. 2024 Aug 1;35(8):634–642. doi: 10.5152/tjg.2024.23531

Table 4.

Agreement Analysis in Stiffness Measurements by iLivTouch and FibroScan

Parameter Study 1 Study 2
Intercept Slope Intercept Slope
Coefficient 4.84 0.71 1.58 1.17
Std. error 0.93 0.10 1.29 0.20
95% CI for coefficient 2.99 to 6.69 0.51 to 0.91 −0.96 to 4.14 0.76 to 1.57
Regression equation y = 4.84 + 0.71x y = 1.58 + 1.17x
Correlation coefficient, r 0.75 0.57
95% CI for r 0.69 to 0.80 0.46 to 0.66
CCC 0.70 0.46
95% CI for CCC 0.64 to 0.75 0.36 to 0.54
Descriptive Statistics of Stiffness Measurements of iLivTouch and FibroScan
Variable FIBR LIVT FIBR LIVT
Sample size (n) 231 185
Lowest value (kPa) 2.5 2.9 2.5 2.9
Highest value (kPa) 75 46.3 24.2 26.4
Arithmetic mean (kPa) 11.28 12.91 7.21 10.04
95% CI for the arithmetic mean 9.84 to 12.71 11.88 to 13.93 6.65 to 7.77 9.38 to 10.69
Median (kPa) 7.2 10.3 5.9 8.6
95% CI for the median 6.11 to 8.00 9.11 to 11.98 5.50 to 6.43 7.90 to 9.56
Standard deviation 11.05 7.90 3.86 4.53
Relative standard deviation 0.98 (98.02%) 0.61 (61.21%) 0.53 (53.63%) 0.45 (45.15%)
Variance ratio 1.95 0.72

CCC, concordance correlation coefficient; FIBR, FibroScan; LIVT, iLivTouch.

†Deming rRegression, x: FIBR(kPa) and y: LIVT(kPa)