 347.5 | 311.7 | 326.0 |
Maximum current through rescuer, mA | ||||
Total energy, J | 0.35 | 0.34 | 0.26 | 0.32 |
Proportion of energy*, % | 0.10 | 0.09 | 0.07 | 0.09 |
Represents the fractional energy delivered to the rescuer arising from the electric potential difference on the rescuer between the two points of contact with the patient and/or common ground, calculated as the energy experienced by the rescuer/the total energy output of the defibrillator.