Skip to main content
. Author manuscript; available in PMC: 2025 Aug 1.
Published in final edited form as: J Magn Reson. 2024 Jul 26;365:107741. doi: 10.1016/j.jmr.2024.107741

Figure 6. Influence of Undersampling on In-vivo aT2* Mapping.

Figure 6.

Panels (a-d) display axial lung slices (TE = 0.080 ms) sampled at 100%, 70%, 50%, and 10%. At 70% and 50% sampling rates, images closely resemble the fully sampled image, while significantly degraded resolution is observed in the 10% sampled image, particularly near lung parenchyma edges and pulmonary vasculature. Representative parenchymal aT2* maps in panels (e-h) demonstrate regions affected by increasing undersampling. Panels (e-h) exhibit consistent aT2* mean values, while panel (h) displays areas with both increased and reduced aT2* regions, indicated by red arrows and a red-dashed arrow, respectively. (i-l) Histograms of parenchymal aT2* across 10 iterations of undersampling depict distributions at sampling levels of 100%, 70%, 50%, and 10%. Scatter plots (panels m-o) show voxel-level correlations between aT2* values calculated from full sampled data with values calculated at various sampling levels. As sampling level decreases, the correlation slope diminishes, indicating a shift towards higher aT2* values compared to fully sampled data.