n FibroScan iLivTouch Mean of Differences 95% CI for MoD t P r
LSM 231 11.28 ± 11.06 12.91 ± 7.9 −1.63 ± 7.27 −2.57 to −0.69 −3.408 .001 0.754
CAP/UAP 229 258.92 ± 75.27 245.33 ± 45.81 13.59 ± 62.37 5.47-21.71 3.297 .001 0.562
IQR/MED 226 17.76 ± 9.15 15.76 ± 9.78 2.01 ± 12.13 0.42-3.6 2.485 .014 0.180