Skip to main content
. 2024 Aug 28;24:382. doi: 10.1186/s12886-024-03635-y

Fig. 1.

Fig. 1

Corneal deformation during the Corvis ST measurement. The whole eye movement (WEM) was measured at a distance of 4 mm from the cornea center. (A) The cornea deforms and moves backward when the jet pulse hits the front surface of the eye. (B) From left to right: Initial state; first corneal applanation (A1); highest concavity (HC); second corneal applanation (A2). (C) Relationships among DeformAmp, DeflAmp, and WEM. Upper left of C: The blue cornea represents the initial moment, the red cornea represents the moment of HC_DeflAmp, and the yellow cornea represents the moment of WEM_Max. Since WEM is still progressing when the cornea is at its maximum deflection, HC_DeflAmp occurs before WEM_Max. Red arrow: DeformAmp; green arrow: WEM; purple arrow: DeflAmp. The red, green, and purple curves in the right panel represent DeflAmp, WEM, and DeflAmp, respectively, with WEM occurring after the complete recovery of the corneal flexure. The right side of C is a schematic of these data