Table 2.

HS-GC-MS analytical conditions for analysis of impurities in hand sanitizers.

Analytical System
GC- MS GCMS-QP2020 NX
Headspace Autosampler HS−20
Column PoraBOND Q (25 m x 0.25 mm I.D., d.f. = 3 µm)
HS Parameters
Oven Temperature 85°C
Sample Line Temperature 150°C
Transfer Line Temperature 160°C
Injection Time 1 min
Pressurizing Gas Pressure 90 kPa
Equilibrating Time 10 mins
Shaking Level 2
GC Cycle Time 30 mins
GC Parameters
Column temperature 50°C
Injection Mode Split mode
Split ratio 50
Carrier Gas Helium
Gas Flow Condition Constant linear velocity mode
Linear velocity 43.4 cm/s
Oven Temperature Programming 50°C (1 min)
→ 20 °C/min to 250°C (9 mins)
Column Flow 1.19 mL/min
Injection Volume 1.0 µL
MS Parameters
Ion source temperature 230°C
Interface temperature 300°C
Measurement mode Scan/SIM
Scan range m/z 24–200
MSD Solvent Delay No solvent delay
MSD m/z Settings Start time (0 min)
End Time (20 min)
Acquisition mode (SIM/Scan)