Table 2.
Catheter migration and early discontinuation in each group.
| Outcome measure | Adhesive control (AC) (n = 61)) | LifeBubble (LB) (n = 57) | P-value | Adjusted P-value |
|---|---|---|---|---|
| Umbilical catheter moved more than one vertebral body, n (%) | 34 (55.7) | 7 (12.3) | <0.001a | <0.001 |
| UVC movement more than one vertebral body, n (%) | 24 (39.3) | 3 (5.3) | <0.001a | 0.001 |
| UVC discontinuation, n (%) | 23 (37.7) | 3 (5.6) | <0.001a | 0.002 |
| UAC movement more than one vertebral body, n (%) | 14 (23.0) | 4 (7.0) | 0.022a | 0.01 |
| UAC discontinuation, n (%) | 4 (6.6) | 0 (0) | – | – |
aBy Fisher’s Exact test, as appropriate.