Skip to main content
. 2024 Aug 22;29(16):3975. doi: 10.3390/molecules29163975

Table 3.

Analytical figures of merit of the GF AAS method to quantify manganese in leaf extracts of Dioscorea polystachya.

Parameter Value
Linear working range 2–20 µg L−1
Correlation coefficient (R2) 0.9995
Characteristic mass (m0) 0.67 ± 0.02 pg
Limit of detection (LOD) 0.67 µg L−1
Limit of quantification (LOQ) 1.83 µg L−1
Recovery/Precision (2 µg L−1) 98.5%/1.9%