Table 2.
P (W) | Time (min) | Ti (°C) | T1f (°C) | ΔT1 (°C) | T2f (°C) | ΔT2 (°C) |
---|---|---|---|---|---|---|
80 | 3.5 | 20 | 51.3 | 31.3 | 33.4 | 13.4 |
80 | 3.5 | 30 | 62.5 | 32.5 | 39.6 | 9.6 |
80 | 3.5 | 37 | 68.9 | 31.9 | 43.8 | 6.8 |
120 | 5 | 20 | 81.4 | 61.4 | 41.8 | 21.8 |
Ti is the pre-ablation temperature of the bone tissue, T1f is the temperature at sensor T1 at the end of the ablation, ΔT1 is the difference between Ti and T1f, T2f is the temperature at sensor T2 at the end of the ablation, ΔT2 is the difference between Ti and T2f.