Table 1. Ultrasonic stimulation parameters.

Exp. f(kHz) depth (mm) PD(ms) PRF(kHz) DC PTD(ms) Isppa(W/cm2) P(Mpa) MItc
I 500 35 0.3 1000 30% 100/500 32.5/65 1.02/1.44 0.99/1.40
II 250 28 0.3 1000 30% 500 6.35/19.06 0.45/0.78 0.65/1.12
III 500 30 0.1 1000 10% 500 9.26 0.54 0.53
IV 250 50* 0.3 1000 30% 400 4.34/8.69/10.52 0.37/0.53/0.58 0.53/0.76/0.83

f = fundamental frequency, depth = TPO focus setting for distance of free-water full-width half-maximum from transducer exit plane, PD = pulse duration, PRF = pulse repetition frequency, DC = duty cycle, PTD = pulse train duration, Isppa = spatial-peak pulse-average intensity in free-water, P = pressure, MItc = transcranial derated mechanical index. The ramp shape for all experiments was rectangular. For estimated intracranial indices for Experiments I & II see Appendix 5—figure 5.

*

Note: Experiments I-III targeted the hand motor area. Experiment IV targeted the corticospinal white matter.