Agreement Analysis in Controlled Attenuation Parameter (FibroScan) and Ultrasound Attenuation Parameter (iLivTouch) Scores on Steatosis

Parameter Intercept Slope
Coefficient 82.73 0.61
SE 14.78 0.056
95% CI for coefficient 53.61-111.86 0.50-0.72
Regression equation y = 82.73 + 0.61x
Correlation coefficient, r 0.66
95% CI for r 0.58-0.73
Concordance correlation coefficient 0.56
95% CI for CCC 0.49-0.63
Descriptive Statistics of CAP and UAP
Variable CAP UAP
Sample size 224
Lowest value 100 177
Highest value 400 400
Arithmetic mean 262.24 244.13
95% CI for the arithmetic mean 252.67-271.81 238.24-250.02
Median 262.5 236
95% CI for the median -254.54270.45 228.84-246.00
Standard deviation 72.68 44.73
Relative standard deviation 0.27 (27.72%) 0.18 (18.33%)
Variance ratio 2.63

CAP, controlled attenuation parameter; CCC, concordance correlation coefficient; UAP, ultrasound attenuation parameter.

Deming regression; x: CAP and y: UAP.