Table 4.
Parameter | Value |
---|---|
Linear working range | 10–100 µg L−1 |
Correlation coefficient (R2) | 0.9909 |
Characteristic mass (m0) | 21.4 ± 3.4 pg |
Limit of detection (LOD) | 9.94 µg L−1 |
Limit of quantification (LOQ) | 37.64 µg L−1 |
Recovery/Precision (40 µg L−1) | 100.7%/3.8% |