Table 2.
HS-GC-MS analytical conditions for analysis of impurities in hand sanitizers.
Analytical System | |
---|---|
GC- MS | GCMS-QP2020 NX |
Headspace Autosampler | HS−20 |
Column | PoraBOND Q (25 m x 0.25 mm I.D., d.f. = 3 µm) |
HS Parameters | |
Oven Temperature | 85°C |
Sample Line Temperature | 150°C |
Transfer Line Temperature | 160°C |
Injection Time | 1 min |
Pressurizing Gas Pressure | 90 kPa |
Equilibrating Time | 10 mins |
Shaking Level | 2 |
GC Cycle Time | 30 mins |
GC Parameters | |
Column temperature | 50°C |
Injection Mode | Split mode |
Split ratio 50 | |
Carrier Gas | Helium |
Gas Flow Condition | Constant linear velocity mode |
Linear velocity 43.4 cm/s | |
Oven Temperature Programming | 50°C (1 min) |
→ 20 °C/min to 250°C (9 mins) | |
Column Flow | 1.19 mL/min |
Injection Volume | 1.0 µL |
MS Parameters | |
Ion source temperature | 230°C |
Interface temperature | 300°C |
Measurement mode | Scan/SIM |
Scan range | m/z 24–200 |
MSD Solvent Delay | No solvent delay |
MSD m/z Settings | Start time (0 min) |
End Time (20 min) | |
Acquisition mode (SIM/Scan) |