. 2024 Feb 28;17(3):285–301. doi: 10.2174/0118744710293461240219111852

Table 2.

Summary of the specifications and results for the validation of the radio-HPLC method (RAD detection).

Parameter Criteria Specifications Results
Identity RRT 0.95–1.05 Isomer 1:1.01
Isomer 2:1.006
Linearity R2 value R2 ≥ 0.99 0.9975
Specificity R value R > 1.5 1.99–4.76
Repeatability (%) RRT CV ≤ 2 < 1.8
Accuracy (%) Recovery ≥ 95 95.6 ± 0.6
LOQ (MBq/mL) Radio detection Signal/noise ratio > 10 Isomer 2:8
68Ga3+:0.228
LOD (MBq/mL) Radio detection Signal/noise ratio > 3 Isomer 2:1.05
68Ga3+:0.097
HR test Radio detection No extra signal No extra signal