. 2024 Sep 5;13(9):11. doi: 10.1167/tvst.13.9.11

Table 2.

Performance for Eye Level iRORA and cRORA Classification

OCT Device
ZEISS (n = 143) Spectralis (n = 167)
None versus iRORA/cRORA
 Sensitivity 38.5% 84.0%
 Specificity 93.1% 93.7%
 PPV 35.7% 70.0%
 NPV 93.8% 97.1%
None/iRORA versus cRORA
 Sensitivity 60.0% 62.5%
 Specificity 96.4% 97.4%
 PPV 37.5% 71.4%
 NPV 98.5% 96.1%
None versus iRORA versus cRORA
 κw 0.37 0.73

n is the number of OCT volumes, which is equal to the number of eyes and the number of patients.

NPV, negative predictive value; PPV, positive predictive value.