A practical guide to light-sheet microscopy for nanoscale imaging: Looking beyond the cell

Stephanie N Kramer 1, Jeanpun Antarasen 1, Cole R Reinholt 1, Lydia Kisley 1,2,1,2,a)
PMCID: PMC11380115  PMID: 39247785

Abstract

We present a comprehensive guide to light-sheet microscopy (LSM) to assist scientists in navigating the practical implementation of this microscopy technique. Emphasizing the applicability of LSM to image both static microscale and nanoscale features, as well as diffusion dynamics, we present the fundamental concepts of microscopy, progressing through beam profile considerations, to image reconstruction. We outline key practical decisions in constructing a home-built system and provide insight into the alignment and calibration processes. We briefly discuss the conditions necessary for constructing a continuous 3D image and introduce our home-built code for data analysis. By providing this guide, we aim to alleviate the challenges associated with designing and constructing LSM systems and offer scientists new to LSM a valuable resource in navigating this complex field.

I. INTRODUCTION

Fluorescence microscopy techniques offer non-invasive means to study biological samples in situ. An ideal fluorescence microscope should possess high sensitivity, signal-to-background ratio (SBR), and spatial resolution, along with fast acquisition speeds and minimal photodamage. A sensitive microscope typically exhibits lower background intensity relative to the sample emission, allowing for the detection of weaker signals amidst background noise. Fluorescence microscopy inherently benefits from high contrast imaging due to the substantial quantum yield of fluorophores, enhancing image quality and enabling precise visualization of low-abundance molecules or events.1–3 High spatial resolution is crucial for resolving fine structural details, particularly at the subcellular level. Fast image acquisition with low source intensity, facilitated by using conventional detectors like fast frame rate cameras, can help mitigate photodamage. Being able to collect at higher frame rates also improves the resolvable timescales for more dynamics-based analysis.

Epifluorescence, also known as wide-field, microscopy is the original fluorescence microscopy technique developed in the early 1900s and is a common method used for biological imaging.4–7 In epifluorescence microscopy, an area of the order of 100 μm2–10 mm2 of the sample is illuminated along the entire axial detection axis. Consequentially, the extensive excitation causes fluorophores within the sample both above and below the focal plane to emit, resulting in out-of-focus signal.8,9 The broad axial excitation with little selectivity leads to limitations such as a lower contrast image, a limited optical sectioning capability, and an increase in photodamage proportional to the number of planes imaged.10–12 However, an advantage to using epifluorescence microscopy is the ability to increase the field of view (FOV) in a given timescale when measuring sample dynamics. Since epifluorescence excites the entire sample at once, it is possible to fully utilize the fast frame rates of the chosen detector without the need to scan across the sample to generate an image.

Laser scanning confocal microscopy (LSCM), first reported in the mid-1900s, was developed to improve upon the image resolution of epifluorescence microscopy and is also popular for biological samples.13–15 Unlike epifluorescence microscopy, which has a focal plane, LSCM has a focal volume, defined by the beam waist of the focused point-like beam and the aperture of a pinhole along the illumination path.4,16 By varying the pinhole size, it is possible to control the background signal and photodamage.17 The incorporation of a pinhole along the detection path blocks out-of-focus light, thus improving the axial resolution, optical sectioning, and image contrast.10,11 However, while the lateral area illuminated in LSCM is significantly smaller (0.1–1 μm2) than in epifluorescence microscopy, the sample is still illuminated along the entire detection axis, risking photodamage. Another limitation of LSCM is the timescale for image acquisition. Since LSCM uses a point-like excitation, the beam must be scanned across the sample, requiring multiple scanning iterations to construct an image. Thus, imaging a single plane is a relatively slow process at approximately 1 frame every second.18 Moreover, since there is a temporal variation between each point in the image, LSCM is not viable for some dynamic samples. Considering the constraints of both epifluorescence and scanning confocal microscopy, including limited 3D sectioning capabilities, potential for inducing sample photodamage, and susceptibility to detecting excess signal beyond the focal plane, it becomes imperative to seek out and adopt an enhanced imaging technique for more precise and comprehensive analysis.

Light-sheet microscopy (LSM) has gained acclamation in biological imaging within the past few decades due to the selectivity of the illuminating focal plane, reducing photodamage, and improving SBR (Fig. 1).2,3,19,20 The selective illumination is due in part to the thin focal volume of the sheet with beam thicknesses reaching as small as 400 nm21–23 in specialized setups such as lattice LSM while more conventional setups afford thicknesses of a few μm.24,25 The ability of LSM to capture entire wide fields of view simultaneously enhances image acquisition speed, making it suitable for dynamic imaging experiments over extended periods such as monitoring zebrafish embryo development,26,27 cell migration,28,29 or neuronal activity30,31 where the timescales range from ms to hours. In addition to these dynamic samples, LSM has also been used to characterize structural features at high resolution within samples ranging from zebrafish32,33 to single cells.34,35 While this is an impressive display of imaging, all listed example analytes have features of the order of μm and, therefore, only require μm resolutions. As such, a 3D reconstruction of most biological samples can be acquired using commercially available LSM setups. Techniques such as flow cytometry36,37 and two-photon excitation38,39 are well suited for diffraction-limited levels of analysis.

FIG. 1.

FIG. 1.

Basic diagram of a light-sheet microscope setup. The excitation source of a LSM setup is generally a laser, the beam profile of which is either Gaussian, Bessel, or Airy in nature (a). Light-sheet formation can be accomplished using cylindrical lenses or via scanning mirrors (b). Most LSMs use a two-objective system with separate illumination and detection objectives (c). The illumination objective focuses a thin sheet through a sample and the detection objective collects the sample emission. The detectors for LSMs are usually either electron multiplying charge-coupled device (EMCCD) or scientific complementary metal–oxide–semiconductor (sCMOS) cameras (d). Inset: The thin sheet allows for a lower background signal as well as less photodegradation of the sample. Sample mounting will be determined based on the objective system chosen as well as the overall geometry of the LSM.

LSMs with super-resolution capabilities40 are necessary for imaging nanoscale features at or below the Abbe diffraction limit41 of visible light ( <250 nm) or dynamic samples at timescales faster than a few ms. Recent advancements in super-resolution LSM include the incorporation of techniques such as single-molecule localization,42,43 structured illumination microscopy,44,45 and correlation-based microscopy.46,47 Generally, to achieve super-resolution, a more specialized system is needed, which will require a custom setup to be constructed. Designing such systems requires careful consideration of optical components and alignment processes. In our guide, we detail considerations for building a home-built LSM for imaging non-traditional systems at nanoscales. We discuss optical component selection, alignment procedures, and our design choices for measuring biomolecule diffusion dynamics within an extracellular matrix (ECM) analog, covering alignment, calibration, data acquisition, and 3D reconstruction processes.

II. OPTICAL HARDWARE COMPONENTS TO CONSIDER PRIOR TO MICROSCOPE CONSTRUCTION

Each light-sheet setup is unique, tailored to both the chosen sample and the desired spatiotemporal resolutions. When beginning to consider the configuration of the microscope, first consider the characteristics of the sample, as they influence subsequent component choices. For instance, if the sample requires a specific orientation, this influences objective selection and the design of the sample chamber. Imaging zebrafish necessitates adherence to a substrate to minimize movement while for smaller entities like cells, immersion in a suitable buffer solution and substrate adherence are essential. Dynamic data collection, such as protein diffusion dynamics, imposes distinct experimental requirements, notably in resolving diffusion timescales (10–100 μm2 s−1), compared to static samples. Regardless of the sample, identifying necessary experimental conditions prior to construction is crucial. In this section, we outline hardware considerations encompassing laser properties, sheet dimensions, objective geometry, sample immersion and orientation, and camera sensitivity and timescales.

A. Beam profiles that have been used in LSM

Three beam profiles— Gaussian,23,30 Bessel,24,48 and Airy49,50—have proven effective in LSM (Fig. 2). Each offers distinct advantages, although no single profile suits every experiment perfectly. The primary considerations for profile selection are the desired axial resolution and lateral FOV. Here, we will explore the characteristics, advantages, and drawbacks of these beam profiles for LSM setups. Notably, we will omit the discussion of less conventional options like lattice profiles, which entail complexities to integration into specialized LSM setups, requiring custom objectives and programmed coordination between electronically-controlled mirrors, objectives, and the sample stage.22,51–53 Commercial lattice LSM systems are available to address such needs.54–56 Among the three discussed profiles, Gaussian profiles are the simplest, with Bessel and Airy profiles typically formed from Gaussian beams. While the intensity profile of a Gaussian beam can be solved analytically in all three dimensions, such analysis is not straightforward for Bessel and Airy profiles despite their derivation from Gaussian beams. Thus, we simplify the analysis by considering two dimensions for Bessel and Airy intensity profiles.

FIG. 2.

FIG. 2.

Profiles of Gaussian [(a) and (d)], Bessel [(b) and (e)], and Airy [(c) and (f)] beams. Here, we show the xy cross-sectional profiles at z=0 [(a)–(c)], (intensity profiles in white) and the axial profiles [(d)–(f)]. The top row illustrates the overall difference in illumination between each beam, while the middle row shows the differences in the confocal parameters. The Gaussian beam is generated from the theoretical model using Eq. (1) with ω0g=4μm. The Bessel beam is generated through a Gaussian beam with ω0g=6μm propagating through an axicon, displaying energy dissipation along the lateral axis. The Airy beam is also simulated using the theoretical model based on the propagation of a Gaussian beam with ω0g=0.1μm through a cubic phase modulator. The confocal parameters for Gaussian ( bg), Bessel ( bb), and Airy ( bAi) are all roughly 200  μm. The scale bar for (a)–(f) is 10  μm. The bottom row depicts images of a juvenile amphioxus, with F-actin stained using Alexa488 phalloidin (green) and propidium iodide (magenta), when excited by Gaussian (g), Bessel (h), and Airy (i) beams to demonstrate the penetration ability of each beam. The data in (g)–(i) are reproduced with permission from Vettenburg et al., Nat. Methods 11, 541–544 (2014). Copyright 2014 Springer Nature.

1. Gaussian

Gaussian beams are common due to their simplicity and the prevalence of lasers with Gaussian or closely approximated two-dimensional intensity profiles [Fig. 2(a)]. The beam profile is described by a Gaussian function

I(r,z)=I02πω0g2(1+(zzRg)2)exp(2r2ω0g2(1+(zzRg)2)), (1)

where I0 is the maximum intensity of the beam at z=0, ω0g the radius of the beam where I=I0/e2 (13.5%), zRg the Rayleigh range, and r the radial distance from the center axis of the beam.57 The beam waist ω0g can be characterized using Eq. (2),

ω0g=λπθ, (2)

where λ is the laser wavelength and θ the divergence angle using the far field approximation.

When opting for a Gaussian beam, achieving both reasonable axial resolution, determined by ω0g, and a wide FOV, based on propagation length, involves a trade-off. This balance is described by the Rayleigh range, defined as the distance from ω0g along the illumination axis where ωz2ω0g. The Rayleigh range is calculated via Eq. (3) with refractive index n, but zRg represents only half of the full propagation length. Therefore, doubling zRg yields the full propagation length [Fig. 2(d)], known as the confocal parameter bg [Eq. (4)],

zRg=πω0g2nλ, (3)
bg=2zRg. (4)

Gaussian beams with narrow waists offer higher axial resolution but have smaller propagation lengths along the illumination axis. While the sectioning will be better in these systems, it will inevitably limit the possible FOV when imaging. Conversely, wider beam waists sacrifice axial resolution for increased propagation lengths and FOVs.24,58 Additionally, Gaussian beams encounter scattering and absorption in biological samples, resulting in irregular patterns like shadows and streaks as light propagates through tissues.59–61 The penetration depth inside tissues is, thus, decreased, making imaging the entire volume more challenging for thicker samples.

2. Bessel

A Bessel beam, unlike a Gaussian beam, exhibits non-diffracting properties, maintaining its waist size without spreading out as it propagates.62–64 To use this beam profile, it must be generated from a Gaussian laser beam, a process achievable through various methods such as employing a diffraction grating65 or using an axicon.24 An axicon is an optic that produces a line focus along the optical axis rather than a point focus.66,67 Various optical elements, such as a narrow annular aperture,68 refractive cone,69 doublet lens,70 conical mirror,71 or circular grating,72 can serve as an axicon. Here, we mathematically demonstrate the generation of a Bessel beam by propagating a Gaussian beam through an axicon, resulting in a finite lateral intensity profile. However, in practical applications, it is not possible to generate infinitely intense non-diffracting beams.73 The intensity profile of a Bessel beam generated from an axicon with refractive index n and an angle τ is described by the function

I(r,z)=4I0ksinαoω0gzbbexp2z2bb2J02(krsinαo), (5)

where αo is the axicon parameter defined as (n1)πτ2, J0 a Bessel function of zeroth order, k=2πλ the wave vector, and r and z the radial and longitudinal coordinates, respectively.

Bessel beams produced from an axicon exhibit multiple concentric rings around the core [Fig. 2(b)], unlike an ideal Gaussian beam with a single mode. These concentric rings maintain consistent intensity within a certain range bb, analogous to the confocal parameter albeit analytical derived differently,

bb=ω0gcosαosinαo. (6)

Here, bb is expressed as a function related to the axicon top angle α0 and the beam waist of an incident Gaussian beam ω0g, assuming the axicon is positioned in air.63 In the case of a Bessel beam generated from a circular aperture, the Rayleigh range and beam waist can be analytically calculated.64

In addition to maintaining axial resolution while propagating over longer distances [Fig. 2(e)], Bessel beams penetrate scattering media more deeply compared to Gaussian beams. Bessel beams possess the ability to “self-heal” in scattering tissue, allowing for increased penetration depth and retention of beam shape.24,58,63,74 This difference in penetration depth was demonstrated by Purnapatra et al. by imaging fluorescent polymer-coated yeast cells suspended in a tissue-like gel.75 From Eq. (6), the confocal parameter of a Bessel beam depends on the beam waist of the Gaussian beam. To achieve the same confocal parameter, the beam waists of the input Gaussian beam are different and based on the chosen optical components. Thus, the penetration of each beam will also differ. In practical use, the Bessel beam penetrates to approximately 650  μm compared to the limit of the Gaussian beam of around 200  μm. However, it is worth noting that the overall image contrast collected with a Bessel beam may be reduced compared to a Gaussian beam due to the multi-lobe nature of a Bessel beam, potentially resulting in out-of-focus excitation.76,77

3. Airy

An Airy beam is non-diffracting and typically generated from a Gaussian profile,78 similar to a Bessel beam. Conventionally, Airy beams are formed using a spatial light modulator (SLM), a device capable of applying a spatially varying phase pattern to light, enabling a more compact Airy-based LSM setup compared to lens-based methods.79,80 However, Airy beams can still be formed using lenses.31,81 Unlike symmetric Gaussian or Bessel profiles, the intensity profile of an Airy beam is parabolic [Fig. 2(f)] and asymmetric in nature [Fig. 2(c)]. While an ideal Airy beam possesses infinite power, experimentally it exhibits finite energy and is best described by an exponentially apertured ideal Airy beam.78 It is then possible to numerically propagate the beam in the z-direction. To create an Airy beam from a Gaussian beam, a cubic phase generated by the SLM is utilized. The intensity profile of the Airy beam can be theoretically expressed as

I(x,y,z)=exp(2a(xx0z22k2x04))exp(2a(yx0z22k2x04))×(Ai(xx0z24k2x04+iazkx02)Ai(yx0z24k2x04+iazkx02))2, (7)

where Ai(x) is the Airy function ( d2xdx2xy=0), x0 the characteristic length, a the apodization factor which describes the amplitude decrease relative to the radial coordinate of the pupil.82

In Airy LSM, the propagation length can be derived based on the full width half maximum (FWHM) of the maximum intensity of the mainlobe of the beam, akin to the definition of bg and bb,

bAi=122ln2ω0g2π2nγλ. (8)

The number of phase wraps at the beam waist γ of a Gaussian beam, which can be used to determine a and x0, can be optimized based on the numerical aperture (NA) of the system or as the ratio of the focal length to the waist of the Gaussian beam ω0g for a specific bAi.82

Similar to Bessel beams, Airy beams maintain comparable axial resolution and exhibit “self-healing,” enabling increased maximum depth of penetration compared to Gaussian beams.58,82–84 Nylk et al. observed that Airy beams could penetrate up to approximately 330  μm into mouse brain tissue, whereas Gaussian beams reached only about 250  μm with the same beam waists of approximately 1.2  μm.85 However, the depth achieved with Airy beams, although significant, is still less than the approximate 650  μm penetration depth reported for Bessel beams by Purnapatra et al.75,85 Additionally, Airy beams may yield lower contrast images due to their multi-lobe intensity structure as well as necessitate the use of deconvolution methods to correct for asymmetric sidelobe structures.49,76,77

B. Sheet dimensions determine the microscope sectioning capabilities

In this section, we take a close look at the important considerations when gathering components for constructing a microscope, ranging from optical parts to the practicalities of assembly. Additionally, we describe some calculations to help estimate the dimensions of the light-sheet based on the planned optical path. These calculations are crucial for optimizing a LSM setup and ensuring it meets the desired imaging needs.

1. Static or dynamic sheets can be formed

Light-sheets are created through the manipulation of Gaussian, Bessel, or Airy beams.3,31,86 As depicted in Fig. 3, cylindrical lenses focus the excitation beam into a light-sheet in selective-plane illumination microscopy (SPIM),19,61 while in digitally scanned laser light-sheet microscopy (DSLM), a virtual light-sheet is formed by rapidly moving a focused beam at the focal plane of the detection lens.87,88

FIG. 3.

FIG. 3.

Ray diagram of representative selective-plane illumination microscopy (SPIM, a) and digitally scanned laser light-sheet microscopy (DSLM, b) imaging systems. In SPIM, a cylindrical lens is employed to illuminate the back focal plane of an infinity-corrected objective, creating a 2D light-sheet within the specimen. In DSLM, a thin Gaussian beam is scanned laterally to form a 2D light-sheet over time.

Cylindrical lenses produce an anisotropic light-sheet. Unlike spherical lenses [Fig. 4(a)], which interact with light symmetrically [Fig. 4(c)], cylindrical lenses [Fig. 4(b)] converge or diverge light asymmetrically depending on orientation [Fig. 4(d)].89,90 One cylindrical lens can generate a triangular sheet along the optical axis, while a pair of either concave or convex lenses oriented orthogonally are needed to produce a diverging rectangular light-sheet.91 In general, SPIM systems are less expensive and simpler to construct than their DSLM counterparts. However, SPIM light-sheets may exhibit intensity variation based on the beam profile and detection objective FOV. Additionally, the detection path may collect some out-of-focus light depending on the sheet thickness.

FIG. 4.

FIG. 4.

Diagram of a convex spherical (a) and a convex cylindrical (b) lens, the names of which stem from the geometric shape the lens would be a slice of. Spherical lenses will focus light symmetrically (c) while cylindrical lenses will focus light asymmetrically (d).

A virtual light-sheet is generated by swiftly scanning an isotropic beam laterally to simulate a planar light-sheet. This method in DSLM requires minimal prior manipulation, resulting in fewer optical aberrations.87 An f θ lens is employed to convert the scan range of the beam into a vertical array of parallel beams.88 The control of the scanning range theoretically enables variation in sheet thickness. Additionally, DSLM ensures uniform illumination across the sample sections, unlike SPIM, but also increases the overall laser power necessary to provide a comparable signal as the beam becomes essentially time-shared.3,87 Moreover, the need for rapid beam movement entails the use of electronic scanning mirrors, escalating system cost and complexity. However, the scanning can be synchronized with a “rolling shutter” exposure mode (Sec. II E), effectively functioning as a digital confocal slit to eliminate out-of-focus scattered light, enhancing image contrast and depth.57

2. Theoretical calculations to determine the sheet thickness

The equations presented here assume a Gaussian beam profile. To calculate the beam diameter of the sheet, Eq. (2) is redefined considering the specific lens and laser beam quality. The redefined equation incorporates additional variables: f is the lens focal length, λ the excitation wavelength, M2 the laser beam quality, and ω(z) a spreading function

2ωz=4fλM2πω(z). (9)

In Eq. (9), the beam diameter is defined as 2ωz rather than 2ω0g to adjust for optics not precisely positioned at the focal point of the preceding component. The spreading of the beam waist is determined using ω(z) to accurately calculate subsequent beam diameters as the optical system progresses,

ω(z)=ω0g1+(zzRg)2. (10)

Parameter z is the distance of the lens from the focal point of the preceding lens. This approach, applicable in traditional microscopy, enables the determination of the excitation laser diameter along the optical path. In DSLM, which generates a symmetric excitation beam akin to traditional microscopy, Eqs. (9) and (10) are utilized once each. However, in a SPIM configuration using cylindrical lenses to produce an asymmetric beam, these equations must be applied separately to ascertain beam diameters along both the x- and y-axes.

C. Objective lenses will dictate the image quality and LSM geometry

The objective lens serves as a crucial element in a LSM, impacting various components and capabilities of the microscope (Fig. S1 in the supplementary material). The magnification of the objective determines the FOV achievable for imaging the sample as well as the resolving power, which is especially important in correlation-based super-resolution imaging where pixel size determines resolvable feature size.92–94 Moreover, magnification influences the focal length of the objectives, with higher magnifications corresponding to shorter focal lengths, as described in Eq. (11),

fobj=ftlM, (11)

where ftl is the focal length of the tube lens and M the magnification of the objective. This relationship bears significance when deciding on an objective orientation to ensure the sample is within the working distance (WD), i.e., the distance from the front lens of the objective to the focal point. For super-resolution imaging, the WD typically ranges from 0.01 to 10 mm,95–98 posing challenges due to the limited space between the sample and the lens, especially with non-standard objective orientations.

The resolving power of the objective is determined by the numerical aperture (NA), which is dependent on the refractive index n of the imaging media and the half-angle θ of the cone of light entering the lens [Eq. (12)],

NA=nsinθ. (12)

A higher NA yields greater lateral ( Rx,y) and axial ( Rz) resolution, allowing for finer feature discrimination, albeit achievable only when imaging through specific media,

Rx,y=λ2NA, (13)
Rz=2λNA2. (14)

Commercial objectives exhibit variations in properties such as magnification and NA, greatly affecting resolution capabilities. For instance, comparing two objectives with identical magnification but differing NA values illustrates significant differences in lateral and axial resolutions. An example comparison between objectives of NA1=1.5 and NA2=0.65 with an emission wavelength of 532 nm, the resolutions would be Rx,y177 nm and Rz473 nm for objective one, and Rx,y409 nm and Rz2.52μm for objective two. In addition, the NA of the objective will affect the point spread function (PSF), which in turn, will impact the image quality. In two-objective LSM systems, discussed in Sec. II C 2, the PSF of the system is the elementwise product of the individual PSFs of the illumination and detection objectives [Eq. (15)],3,99

PSFLSM=PSFIllXPSFDet. (15)

The closer the NAs of the objectives are, the more circular the PSF of the system while dissimilar NAs will result in a more elongated shape.100 Finding a balance between magnification, WD, and NA is crucial in designing an optical setup. Sections II C 1 and II C 2 delve into available immersion media, objective orientations, and strategies for selecting appropriate options that will work well in the desired experimental parameters.

1. Immersion media is dependent on the sample

The main immersion media for light microscopy are oil, air, and water. The application of each media to samples, such as zebrafish, cells, and diffusing proteins, is discussed here considering what conditions will necessitate the use of each immersion medium. The NA of an objective is also influenced by the refractive index of the medium [Eq. (12)]. A comparison of NAs for commercially available objectives with each immersion medium is presented in Table I.

TABLE I.

Comparison of the NA and resolution of commercially available air, oil, and water objectives.

Immersion media NA rangea Resolution rangeb
Oil 0.5–1.5 0.192–0.575
Air 0.04–0.95 0.303–7.19
Water 0.3–1.2 0.240–0.958

Oil, with its high refractive index ( 1.52),101,102 is often used in super-resolution imaging due to its compatibility with high NA objectives. This medium is commonly employed in inverted microscopes with samples mounted on glass with a similar refractive index (1.518).103–105 Oil immersion objectives require direct objective/oil and oil/glass interfaces, limiting usage primarily to one-objective setups,106,107 making oil suitable for all three experimental examples.

Air offers versatility in objective orientation but has the lowest refractive index (1.00), resulting in lower NA objectives and limited application in nanoscale super-resolution tasks, such as sub-cellular imaging or tracking protein diffusion. Samples with larger micrometer to millimeter features, such as zebrafish,108 are well-suited for air objectives. Although not commonly used in LSM for biophysical imaging, given samples are commonly in aqueous environments, air objectives are simpler to implement if super-resolution is not required. If air objectives are used, they are regularly paired with another immersion medium.109,110

Water, while less versatile than air, provides a more adaptable geometry than oil. Proper index matching still necessitates an aqueous medium, but water objectives are not restricted to one-objective setups. With a refractive index of 1.33, water objectives offer a decent range of NAs and are either water immersion or water dipping. Water immersion objectives are designed to be used with a drop of water placed between a glass coverslip and the objective lens. Water dipping objectives do not use a coverslip and instead are meant to have the front lens either dipped or submerged into the sample medium itself. Water objectives can utilize various aqueous media, such as buffer solutions, offering flexibility. Given that both single cells and diffusing proteins typically require a buffer solution, water serves as an ideal medium for these experiments.111,112 Overall, water immersion is a compromise between the versatility of air and the higher resolution capabilities of oil.

2. Objective orientation is dependent on immersion media

In most LSM setups, the objective geometry consists of two separate optical paths, resulting in an uncoupled illumination and detection. Typically, one objective forms the light-sheet while the other collects the emitted light for detection [Figs. 5(a)5(e)]. It is crucial to consider the WD of both objectives to ensure the confocal parameter [Eq. (4)] of the illumination objective aligns with the focal plane of the detection objective. Although using identical objectives facilitates objective housing and WD matching,23,29 objectives with different magnifications, NAs, and/or WDs can still be utilized.87,113 The choice depends partly on sample size and the physical gap between the objectives, which is often limited, requiring both objectives be designed for the same immersion media, such as air/air or water/water. However, some geometries do allow for air/water,114,115 air/oil,116 or oil/water117 mismatches in media. It is also worth noting that two-objective orientations are not typically designed to work with oil immersion media, with some exceptions mentioned later in this section.

FIG. 5.

FIG. 5.

Diagram of LSM objective orientation options. L-selective-plane illumination microscopy (SPIM, a) is the easiest two-objective orientation to implement. V-SPIM has several options for objective orientations such as inverted SPIM (iSPIM, b), open-top SPIM (otSPIM, c), and dual-illuminated iSPIM (diSPIM, d). Reflective LSM (RLSM, e) allows for a more straightforward mounting arrangement. Single-objective LSM can be achieved using single-objective SPIM (soSPIM, f) or highly inclined and laminated optical sheet (HILO, g) orientations. Green and orange arrows indicate the directions of the excitation beam and resulting emission. Note: This figure is not comprehensive of all geometries. See Refs. 122–124 for a more in-depth discussion on objective geometry options.

Some LSMs feature a single optical path where illumination and detection are coupled through a single objective [Figs. 5(f) and 5(g)].98,106,118,119 These one-objective configurations are more conventional and easily integrated into standard inverted microscope bodies. Since there is only one objective, immersion media and WD matching are not concerns, allowing for the use of any immersion media when designing a one-objective LSM system. Less common LSM setups use three or more optical paths,60,61,120,121 but these are beyond the scope of this tutorial. Here, we present one- and two-objective orientations, starting with the more complex two-objective options.

The standard light-sheet configuration, L-SPIM, features orthogonal optical paths, with one objective forming the light-sheet and the other collecting emitted light [Fig. 5(a)]. First reported in the early 1900s, L-SPIM is the oldest two-objective setup.125 Although originally the sheet was formed with just a cylindrical lens in the illumination path,19,99,126 modern setups often include an objective after the cylindrical lens to further focus the light-sheet.127–129 L-SPIM is the simplest geometry and is relatively easy to incorporate into an inverted microscope body, with either the detection or illumination objective being freely mounted above the sample stage. This orientation is suitable for analysis ranging from monitoring the mobility of HP1 α in cell nuclei23 to determining the structure of whole insects.130 Additionally, L-SPIM can accommodate both air and water immersion media, though the latter requires a special sample chamber as discussed in Sec. II D.

Another two-objective configuration, V-SPIM, maintains objective orthogonality with a 45° orientation relative to the sample [Figs. 5(b)5(d)]. Various V-SPIM orientations, such as inverted SPIM (iSPIM), open-top SPIM (otSPIM), and dual-illuminated iSPIM (diSPIM), are primarily suited for air or water immersion media. However, one thing to consider if choosing any of the V-SPIM configurations is the WD of the objectives which should be long enough to ensure the sample does not contact the lenses themselves.

iSPIM and iSPIM-like geometries are frequently reported two-objective orientations, often due to the prevalence of lattice LSM.22,51–53 It may be slightly misleading to those accustomed to traditional microscopy techniques, as the “inversion” in iSPIM refers not to the orientation of the objectives themselves, but rather to the capability of integrating the SPIM method onto an inverted microscope.131 To mount the two objectives onto the microscope, they are suspended above the sample [Fig. 5(b)].131–133 The overall geometry of iSPIM accommodates the use of both air and water as immersion media, with water being most common in super-resolution applications (see discussion in Sec. II C 1). The reduced distance between the objective and the sample can pose challenges, making iSPIM less ideal for super-resolution imaging of larger samples such as adult zebrafish or cleared tissue samples at least 1 cm in size. However, iSPIM inherently avoids issues like media mismatch or off-axis optical aberrations caused by conventional coverslip mounting, as the image is collected from above, eliminating the need to interact with the sample mount.

otSPIM facilitates easy sample loading and manipulation within the objective gap in an iSPIM setup. In otSPIM, the objectives are positioned below the sample mount, creating an “open-top” design [Fig. 5(c)] that is easily accessible for a multitude of samples and sample chambers.134,135 By situating the objectives at a 45° angle beneath the sample stage, they no longer interface directly with the sample. However, this tilted interface geometry presents challenges due to media mismatch. An additional refractive optic, such as a liquid-filled prism,136,137 solid immersion lens,138,139 or solid immersion meniscus lens,140,141 is required in otSPIM to mitigate off-axis optical aberrations caused by media mismatch. Despite the improvements introduced by the refractive optic, the sample mount itself can introduce astigmatism since the optical axis of the detection objective is tilted. To counter this astigmatism, a single cylindrical lens can be inserted into the imaging path before the detector.134,136 Most objectives selected for otSPIM applications are designed for use in either air or water due to the availability of more reasonable WDs in these immersion media. While it is theoretically feasible to use an oil objective when coupled with a custom oil-matching refractive index prism, this option should only be pursued if the WD is physically practical.

diSPIM effectively eliminates artifacts by isotropically exciting and collecting emission. Prior to this, the discussed objective geometries involved uncoupled illumination and detection paths. However, with side-on illumination from a single objective dedicated to sheet formation, shadow stripes may occur in images due to sample absorption of the excitation source.57,142,143 To mitigate these artifacts, a diSPIM configuration, akin to iSPIM or otSPIM but with both objectives serving as both illumination and detection [Fig. 5(d)], can be employed.144–147 Here, a final image is formed by alternating which objective forms the light-sheet and which one captures the image. When both objectives are identical, the system achieves isotropic resolution and diminishes artifacts caused by sample absorption or scattering.145,148 Given the objective geometry similarities with iSPIM and otSPIM, diSPIM setups encounter similar advantages and limitations. Unique to diSPIM, illuminating the sample from both sides can increase photobleaching and also requires an added level of complexity due to the need for dual illumination and detection paths.

Reflective LSM (RLSM) is a two-objective method that does not require the objectives to be orthogonal. Instead, the illumination objective is positioned nearly in-line with the detection objective, albeit with a slight offset [Fig. 5(e)]. Consequently, an additional reflective component, like a prism149,150 or mirror117,151 set at a 45° angle, is necessary after the illumination objective to direct the light-sheet orthogonally through the sample, hence the name reflective LSM. RLSM accommodates objectives designed for different immersion media within the same setup. With its predominantly vertical geometry, RLSM can be seamlessly integrated into a standard inverted microscope, enabling the use of conventional sample mounting techniques. In this configuration, the sample is positioned at the detection objective, which offers the option of employing objectives with higher NA, such as those designed for oil immersion.151–153 However, the incorporation of a small prism or mirror within the narrower WDs of these objectives introduces additional complexity.

One-objective configurations require supplementary optical components to guide the light-sheet through the sample while simultaneously enabling the objective to collect sample emission. Single-objective SPIM (soSPIM) resembles RLSM in using a micro-fabricated mirror98,118,119 or prism,154,155 set at a 45° angle to direct the light-sheet orthogonally through the sample [Fig. 5(f)]. soSPIM offers the advantage of requiring only a single objective, facilitating easier integration of the light-sheet into a conventional inverted microscope. However, only using one objective can be limiting since the additional reflective component requires careful sample mounting to ensure the sample is within the FOV of the objective while also allowing the light-sheet to reach the reflective optic before entering the sample. Custom sample holders with embedded reflective optics are commonly used to ensure orthogonal penetration of the light-sheet into the sample.156–158 While theoretically compatible with all immersion media, soSPIM may still encounter refractive index mismatches depending on the chosen sample holder.

Highly inclined and laminated optical sheet (HILO) microscopy, often implemented on total internal reflection fluorescence (TIRF) microscopes [Fig. 5(g)], utilizes a single-objective configuration. HILO employs an illumination beam that enters the objective just below the critical angle, generating an inclined sheet.159–161 This tilting of light is achieved with a higher NA objective in conjunction with a translation optic, typically already incorporated into a conventional TIRF microscope setup. The excitation beam is translated somewhere between the edge of the back focal plane of the objective (where TIRF occurs) and the center of the objective (where epifluorescence occurs).106,162,163 Its easy integration into existing TIRF microscope setups commonly found in optics labs renders HILO the most widely adopted light-sheet method in many core facilities.164 TIRF can theoretically be achieved with any of the previously discussed immersion media, making HILO also a viable option. However, water and oil are better suited to this geometry due to their higher inherent NAs. Since the sample is mounted above the objective without additional optical components limiting its placement, various mounting options are available. The primary limitation of the HILO method is the shallower penetration depth of the light-sheet, reaching up to approximately 10  μm,165 compared to other LSM options compatible with samples ranging from 10  μm to 1 mm thick.166 Additionally, since HILO requires the beam to be at an angle between the critical angle of TIRF and the 0° angle of epifluorescence, there is variability on the sheet thickness, which also results in the sheet no longer being aligned with the image plane of the objective.

D. Objective orientation determines the compatible sample size and mounting for imaging

LSM requires a level of creativity in how to introduce the sample to the imaging plane, considering both objective orientation and sample size. Minimizing optical aberrations due to media mismatch is also crucial. Figure 6 outlines commonly reported sample mounting techniques. Regardless of the chosen method, it is essential to plan how to acquire a 3D image. Typically, the sample is translated or rotated within the focal plane of the objective(s). Here, we explore a subset of mounting options, focusing on those not tied to specific experiments. Table II includes common immersion media and objective geometries paired with each technique.

FIG. 6.

FIG. 6.

Diagram of sample mounting techniques. Capillary tubes (a) have been used to mount samples such as zebrafish embryos. This sample holder relies on index matching between the capillary tube, hydrogel, and media to decrease optical aberrations and is commonly used in water immersion systems. Cuvettes (b) are similar to capillary tubes but have been used for samples ranging in size from seaweed to cancer cells. These sample holders work well for air and water but will suffer from optical aberrations due to an index mismatch. Coverslips (c) and Petri dishes (d) are common sample mounting options in conventional microscopy. Both work well for smaller samples such as single cells. Depending on the objective geometry, both coverslips and Petri dishes can be imaged from above or below.

TABLE II.

Comparison of the applicable detection objective immersion media and overall objective geometries for common sample mounting techniques.

Sample mounting Immersion media Objective geometry
Capillary Water L-SPIM, RLSM
Cuvette Air, water L-SPIM, RLSM
Coverslip Air, water, oil iSPIM, otSPIM, diSPIM, RLSM, soSPIM, HILO
Petri dish Air, water, oil iSPIM, otSPIM, diSPIM, RLSM, soSPIM, HILO

Embedding techniques are employed for imaging larger samples like zebrafish, involving the placement of the sample within an optically transparent medium encased in a transparent holder [Fig. 6(a)]. In many LSM applications, water serves as the immersion medium, leading to the common use of agarose hydrogel to physically immobilize the sample. Agarose has a high water content ( 95%), resulting in a refractive index (1.335) similar to water (1.33),99,167 facilitating imaging through the hydrogel with minimal aberrations. Capillary tubes are often utilized as sample holders in this mounting technique, offering flexibility in their application. Initially, these tubes were used to shape the agarose into a cylindrical form, with a small plunger extruding the gel to align it within the focal plane of the objective.19,168,169 However, there is a delicate balance in agarose concentration (wt. %) to ensure structural integrity without adversely affecting the sample. Higher concentrations ( 1 wt. %) may exert compression forces on or restrict the movement of living samples, potentially hindering studies focused on long-term dynamics such as embryo development.20,87 A modern alternative involves utilizing thin fluorinated ethylene propylene (FEP) capillary tubes instead of silica glass. FEP shares a refractive index of 1.338 with water and agarose, enabling direct sample imaging through the tube.170–172 With no need for hydrogel extrusion, lower agarose concentrations ( <1 wt. %) can be used, effectively immobilizing the sample without impeding any dynamic studies.170 Additionally, to match the refractive index of water, capillary tubes are often suspended within a water-tight chamber, ranging from custom-made173 to commercially available.174

Four-sided cuvettes offer structural stability benefits akin to capillary tubes [Fig. 6(b)]. Typically, cuvettes are treated similarly to capillary tubes, employing a gelling agent to suspend the sample within the chamber. This method accommodates various sample sizes, from seaweed fragments175 to breast cancer cells.176 Alternatively, cuvettes can serve as pseudomedia baths for samples. Here, cells are cultured on a conventional coverslip, which is affixed to the bottom of the cuvette, followed by filling with the desired media.177 While cuvettes provide a straightforward commercially available mounting option, they have inherent drawbacks. Primarily, although optically clear, four-sided cuvettes are commonly made from materials with higher refractive indices than the sample. Most commercially available cuvettes are crafted from materials like quartz, borosilicate glass, or polystyrene, with refractive indices around 1.459, 1.519, and 1.55, respectively, whereas LSM-imaged samples typically have refractive indices ranging from approximately 1.33 to 1.47.170,178,179 Consequently, this refractive index mismatch leads to spherical aberrations.180 To address this, one might consider using objectives designed for higher refractive index media, such as oil. However, oil immersion objectives, as discussed in Sec. II C 1, have the shortest working distance among immersion media, typically in the range of a few hundred micrometers.95 Given that the thinnest wall of a cuvette is 1 mm, oil immersion objectives would not be suitable in this context.

The simplest mounting techniques, familiar to those knowledgeable of conventional microscopy, include glass coverslips [Fig. 6(c)] and Petri dishes [Fig. 6(d)]. Coverslips, while common, introduce complexity to sample mounting due to objective orientation. When deciding between one- and two-objective orientations, consideration of physical space for the sample is important, with more constraint typically in two-objective geometries (see Sec. II C 2). The physical limitations of a coverslip also impact the size of imageable samples. In standard inverted orientations, samples are imaged from below, requiring the light-sheet to pass through the coverslip, leading to having to select an appropriate coverslip thickness. Typically, a commercially available No. 1.5 coverslip with a thickness range of 160–190  μm is suitable. Opting for an objective geometry that images the sample from above affords more flexibility in sample preparation and immersion media. Since the coverslip is not in the imaging path, refractive index mismatches due to the substrate are less likely, and any glass thickness can be chosen. In such cases, oil immersion is not viable, while both water and air are. When using water, maintaining the objective and sample in the media throughout imaging is necessary, often facilitated by a media bath component, ranging from custom fabrication51 to a commercially available Petri dish.23

Petri dishes have been used in conventional optical microscopy to facilitate the imaging of biological samples under controlled environments. These dishes accommodate various requirements, from cell culturing to serving as a media bath or a combination of both functions.23,25 Similar to glass coverslips, Petri dishes offer flexibility in imaging orientations, supporting sample observation from either above or below. When imaging from below, it is optimal to use glass-bottom dishes equipped with a No. 1.5 coverslip window to match an oil refractive index for improved imaging quality. For objectives that image from above, standard plastic Petri dishes are sufficient. However, it is crucial to ensure that the dish dimensions align with the physical constraints of the objectives and samples. The dish diameter should accommodate the objective(s) without making contact, while the dish depth must accommodate the sample and immersion media. Fortunately, there is a wide range of commercially available Petri dish dimensions that fit the requirements of specific LSM objectives and samples.

E. Cameras are chosen based on the desired sensitivity and timescale of the measurements

Two-dimensional scientific cameras are used to detect the collected emission from the sample. Electron multiplying charge-coupled device (EMCCD) or scientific complementary metal–oxide–semiconductor (sCMOS) cameras are the two options for nanoscale imaging due to their high (up to 95%) quantum efficiencies. Cameras are selected based on the noise, sensitivity, and frame rate capabilities.

EMCCD cameras find extensive application in various low-light imaging scenarios, spanning from biological samples92,181 to single-molecule measurements182,183 to astronomy,184,185 where high sensitivity is crucial for accurate imaging. Here, the definition of “sensitivity” is the ability to distinguish the signal from the noise of the measurement.186 The inherent high sensitivity of EMCCD cameras stems from an on-chip multiplication gain mechanism that enables the detection of single-photon level intensity. However, this gain mechanism introduces additional noise, potentially lowering the overall signal-to-noise ratio (SNR) and image quality.

To enhance sensitivity, EMCCD cameras commonly feature larger pixels, typically around 13  μm. While this increases photosensitivity, it comes at the expense of resolution. The challenge lies in selecting an appropriate magnification objective to ensure that the image pixel size satisfies the Nyquist sampling criterion, which dictates that the pixel size should be at least half the size of the object being resolved.187 In fluorescence microscopy, the “object” size is determined by the diffraction limit of light, approximated by Eq. (16),

d=λ2NA, (16)

where d is the diffraction limit, λ the emission wavelength, and NA the numerical aperture of the detection objective. In systems with an emission wavelength at or above 500 nm, and assuming an NA of 1, the correlating “object” size is at least 250 nm. Therefore, to meet the Nyquist criterion, the pixel size should ideally be close to or less than 100 nm. For an ideal system, the required magnification can be calculated using Eq. (17),

PPM=PI, (17)

where the physical pixel size ( PP) is divided by the objective magnification ( M) to obtain the image pixel size ( PI). For most EMCCD cameras with a PP of 13 μm, achieving a PI of 100 nm needs a magnification of 130 ×. While this magnification may seem high, it is feasible for many applications and can be achieved with a 100 × objective supplemented by additional magnification optics in the detection path. However, the pixel size is proportional to SNR.188,189 Thus, the impact of a smaller pixel should be taken into account as some super-resolution techniques, such as single-molecule localization, rely on higher SNR.190,191

The frame rates achievable for EMCCD cameras are limited due to a number of factors. First, the on-chip multiplication process restricts the speed at which the camera can read and clear the signal.192 Additionally, EMCCDs typically use a “global shutter” exposure. Each pixel will simultaneously collect charge and then transfer that charge to the readout nodes. The overall signal processing in EMCCDs is pixel by pixel, further slowing the image acquisition. However, since this exposure mode results in a single “snap-shot” in time for each frame, there is very little if any spatial distortion. Moreover, it is possible to temporally correlate between different regions of image, which may be required for some super-resolution techniques such as higher-order cross correlation. While imaging the full-chip, these cameras typically operate at 61 fps, but can reach speeds as high as 4000 fps with a smaller FOV as well as using pixel binning.193 Accordingly, EMCCD cameras are well-suited for imaging static or slow-moving samples with low-emission, where resolution below 130 nm is not essential.

sCMOS cameras, a newer technology compared to their EMCCD counterparts, are increasingly employed in biological imaging194,195 and dynamic studies.196,197 While they are less photosensitive than EMCCDs due to the absence of a gain mechanism, sCMOS cameras exhibit lower overall read noise. However, sCMOS cameras do suffer from pixel-dependent noise, which is difficult to correct.198 The primary advantages of sCMOS cameras lie in their higher inherent resolution and fast achievable frame rates. The enhanced resolution is attributed to their smaller average pixel size, approximately 5  μm. Referring back to Eq. (17), and maintaining the same assumptions, a resolution of 100 nm can be easily attained with a 50 × magnification. Considering typical magnifications of 40–100 × are used in many microscopy setups, this is not an unreasonable magnification to achieve without additional optics following the objective.

The faster frame rates achievable with sCMOS cameras are primarily due to both the exposure mode and the clearing mode. The typical exposure mode used in sCMOSs is a “rolling shutter.” The “rolling shutter” mode has an initial parallelized exposure phase in which each column is activated in sequence, originating in the middle of the sensor and progressing toward the edges. After the initial phase, all activated columns are exposed simultaneously. Finally, the camera reaches the readout phase. Unlike EMCCDs, which clear serially, sCMOS cameras clear in the same manner as the initial exposure phase—in parallel—enabling full-chip frame rates exceeding 100 fps.199 However, the temporal nature of the “rolling shutter” introduces some disadvantages. Since the columns are not exposed simultaneously throughout the entire exposure window, there is a greater chance for spatial distortion as well as less reliable temporal correlation between the center and edges in dynamic samples. The distortion is more apparent when objects larger than a few pixels and are moving faster than the frame rate. However, this drawback is less concerning when imaging smaller objects with a faster frame rate that allows for oversampling. These cameras often come equipped with various settings optimized for SNR, speed, or dynamic range. By imaging a smaller FOV, frame rates of up to 10 000 fps can be reached using a speed setting.200 However, this increased speed comes at the expense of sensitivity. The smaller regions that can reliably be correlated in time limit this exposure mode to techniques such as lower-order cross correlation. Consequently, sCMOS cameras are best suited for imaging samples with high emission efficiencies, relatively fast movement typical of dynamic studies, and resolution requirements below 130 nm.

III. ALIGNMENT AND CALIBRATION PROCEDURES FOR LSM

A. Component selection for the home-built LSM: Optics, sample, and acquisition

The upcoming sections on microscope alignment and calibration are intricately linked to the specific LSM system in use. To provide a practical example, we will first outline the setup of our home-built system (Fig. 7, and Fig. S2 and Table S1 in the supplementary material). As previously mentioned in Sec. I, our microscope aims to capture features beyond the cell, particularly monitoring protein diffusion within the ECM.201–204 Consequently, considerations for resolving nanoscale features and dynamics of 10 μm2 s−1 diffusion in situ dictate our choices regarding beam profile, immersion media, objective orientation, sample mounting, and camera selection.

FIG. 7.

FIG. 7.

Simplified beam path of our light-sheet microscope. A linearly polarized 1.14 mm 532 nm Gaussian beam is introduced into the optical path via a fiber optic coupling (FO), compressed symmetrically by a pair of convex spherical lenses (S1,2), stretched along the y axis by a pair of convex cylindrical lenses (C1,3), and then compressed along the x axis by a convex cylindrical lens (C2) to form a sheet. The sheet is directed through an f θ lens (SL) via two galvanometer mirrors (GMs), passed through a tube lens (T1), and enters the back pupil of the illumination objective (IO). The light-sheet excites the fluorophores within a sample and the emission is imaged with a detection objective (DO), tube lens (T2), relay lens (S3), and sCMOS camera. The intensity of the beam is varied using a neutral density filter wheel (ND) along the illumination path. To ensure only the sample emission reaches the camera, a 550 nm long-pass filter (LP) is placed along the detection path. Here, green denotes the illumination path while orange denotes the detection path. For our imaging procedure, the sample is translated along the z axis through the light-sheet via a piezostage mounted on top of an xy-translation stage (TS). For course adjustments along the z axis, we use a lab jack (LJ) custom fit with the sample stage. Inset: The 10 mm sample (light blue circle) is mounted on a custom stainless steel arm suspended within an aqueous media bath. The arrows represent the locations of the objectives relative to the sample.

For excitation in our LSM, we opt for a simple Gaussian beam profile for its ease of implementation. Since we focus on nanoscale imaging rather than deep tissue penetration, complex beam profiles like Bessel or Airy are unnecessary, as we will not be working with larger mm-scale tissue samples. Our sheet formation involves a combination of cylindrical lenses (Fig. 7, C1–3) and a scanning galvanometer mirror system (Fig. 7, GM). A galvanometer mirror consists of a lightweight mirror attached to a coil within a magnetic field. By varying the current through the coil, the mirror can rapidly and precisely scan the laser beam across a surface in one dimension per mirror. This hybrid approach is meant to mitigate shadow formation issues inherent in Gaussian beams. In the context of scattering illumination, one could reduce the shadowing and striping artifact by moving the light-sheet.205 With the millisecond time scale of protein diffusion, we can reduce the striping and shadowing artifact by adjusting the oscillation rate of the galvanometer mirror to match the camera exposure time, which is 2 ms in this Tutorial. Using Eqs. (9) and (10) from Sec. II B 2, we calculate theoretical beam dimensions based on the lenses and beam path of our microscope, resulting in dimensions of 11.23μm×4.24μm. Later, in Sec. III D 4, we will describe our method for determining the experimental beam dimensions and compare them to these theoretical values.

We opt for water as our immersion media due to its refractive index compatibility with most biological environments and their mimics. This choice allows flexibility in objective geometry, as discussed in Sec. II C 2. Considering our intended samples, such as proteins diffusing within an ECM-like matrix, we employ identical 40 × water dipping objectives (refer to Table S1 in the supplementary material) oriented in an iSPIM configuration. However, the limited space between objectives in the iSPIM orientation requires careful sample mounting. For the objectives we chose, the WD is considered relatively long at 3.3 mm. This WD does not afford a very large objective gap ( 12 mm at the base), resulting in the footprint of the sample having to be smaller. Additionally, as depicted in Fig. 5(b), the objective gap is triangular. Therefore, we designed a method (detailed in Methods in the supplementary material) to create a 3 mm high domed sample on a 10 mm diameter glass coverslip, fitting within the narrow objective gap while still reaching the focal plane. The sample is immobilized by covalently bonding the ECM-like matrix to the glass (Methods in SI) to minimize sample drift and blurring during imaging and then mounted on a custom arm within a media bath (Fig. 7 inset and CAD files S1 and S2 in the supplementary material).

To acquire 3D datasets, we move the sample through the light-sheet while collecting emission intensity. Sample movement is achieved by mounting the media bath to a xy-translation stage (Fig. 7, TS) and mounting the sample arm onto a piezostage for vertical translation. With an NA of 0.8 and an emission wavelength of 575 nm, we anticipate sub-micrometer axial resolution (360 nm) and micrometer lateral resolution (1.8  μm) [Eqs. (13) and (14)]. Given our focus on fast dynamics, we have chosen a sCMOS camera over an EMCCD for its higher frame rate capabilities. Additionally, the higher resolution of the sCMOS suits imaging nanoscale structures with our correlation-based super-resolution technique, discussed later in Sec. IV A 2, since the sampling is increased.204 However, if using another super-resolution technique that is more sensitive to the SNR, such as localization, one should consider choosing an EMCCD instead. For further specific component descriptions and dimensions, refer to Table S1 in the supplementary material and CAD files S1 and S2 in the supplementary material.

B. Standards for alignment should be detectable by eye

Aligning a microscope does not require a specific standard until the detection path. It is recommended to start with a highly fluorescent bulk bead or dye solution, typically of the order of μM for dyes and lower for beads, allowing the beam to be visible to the naked eye (Fig. S3 in the supplementary material), aiding in troubleshooting. For our standard, we utilized a 2.0  μm bead solution diluted to 1:10 of the stock which resulted in a 1 pM concentration (Methods in the supplementary material). Moreover, using a bulk sample eliminates the possibility of imaging molecular diffusion. Once the sample emission reaches the camera, a less concentrated solution should be used. This lower concentration offers a clearer understanding of alignment by revealing any aberrations of individual emitters in the camera. To counter potential blurring from diffusion, suspending the particles within a gel with a refractive index similar to that of the solvent can be helpful.

C. Alignment is an iterative process

The alignment process may appear daunting initially, but following a systematic approach can simplify it considerably. In this section, we will walk through our LSM alignment process and provide useful tips for other setups.

To begin alignment, map out the optical mount locations on the breadboard. The spacing between lenses, determined by their functions and focal lengths, should be calculated beforehand (Sec. II B 2). Since this spacing may not align with the inherent spacing of the breadboard, incorporating rails or movable post mounts can be advantageous. Additionally, decide whether optics will be mounted horizontally or vertically. For our LSM, we opted for vertical mounting to accommodate the iSPIM orientation of the objectives, which was achieved using a custom “U”-shaped breadboard (Base Lab Tools, Fig. S2 in the supplementary material).

After mapping the path, attach the optical mounts onto the breadboard. Ensure consistency in the center of each mount component to aid lens alignment later. Although components of the same shape and size are helpful, this is not always feasible, so adjustments may be needed for those with different dimensions. Using post mounts with adjustable heights (e.g., ThorLabs, PH2) and a non-reflective ruler (e.g., ThorLabs, BHM3) can be beneficial in ensuring each component is centered at the same height. However, for vertical mounting, there is potential for misalignment due to gravity and the post is only secured by a single side screw within the mount. The impact of gravity can be mitigated by using posts that have been machined to the desired height and secured directly to the breadboard. During mounting, ensure all mirror mounts are set to 45°, while lens mounts are either at 90° or 180° relative to each other for easy beam path tracing. While some LSM geometries may require deviations, maintaining this rule for most components simplifies alignment with irises.

Once optical mounts are in place, ensure the laser passes through the center of each mount to facilitate overall microscope alignment and prevent edge aberrations. Start with the laser at its lowest power setting, using fluorescent targets (e.g., ThorLabs, VRC2RMS) along the optical path for beam location identification. If this value is greater than 5 mW, neutral density (ND) filters should be used to decrease the overall laser intensity. Protective eye-wear (e.g., ThorLabs, LG12) should also be worn during this process.206 In this first alignment iteration, a multi-iris system with optical post collars (e.g., ThorLabs, R2) to fix the iris height can be used to ensure the laser beam is centered on each mounting component. Mirrors should be used to direct the beam through each iris until it reaches the objective.

In subsequent alignment iterations, we introduce optics into the path gradually to avoid deviations caused by a slight tilt in the optic. If all the lenses are introduced at once, isolating the exact optic(s) causing the deviation will be difficult. In an SPIM setup, a two-iris system can aid alignment until the sheet formation stage at the cylindrical lenses. The two-iris system can generally be used throughout the illumination path of a DSLM since the beam remains circular. For aligning sections with asymmetric beam shapes, a rail system may be necessary. If the beam is aligned prior to entering the first lens, a rail system should allow the beam to maintain its trajectory through the other lenses, which is advantageous when the lens placements do not afford enough room for an iris to be mounted. After the cylindrical lenses, the alignment of the beam can be determined using the fluorescent targets. Using the previous mirror within the optical path, it is possible to direct the beam to the center of the next mirror. This method was used from the cylindrical lenses (Fig. 7, C1–3) to the f θ lens (Fig. 7, SL) and then again from the f θ to the illumination tube lens (Fig. 7, T1). The illumination tube lens and illumination objective (Fig. 7, IO) should be in line with one another. Alignment typically requires multiple attempts and refinements, but once everything seems correct, confirm the beam exits the objective at a 45° angle relative to the sample stage using the highly fluorescent alignment standard (Sec. III B).

The detection path alignment, theoretically simpler than illumination, involves centering the emission collected by the detection objective (Fig. 7, DO) on the detection tube lens (Fig. 7, T2). Use a mirror to direct emission light through a relay lens (Fig. 7, S3), ensuring no warping of the image with the aid of an iris. After the relay lens is a mirror used to direct the light into the camera. Once emission reaches the camera, fine-tune the sheet orientation using a lower concentration of the alignment standard solution, where individual particles can be seen, to better diagnose any aberrations caused by slight misalignments.

D. Calibrations for nanoscale imaging require single particle and molecule samples

Once the microscope is well aligned, calibration is essential to begin data collection. Understanding the FOV based on the beam shape and size, along with the precise pixel size of the setup, is crucial for orienting within the sample. Additionally, it is important to acknowledge that there may be image skewing inherent to 3D imaging. Before reconstructing the structure of the sample, identifying and accounting for this offset is necessary.

1. Standards for calibration should be highly emissive beads of known size

Our standard calibration sample differs from the alignment sample to enable visualization of individual fixed particles for accurate determination of the microscope pixel size and inherent axial skewing in LSM. To accomplish this, we suspended beads in a 2 wt. % agarose hydrogel. Agarose is well-suited for our water dipping objectives due to its refractive index matching, as discussed in Sec. II C 2.99,167 The bead sizes range from 2 to 0.25  μm, facilitating straightforward determination of pixel size. We selected a bead concentration that ensures sufficient density for statistical analysis of pixel size yet avoids overcrowding that would hinder the distinction between individual beads.

2. Pixel size can be measured using beads

To determine the pixel size of our microscope setup, we cannot use conventional methods like a USAF target due to our “non-conventional” geometry.207,208 The flat glass USAF targets are beyond the 3.3 mm WD of the two objectives in our iSPIM setup, and we are unaware of any 3D AF or NIST imaging target compatible with “non-conventional” geometry LSM. Therefore, we have determined the pixel size by measuring the PSF of highly fluorescent beads of varying sizes, specifically 2, 1, and 0.25  μm beads suspended in a 2 wt. % agarose hydrogel (actual sizes reported by the manufacturer are listed in Table III). For this method, we first assume the PI is ideal based on Eq. (17), where PP is 6.5  μm and M is 40 ×, resulting in a PI of 0.163  μm. Using this image pixel size, we estimate the approximate PSF for each selected bead size. Once we estimate the number of pixels that each bead size should encompass, we measure the actual pixel size using our localization software and extract the average pixel size of each bead sample, as detailed in Table III.209 The expected PSF values are statistically within the range of the experimental average PSF values, confirming our alignment in terms of magnification. To determine the actual size of our image pixel for our microscope, we divide the bead size by the average PSF (Table III). The average image pixel size across all bead sizes is 0.18  ± 0.04  μm, which is larger than the Nyquist criterion due to using a 40 × magnification rather than a 60 ×.

TABLE III.

Comparison of the expected and experimental PSFs for each bead size measured.

Bead sizea Expected PSFb Experimental PSFc Pixel sized Overall average pixel sizee
2.10 ± 0.09 12.9 ± 0.6 11 ± 2 (3461) 0.18 ± 0.02 0.18 ± 0.04 (3)
0.950 ± 0.024 5.83 ± 0.15 5.4 ± 0.8 (2140) 0.18 ± 0.03
0.25 ± 0.05 1.5 ± 0.3 1.4 ± 0.1 (2439) 0.18 ± 0.07

aManufacturer reported, μm.

bBased on an ideal pixel size of 160 nm, px.

cStandard deviation based on (x) measurements, px.

dCalculated from the bead size and experimental PSF, μm.

eStandard deviation based on (x) measurements, μm.

3. Axial offsets must be quantified for correct 3D reconstruction

In our iSPIM setup, the excitation and detection paths intersect at right angles, offset from the sample itself, resulting in a parallelogram-shaped sample region [Fig. 8(a)]. Consequently, when imaging in 3D, we perform three separate calibrations to correct for the skewed data in the z, y, and x axes [Figs. 8(b)8(d)].

FIG. 8.

FIG. 8.

Generalized diagram of the light-sheet passing through and exciting a sample (a). Notice how the illumination is not perpendicular to the sample itself, leading to a more complicated geometrical offset of fluorophores during the collection and reconstruction process. The portion highlighted in black depicts the section of the sample within the FOV of the detection objective. This subsection will be used to demonstrate the positions of the fluorophores within the FOV relative to the detection objective as the sample is moved within the z, y, and x axes, respectively. Diagram depicting the expected skewed trajectory of the bead standard along the z axis (b). Here, the sample is vertically within the sheet. The inset demonstrates the expected geometry of the fluorophore offset as it moves along the z axis. Diagram depicting the expected skewed trajectory of the bead standard along the y axis (c). Here, the sample is closer to and further from the detection objective. Diagram depicting how the sample stays within the focal plane when moving along the x axis resulting in no expected offset (d).

As the sample moves through the light-sheet, the same region is detected multiple times in the FOV, irrespective of its orientation relative to the detection objective. This redundancy is accounted for during 3D image reconstruction. Initially, we focus on calibrating the z-offset, which is the most common type of skewing encountered. This is achieved by moving a 2  μm bead calibration standard along the z axis in 5  μm steps using a piezostage [Fig. 8(b)]. However, experimental analysis (Fig. S4a in the supplementary material) reveals a discrepancy between the expected ( 5μm) and observed (2.2  ± 0.1  μm) shifts, indicating that the angle of the light-sheet relative to the sample differs from the assumed 45° (calculations in the supplementary material).

Although the light-sheet angle minimally impacts z- and x-translations, it significantly affects y-translation and the orientation of the captured image on the camera. Through trigonometric calculations, we determine that the actual angle of the light-sheet relative to the sample is 66°, not 45° as initially assumed. The discrepancy between the ideal angle and the actual angle of the light-sheet is due to the sheet entering the back pupil of the objective slightly higher than the optical axis ( 2 mm). However, the off-axis effect can be partially mitigated by incorporating scanning into the image acquisition. Consequently, when not scanning, the image appears tilted on the camera (Fig. S5 in the supplementary material), with shallower emitters appearing closer to the detection objective than deeper ones.

Next, we address the skew along the y axis, which requires a different correction approach due to geometric considerations [Fig. 8(c)]. Using the previously determined angles, we calculate the expected shift along the y axis (calculations in SI in the supplementary material). Trigonometric calculations yield an expected average shift of approximately 4  μm, consistent with the experimental observation of 3.8  ± 0.2  μm (Fig. S4b in the supplementary material).

Finally, we correct for image skew along the x axis, which is straightforward since it is parallel to the image plane [Fig. 8(d)]. Here, we ensure that the physical shift aligns with the pixel shift. Experimental results confirm that moving the sample 5  μm along the x axis corresponds to an average pixel shift of 5.3  ± 0.2  μm (Fig. S4c in the supplementary material), as expected. The quantification of the entire imaging area can then be applied to reconstruct the image. In addition, physical shifts in the excitation volume relative to the sample (i.e., drift) can occur during measurements. Inclusion of fiducial markers such as fluorescent beads,210 luminescent nanoparticles,211 or nanofabricated patterns212 in the experimental sample can quantify drift. Image registration of the marker locations in post-processing can then be applied to accurately reconstruct the three-dimensional image.213 Alternatively, active stabilization to reduce drift during measurements can be achieved with objective nanopositioner or piezostage hardware and programmed feedback loops.214,215

4. Sheet thickness can be measured with a beam profiling camera

Originally, our aim was to determine the sheet thickness by tracking the duration a bead remains in focus as we move the sample along the z axis through the sheet. However, this approach proved impractical due to the geometrical factors discussed in Sec. III D 3, where the initial position of a feature in the fluorescent sample within the sheet affects its observability, rather than the sheet thickness itself. Another indirect measurement option is programming a piezostage to account for the angle of the sheet as the sample is moved along the z axis, which requires prior knowledge of the axial offset of the system (Sec. III D 3). In addition, this method is complex and relatively expensive. Direct measuring options for determining the sheet thickness at the sample plane include rotating the cylindrical lenses 90° or mounting a mirror 45° relative to the objectives. Rotating the optics requires the use of a highly fluorescent sample, such as those discussed in Sec. III B, resulting in imaging the side-profile of the sheet. However, the PSF of the microscope will broaden the measurement and require deconvolution. This method also requires that the optical path be realigned each time the sheet is rotated, introducing the possibility of misalignment. Mounting a mirror will result in the light-sheet appearing as a line on the camera. Here, it is possible to measure the thickness, of a given position, by measuring the FWHM of the line so long as any emission filters within the detection path are removed to allow the laser to reach the sensor. To limit the intensity and avoid saturating the camera, the laser should be set to the lowest power and ND filters should be added to the illumination path during measurement. Additionally, the mirror must be small enough to fit within the objective gap. For our LSM, this would require the diameter of the mirror to be <12 mm, the widest portion of the objective gap. Consequently, we opted to physically measure the beam using a beam profiling camera capable of resolving dimensions down to 20  μm (e.g., Thorlabs, BC207VIS).

To ensure accuracy given the beam size is smaller than the resolution limit of the camera, we measure prior to the objective (Fig. 9) and then calculated the final size using the magnification of the objective. Applying Eq. (9), we determined the experimental beam width (7.1  ± 0.4  μm) and thickness (2.15  ± 0.03  μm). Ideally, the sheet thickness should align closely with the depth of field (DOF) of the detection objective to reduce the detection of as much out-of-focus light as possible. Utilizing Eq. (18),

DOF=λnNA2+nPIMNA, (18)

where λ is the emission wavelength, n the refractive index, NA the numerical aperture of the objective, PI the image pixel size, and M the objective magnification, for our selected 40 ×, NA 0.8 water dipping objective, assuming an emission wavelength of 575 nm, we computed a DOF of 1.20  μm. Given that the DOF is smaller than the beam thickness, some background signal is expected in the images.

FIG. 9.

FIG. 9.

Beam profile of the light-sheet measured by a profiling camera prior to reaching the illumination objective.

The final measured sheet dimensions (7.1 ± 0.4 × 2.15 ± 0.03 μm2) closely approach, but were slightly overestimated by, the theoretical values calculated in Sec. III A (11.23 × 4.24 μm2). However, the sheet dimensions we report here are not completely accurate due to the beam under-filling the back pupil of the objective lens ( 7.2 mm ). In order to take full advantage of the higher NA, the back aperture must be slightly over-filled as the inherent nature of a Gaussian beam dictates only the 1/e2 section of the beam should enter the objective. If the aperture is under-filled, the effective NA of the objective is smaller. Following Eq. (16), a decrease in the NA will also decrease the resolution, with the effect primarily observed in the axial resolution of the system.

IV. DATA COLLECTION AND IMAGE RECONSTRUCTION METHODS FOR LSM

When acquiring data at different depths, the end goal of the imaging determines the axial size of the steps between each dataset, known as “z-slices,” and the acquisition time. These steps should be sufficiently small for continuous imaging, tailored to the size of the feature of interest. Typically, smaller step sizes (e.g., 10–100 nm) are ideal. However, reconstructing samples with larger features (e.g., 1–1000  μm) using fine steps could result in excessively large (e.g., 1 TB) data files. Thus, there must be a balance between the number of steps required for accurate reconstruction and managing file size, which varies depending on the imaging method and information sought.

After data collection, the data must be reconstructed and analyzed. For our LSM data, we have developed a 3D image reconstruction code, available on GitHub and accompanied by a user guide.216 Here, we outline the steps for reconstructing spatial information from a 2  μm bead calibration sample (described in Sec. III D 1) as well as spatiotemporal information collected from 155 kDa dextran diffusing within a 2 wt. % agarose gel.

A. Data collected on the LSM must be converted and corrected

The following discussion will begin with the data workup procedure for static samples. In general, the method for static data analysis is simpler than that of the method for dynamic data. Therefore, it provides a good introduction to the overall analysis process. A flow chart of our data acquisition and analysis steps, as well as representative images of raw data and data prior to deconvolution, are included in the SI (Figs. S6 and S7 in the supplementary material).

1. Static data requires de-skewing and de-blurring corrections of the raw images

In our setup, Micro-Manager is used for data collection, saving raw data files in .tiff format that are subsequently imported into the reconstruction code.217,218 Users have the flexibility to import individual z-slices or a complete sequence comprising all collected z-slices. When importing slices individually, they are consolidated into a single sequence and then converted into a unified .mat file format. Before initiating the code, users input specific frames desired and define the region of interest (ROI).

Upon converting the data to .mat format, we address skew, as discussed in Sec. III D 3, where each subsequent z-slice is offset based on the piezostage position and our calibration data. Our code corrects for skew by shifting each z-slice relative to the location of the first slice in the image sequence. This correction involves translating the image matrix along both x and y axes according to the pixel shift measured for a 5  μm z-step, scaled to account for the actual z-step size between images.

3D deconvolution improves resolution by de-blurring the contribution of the LSM PSF using either an instant deconvolution or an iterative approach. Instant deconvolution, also known as inverse filtering, applies a deconvolution algorithm (e.g., Wiener deconvolution) to the image in order to back-calculate to the de-blurred image. This method yields relatively fast results and is more suitable for real-time applications but may also be less accurate with poorly characterized PSFs.219,220 In contrast, iterative deconvolution assumes the observed image is a convolution of the PSF and real image. Here, the algorithm (e.g., Richardson–Lucy) estimates the de-blurred image, blurs it using the PSF, and then compares the result to the experimentally obtained image. The process is repeated, with an adjusted estimate each iteration, until convergence is reached.221,222 While iterative methods offer higher accuracy for features that are not point-like, they can be computationally intensive, particularly for large or complex images.223,224 In an attempt to alleviate the computational burden, iterative approaches are generally performed within the frequency domain as it is less computationally intensive.225

We adapted an iterative code written originally for depth-resolved holographic reconstruction, leveraging its ability to restore continuous, non-bead-like features.226 The PSF is defined based on the diffraction-limited 2D PSF of the microscope, determined in Sec. III D 2, and extended to a 3D PSF. Using the MATLAB function “fspecial3,” we generate a 3D PSF based on the 2D PSF and matching the object data file dimensions and z-range. The relationship between the 3D PSF and the unknown 3D object guides the iterative loop, with each subsequent iteration approaching closer toward reconstructing the true 3D object dimensions.

Given the sheet thickness (2.15  μm) exceeds the DOF (1.44  μm), some background intensity remains. To enhance the image quality, we background correct via thresholding. Our thresholding process, based on sparse emitters, defines an ROI and calculates local background and standard deviation intensities.227 The code iteratively analyzes the entire image and generates a threshold map by adding three standard deviations to the average background intensity. The corrected frame is generated by subtracting the threshold map from the original frame. This process is performed for each frame in the sequence, ensuring intensity correction for optimal 3D structure reconstruction.

2. Dynamic data can be analyzed by correlation to resolve diffusion and structure in 3D

The dynamic dataset will undergo a similar analysis procedure as the static dataset, with the primary distinction being that instead of containing a single image at each z-step, it now consists of a movie. Prior to de-skewing and deconvolution, the dataset will be transformed into a single image per movie, accomplished through fluorescence correlation spectroscopy super-resolution optical fluctuation imaging (fcsSOFI) analysis.201–204 fcsSOFI allows for simultaneous quantification of molecule diffusion speeds in heterogeneous media while recovering the matrix structure. Current temporal and structural resolutions of fcsSOFI are approximately 1  μm2 s−1 and 100 nm, respectively.203

In our prior publications, we thoroughly explained the theories and methodologies underlying FCS and SOFI correlation analysis techniques.201,202,204 To summarize, the data of intensity against time at individual pixels undergo a second-order correlation. For SOFI spatial analysis, cross correlation is used to reduce aliasing in spatial information.204 Subsequently, an image is generated where each new pixel intensity is determined by the cross correlation function value at the initial time lag. This is unlike conventional optical imaging methods, where reported intensity corresponds to detected photon counts. For FCS diffusion analysis, the correlation curves at each pixel are fit to pre-selected diffusion model like single- or two-component Brownian or anomalous diffusion.202 Fitting the auto-correlation curves yields the diffusion time τD for each pixel, which is then used to calculate the diffusion coefficient D using Eq. (19),

D=ω24τD, (19)

where ω2 is the focal area characterized by the standard deviation of the 2D PSF. A diffusion map is then generated containing D values for each pixel. The fcsSOFI image is then formed in an image fusion step of the SOFI image indicating the saturation and the FCS image indicating the hue for a final super-resolved image of diffusion rates at each 2D image size.

B. Intensity corrected z-slices can be stacked to recreate the 3D structure

The final corrected images are stacked to reconstruct the 3D volumetric image. This process is accomplished in our code using the function “vol3d.”228 Briefly, the “vol3d” function creates a volume render from the inputted 3D z-slice sequence. The resulting 3D image (Fig. 10) is scaled in the xy-plane based on a factor of the number of pixels the intensities cover and the value of PI. The z axis scaling is based on a factor of the number of slices and the physical step between them.

FIG. 10.

FIG. 10.

3D reconstruction with deconvolution of stationary 2  μm beads (a) and dynamic 155 kDa dextran (b) in a 2 wt. % agarose gel generated using our 3D reconstruction code from data collected on our home-built LSM.

Starting with the reconstruction of the static 2  μm bead data [Fig. 10(a)], we see a heterogeneous distribution of the beads within the hydrogel environment. Performing line section analysis on the beads in view (Fig. S8 in the supplementary material) reveals the average diameter is 1.6  ± 0.5  μm (x axis) and 1.7  ± 0.7  μm (y axis), slightly less than the manufacturer reported size (2.10  ± 0.09  μm, Table III), but still within variation. The lower average diameter is most likely caused by the deconvolution step, which is known to over-correct the image. Additionally, there is variability in the bead locations within the light-sheet with beads ranging from partially sectioned to fully illuminated by the sheet. This effect is demonstrated in Fig. 10(a) through the variability in saturation where beads that are partially in the light-sheet are less intense than those that are fully within.

Next, we examine the validity of using our home-built LSM to measure spatiotemporal information from 155 kDa dextran diffusing within a 2 wt. % agarose gel. Here, we rely on the complex porous structure of agarose for our nanoscale imaging. These hydrogels are known to have a wide range of structural features including isolated cavities and interconnected channels with an average size of approximately 150 nm.229,230 When allowed to freely diffuse in water, the diffusion coefficient of 155 kDa dextran should be around 24 μm2 s−1 given the molecule has a hydrodynamic radius of 9 nm.231 Introducing a porous structure should induce a confinement effect, reducing the overall diffusion coefficient of the molecule. Previously, we have shown 2 wt. % agarose slows the diffusion of 155 kDa dextran to 2.8 ± 0.4 μm2 s−1. However, we noted this speed is slower than expected and suspect interactions at the glass interface to be the cause.203 Here, the average diffusion coefficient of our 3D system is 6.4  ± 0.9 μm2 s−1 [Fig. 10(b)], faster than that of our previously reported TIRF results due to the lack of interactions at an interface. We have also previously reported the pore structure of the agarose gel, albeit limited to the glass interface, using our fcsSOFI technique.203 Here, we are able to recover not only isolated cavities but also interconnected channels [Fig. 10(b)]. Performing line section analysis, we find the average cross section is 340  ± 50 nm (Fig. S9 in the supplementary material).

V. CONCLUSION

This paper provides an accessible practical guide to LSM for scientists new to this technique. We emphasize that LSM is suitable not only for imaging static microscale features but also nanoscale features and diffusion dynamics. We cover fundamental aspects of microscope conception, beam profiling, and guide readers through the entire process, concluding with image reconstruction. In Sec. II, we discuss essential practical considerations for designing a home-built LSM system. This includes decisions on beam profiling, static or dynamic sheet formation, objectives, sample mounting, and camera selection. We highlight the importance of objective orientation, as this choice impacts the overall design and adaptability of LSM for various samples and imaging conditions. Following component selection, Sec. III outlines our home-built setup to illustrate the alignment and calibration process. While alignment procedures are similar to other optical microscopes, LSM calibration differs due to its 3D imaging capability and axial skewing during sample movement within the light-sheet. Correcting this offset is crucial for accurate 3D image reconstruction. Given the custom nature of our system, we detail the conditions and steps required for constructing a continuous 3D image using our home-built software in Sec. IV. Each LSM system is unique, presenting challenges in design and construction tailored to specific research needs. We hope this guide serves as a valuable resource, offering clarity and guidance in the complex field of light-sheet microscopy.

SUPPLEMENTARY MATERIAL

Methods including sample preparations and imaging conditions, skew correction calculations, schematic of a typical infinity-corrected objective setup (S1), photograph of our LSM setup (S2), photograph of a highly fluorescent bulk bead standard solution (S3), time-lapse line sections of 2  μm bead (S4), image of 2  μm bead to demonstrate the tilt of the light-sheet (S5), flow chart of data acquisition and analysis (S6), example 3D reconstructed images (S7), line sections of a 2  μm bead after 3D reconstruction (S8), cross section of an agarose pore (S9), SNR vs SBR comparison (S10), graphical representation of the light-sheet geometry (S11), all components used for our LSM setup (Table S1 in the supplementary material), and legends for CAD Files S1–S7 (PDF). CAD S1: Bath Drawing (PDF); CAD S2: Microscope Sample Holder Assembly (PDF); CAD S3: Microscope Stage Mount Knob (STL); CAD S4: Microscope Stage Mount Flat Pattern of Slide Holder (DXF); CAD S5: Microscope Stage Mount Riser (STL); CAD S6: Microscope Stage Mount Base (DXF); CAD S7: Laboratory Jack to Stepper Stage Adapter (PDF).

ACKNOWLEDGMENTS

We acknowledge NIH NIGMS (Grant No. R35GM142466) and the Paul G. Allen Family Foundation’s Allen Distinguished Investigators Award for financial support of this work. The authors also acknowledge the Kisley research group, the Jenkins research group, and Matthew Kramer for helpful discussions.

AUTHOR DECLARATIONS

Conflict of Interest

The authors have no conflicts to disclose.

Author Contributions

Stephanie N. Kramer: Conceptualization (equal); Data curation (lead); Formal analysis (lead); Investigation (lead); Methodology (lead); Software (lead); Writing – original draft (lead); Writing – review & editing (equal). Jeanpun Antarasen: Formal analysis (supporting); Software (supporting); Writing – original draft (supporting); Writing – review & editing (equal). Cole R. Reinholt: Conceptualization (supporting); Visualization (lead); Writing – review & editing (equal). Lydia Kisley: Conceptualization (equal); Funding acquisition (lead); Writing – review & editing (equal).

DATA AVAILABILITY

The data that support the findings of this study are available from the corresponding author upon reasonable request.

REFERENCES

  • 1.Verveer P. J., Swoger J., Pampaloni F., Greger K., Marcello M., and Stelzer E. H. K., “High-resolution three-dimensional imaging of large specimens with light sheet–based microscopy,” Nat. Methods 4, 311–313 (2007). 10.1038/nmeth1017 [DOI] [PubMed] [Google Scholar]
  • 2.Stelzer E. H. K., “Light-sheet fluorescence microscopy for quantitative biology,” Nat. Methods 12, 23–26 (2015). 10.1038/nmeth.3219 [DOI] [PubMed] [Google Scholar]
  • 3.Olarte O. E., Andilla J., Gualda E. J., and Loza-Alvarez P., “Light-sheet microscopy: A tutorial,” Adv. Opt. Photon. 10, 111 (2018). 10.1364/AOP.10.000111 [DOI] [Google Scholar]
  • 4.Renz M., “Fluorescence microscopy—A historical and technical perspective,” Cytometry 83, 767–779 (2013). 10.1002/cyto.a.22295 [DOI] [PubMed] [Google Scholar]
  • 5.Funatsu T., Harada Y., Tokunaga M., Saito K., and Yanagida T., “Imaging of single fluorescent molecules and individual ATP turnovers by single myosin molecules in aqueous solution,” Nature 374, 555–559 (1995). 10.1038/374555a0 [DOI] [PubMed] [Google Scholar]
  • 6.Harms G. S., Cognet L., Lommerse P. H., Blab G. A., and Schmidt T., “Autofluorescent proteins in single-molecule research: Applications to live cell imaging microscopy,” Biophys. J. 80, 2396–2408 (2001). 10.1016/S0006-3495(01)76209-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Pasulka A. L., Howes A. L., Kallet J. G., VanderKelen J., and Villars C., “Visualization of probiotics via epifluorescence microscopy and fluorescence in situ hybridization (FISH),” J. Microbiol. Methods 182, 106151 (2021). 10.1016/j.mimet.2021.106151 [DOI] [PubMed] [Google Scholar]
  • 8.Sieracki M. E., Johnson P. W., and Sieburth J. M., “Detection, enumeration, and sizing of planktonic bacteria by image-analyzed epifluorescence microscopy,” Appl. Environ. Microbiol. 49, 799–810 (1985). 10.1128/aem.49.4.799-810.1985 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Bai C., Qian J., Dang S., Peng T., Min J., Lei M., Dan D., and Yao B., “Full-color optically-sectioned imaging by wide-field microscopy via deep-learning,” Biomed. Opt. Express 11, 2619–2632 (2020). 10.1364/BOE.389852 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Li Y., Dick W. A., and Tuovinen O. H., “Fluorescence microscopy for visualization of soil microorganisms—A review,” Biol. Fertil. Soils 39, 301–311 (2004). 10.1007/s00374-004-0720-z [DOI] [Google Scholar]
  • 11.Thorn K., “A quick guide to light microscopy in cell biology,” MBoC 27, 219–222 (2016). 10.1091/mbc.e15-02-0088 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Icha J., Weber M., Waters J. C., and Norden C., “Phototoxicity in live fluorescence microscopy, and how to avoid it,” Bioessay 39, 1700003 (2017). 10.1002/bies.201700003 [DOI] [PubMed] [Google Scholar]
  • 13.Davidovits P. and Egger M. D., “Scanning laser microscope for biological investigations,” Appl. Opt. 10, 1615–1619 (1971). 10.1364/AO.10.001615 [DOI] [PubMed] [Google Scholar]
  • 14.Shotton D. M., “Confocal scanning optical microscopy and its applications for biological specimens,” J. Cell Sci. 94, 175–206 (1989). 10.1242/jcs.94.2.175 [DOI] [Google Scholar]
  • 15.Bayguinov P. O., Oakley D. M., Shih C.-C., Geanon D. J., Joens M. S., and Fitzpatrick J. A. J., “Modern laser scanning confocal microscopy,” Curr. Protoc. Cytom. 85, e39 (2018). 10.1002/cpcy.39 [DOI] [PubMed] [Google Scholar]
  • 16.Elliott A. D., “Confocal microscopy: Principles and modern practices,” Curr. Protoc. Cytom. 92, e68 (2020). 10.1002/cpcy.68 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Harvath L., “Overview of fluorescence analysis with the confocal microscope,” in Immunocytochemical Methods and Protocols. Methods in Molecular Biology, 2nd ed. (Humana Press, 1999), Chap. 20. [DOI] [PubMed]
  • 18.Jonkman J. and Brown C. M., “Any way you slice it—A comparison of confocal microscopy techniques,” J. Biomol. Tech. 26, 54–65 (2015). 10.7171/jbt.15-2602-003 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Huisken J., Swoger J., Del Bene F., Wittbrodt J., and Stelzer E. H. K., “Optical sectioning deep inside live embryos by selective plane illumination microscopy,” Science 305, 1007–1009 (2004). 10.1126/science.1100035 [DOI] [PubMed] [Google Scholar]
  • 20.Reynaud E. G., Kržič U., Greger K., and Stelzer E. H., “Light sheet-based fluorescence microscopy: More dimensions, more photons, and less photodamage,” HFSP J. 2, 266–275 (2008). 10.2976/1.2974980 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Poola P. K., Afzal M. I., Yoo Y., Kim K. H., and Chung E., “Light sheet microscopy for histopathology applications,” Biomed. Eng. Lett. 9, 279–291 (2019). 10.1007/s13534-019-00122-y [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Malivert M., Harms F., Veilly C., Legrand J., Li Z., Bayer E., Choquet D., and Ducros M., “Active image optimization for lattice light sheet microscopy in thick samples,” Biomed. Opt. Express 13, 6211–6228 (2022). 10.1364/BOE.471757 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Capoulade J., Wachsmuth M., Hufnagel L., and Knop M., “Quantitative fluorescence imaging of protein diffusion and interaction in living cells,” Nat. Biotechnol. 29, 835–839 (2011). 10.1038/nbt.1928 [DOI] [PubMed] [Google Scholar]
  • 24.Takanezawa S., Saitou T., and Imamura T., “Wide field light-sheet microscopy with lens-axicon controlled two-photon Bessel beam illumination,” Nat. Commun. 12, 2979 (2021). 10.1038/s41467-021-23249-y [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Sapoznik E., Chang B.-J., Huh J., Ju R. J., Azarova E. V., Pohlkamp T., Welf E. S., Broadbent D., Carisey A. F., Stehbens S. J., Lee K.-M., Marín A., Hanker A. B., Schmidt J. C., Arteaga C. L., Yang B., Kobayashi Y., Tata P. R., Kruithoff R., Doubrovinski K., Shepherd D. P., Millett-Sikking A., York A. G., Dean K. M., and Fiolka R. P., “A versatile oblique plane microscope for large-scale and high-resolution imaging of subcellular dynamics,” eLife 9, e57681 (2020). 10.7554/eLife.57681 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Trivedi V., Truong T. V., Trinh L. A., Holland D. B., Liebling M., and Fraser S. E., “Dynamic structure and protein expression of the live embryonic heart captured by 2-photon light sheet microscopy and retrospective registration,” Biomed. Opt. Express 6, 2056–2066 (2015). 10.1364/BOE.6.002056 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Bernardello M., Marsal M., Gualda E. J., and Loza-Alvarez P., “Light-sheet fluorescence microscopy for the in vivo study of microtubule dynamics in the zebrafish embryo,” Biomed. Opt. Express 12, 6237–6254 (2021). 10.1364/BOE.438402 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Ritter A. T., Asano Y., Stinchcombe J. C., Dieckmann N. M. G., Chen B.-C., Gawden-Bone C., van Engelenburg S., Legant W., Gao L., Davidson M. W., Betzig E., Lippincott-Schwartz J., and Griffiths G. M., “Actin depletion initiates events leading to granule secretion at the immunological synapse,” Immunity 42, 864–876 (2015). 10.1016/j.immuni.2015.04.013 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Schmid B., Shah G., Scherf N., Weber M., Thierbach K., Campos C. P., Roeder I., Aanstad P., and Huisken J., “High-speed panoramic light-sheet microscopy reveals global endodermal cell dynamics,” Nat. Commun. 4, 2207 (2013). 10.1038/ncomms3207 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Ahrens M. B., Orger M. B., Robson D. N., Li J. M., and Keller P. J., “Whole-brain functional imaging at cellular resolution using light-sheet microscopy,” Nat. Methods 10, 413–420 (2013). 10.1038/nmeth.2434 [DOI] [PubMed] [Google Scholar]
  • 31.Hosny N. A., Seyforth J. A., Spickermann G., Mitchell T. J., Almada P., Chesters R., Mitchell S. J., Chennell G., Vernon A. C., Cho K., Srivastava D. P., Forster R., and Vettenburg T., “Planar airy beam light-sheet for two-photon microscopy,” Biomed. Opt. Express 11, 3927–3935 (2020). 10.1364/BOE.395547 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Yang Z., Mei L., Xia F., Luo Q., Fu L., and Gong H., “Dual-slit confocal light sheet microscopy for in vivo whole-brain imaging of zebrafish,” Biomed. Opt. Express 6, 1797–1811 (2015). 10.1364/BOE.6.001797 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Fei P., Lee J., Packard R. R. S., Sereti K.-I., Xu H., Ma J., Ding Y., Kang H., Chen H., Sung K., Kulkarni R., Ardehali R., Kuo C.-C. J., Xu X., Ho C.-M., and Hsiai T. K., “Cardiac light-sheet fluorescent microscopy for multi-scale and rapid imaging of architecture and function,” Sci. Rep. 6, 22489 (2016). 10.1038/srep22489 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Welf E. S., Driscoll M. K., Dean K. M., Schäfer C., Chu J., Davidson M. W., Lin M. Z., Danuser G., and Fiolka R., “Quantitative multiscale cell imaging in controlled 3D microenvironments,” Dev. Cell 36, 462–475 (2016). 10.1016/j.devcel.2016.01.022 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Wu J., Li J., and Chan R. K. Y., “A light sheet based high throughput 3D-imaging flow cytometer for phytoplankton analysis,” Opt. Express 21, 14474–14480 (2013). 10.1364/OE.21.014474 [DOI] [PubMed] [Google Scholar]
  • 36.Kumar P., Joshi P., Basumatary J., and Mondal P. P., “Light sheet based volume flow cytometry (VFC) for rapid volume reconstruction and parameter estimation on the go,” Sci. Rep. 12, 78 (2022). 10.1038/s41598-021-03902-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Son J., Mandracchia B., Silva Trenkle A. D., Kwong G. A., and Jia S., “Portable light-sheet optofluidic microscopy for 3D fluorescence imaging flow cytometry,” Lab Chip 23, 624–630 (2023). 10.1039/D2LC01024K [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Mahou P., Vermot J., Beaurepaire E., and Supatto W., “Multicolor two-photon light-sheet microscopy,” Nat. Methods 11, 600–601 (2014). 10.1038/nmeth.2963 [DOI] [PubMed] [Google Scholar]
  • 39.Wolf S., Supatto W., Debregeas G., Mahou P., Kruglik S. G., Sintes J.-M., Beaurepaire E., and Candelier R., “Whole-brain functional imaging with two-photon light-sheet microscopy,” Nat. Methods 12, 379–380 (2015). 10.1038/nmeth.3371 [DOI] [PubMed] [Google Scholar]
  • 40.Prakash K., Baddeley D., Eggeling C., Fiolka R., Heintzmann R., Manley S., Radenovic A., Smith C., Shroff H., and Schermelleh L., “Resolution in super-resolution microscopy—Definition, trade-offs and perspectives,” Nat. Rev. Mol. Cell Biol. 25, 677–682 (2024). 10.1038/s41580-024-00755-7 [DOI] [PubMed] [Google Scholar]
  • 41.Abbe E., “Beiträge zur theorie des mikroskops und der mikroskopischen wahrnehmung,” Archiv f. mikrosk. Anatomie 9, 413–468 (1873). 10.1007/BF02956173 [DOI] [Google Scholar]
  • 42.Cella Zanacchi F., Lavagnino Z., Perrone Donnorso M., Del Bue A., Furia L., Faretta M., and Diaspro A., “Live-cell 3D super-resolution imaging in thick biological samples,” Nat. Methods 8, 1047–1050 (2011). 10.1038/nmeth.1744 [DOI] [PubMed] [Google Scholar]
  • 43.Cella Zanacchi F., Lavagnino Z., Faretta M., Furia L., and Diaspro A., “Light-sheet confined super-resolution using two-photon photoactivation,” PLoS One 8, e67667 (2013). 10.1371/journal.pone.0067667 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44.Breuninger T., Greger K., and Stelzer E. H. K., “Lateral modulation boosts image quality in single plane illumination fluorescence microscopy,” Opt. Lett. 32, 1938–1940 (2007). 10.1364/OL.32.001938 [DOI] [PubMed] [Google Scholar]
  • 45.Keller P. J., Schmidt A. D., Santella A., Khairy K., Bao Z., Wittbrodt J., and Stelzer E. H. K., “Fast, high-contrast imaging of animal development with scanned light sheet–based structured-illumination microscopy,” Nat. Methods 7, 637–642 (2010). 10.1038/nmeth.1476 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.Chen X., Zong W., Li R., Zeng Z., Zhao J., Xi P., Chenc L., and Sun Y., “Two-photon light-sheet nanoscopy by fluorescence fluctuation correlation analysis,” Nanoscale 8, 9982–9987 (2016). 10.1039/C6NR00324A [DOI] [PubMed] [Google Scholar]
  • 47.Chen R., Zhao Y., Li M., Wang Y., Zhang L., and Fei P., “Efficient super-resolution volumetric imaging by radial fluctuation Bayesian analysis light-sheet microscopy,” J. Biophotonics 13, e201960242 (2020). 10.1002/jbio.201960242 [DOI] [PubMed] [Google Scholar]
  • 48.Planchon T. A., Gao L., Milkie D. E., Davidson M. W., Galbraith J. A., Galbraith C. G., and Betzig E., “Rapid three-dimensional isotropic imaging of living cells using Bessel beam plane illumination,” Nat. Methods 8, 417–423 (2011). 10.1038/nmeth.1586 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Vettenburg T., Dalgarno H. I. C., Nylk J., Coll-Lladó C., Ferrier D. E. K., Čižmár T., Gunn-Moore F. J., and Dholakia K., “Light-sheet microscopy using an airy beam,” Nat. Methods 11, 541–544 (2014). 10.1038/nmeth.2922 [DOI] [PubMed] [Google Scholar]
  • 50.Corsetti S., Wijesinghe P., Poulton P. B., Sakata S., Vyas K., Herrington C. S., Nylk J., Gasparoli F., and Dholakia K., “Widefield light sheet microscopy using an airy beam combined with deep-learning super-resolution,” OSA Contin. 3, 1068–1083 (2020). 10.1364/OSAC.391644 [DOI] [Google Scholar]
  • 51.Chen B.-C., Legant W. R., Wang K., Shao L., Milkie D. E., Davidson M. W., Janetopoulos C., Wu X. S., Hammer J. A., Liu Z., English B. P., Mimori-Kiyosue Y., Romero D. P., Ritter A. T., Lippincott-Schwartz J., Fritz-Laylin L., Mullins R. D., Mitchell D. M., Bembenek J. N., Reymann A.-C., Böhme R., Grill S. W., Wang J. T., Seydoux G., Tulu U. S., Kiehart D. P., and Betzig E., “Lattice light-sheet microscopy: Imaging molecules to embryos at high spatiotemporal resolution,” Science 346, 1257998 (2014). 10.1126/science.1257998 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52.Shi Y., Tabet J. S., Milkie D. E., Daugird T. A., Yang C. Q., Ritter A. T., Giovannucci A., and Legant W. R., “Smart lattice light-sheet microscopy for imaging rare and complex cellular events,” Nat. Methods 21, 301–310 (2024). 10.1038/s41592-023-02126-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 53.Moore R. P., O’Shaughnessy E. C., Shi Y., Nogueira A. T., Heath K. M., Hahn K. M., and Legant W. R., “A multi-functional microfluidic device compatible with widefield and light sheet microscopy,” Lab Chip 22, 136–147 (2021). 10.1039/D1LC00600B [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 54.3i, see https://www.intelligent-imaging.com/lattice for information about “Lattice LightSheet: A microscope for high-resolution, fast, and gentle 3D live cell imaging.”
  • 55.ZEISS, see https://www.zeiss.com/microscopy/en/products/light-microscopes/super-resolution-microscopes/elyra-7 for information about “Elyra 7: Super-resolution and single molecule localization microscope.”
  • 56.ZEISS, see https://www.zeiss.com/microscopy/en/products/light-microscopes/light-sheet-microscopes/lattice-lightsheet-7 for information about “Lattice lightsheet 7: Long-term volumetric imaging of living cells.”
  • 57.Glaser A. K., Chen Y., Yin C., Wei L., Barner L. A., Reder N. P., and Liu J. T. C., “Multidirectional digital scanned light-sheet microscopy enables uniform fluorescence excitation and contrast-enhanced imaging,” Sci. Rep. 8, 13878 (2018). 10.1038/s41598-018-32367-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 58.Corsetti S., Gunn-Moore F., and Dholakia K., “Light sheet fluorescence microscopy for neuroscience,” J. Neurosci. Methods 319, 16–27 (2019). 10.1016/j.jneumeth.2018.07.011 [DOI] [PubMed] [Google Scholar]
  • 59.Rohrbach A., “Artifacts resulting from imaging in scattering media: A theoretical prediction,” Opt. Lett. 34, 3041–3043 (2009). 10.1364/OL.34.003041 [DOI] [PubMed] [Google Scholar]
  • 60.Royer L. A., Lemon W. C., Chhetri R. K., and Keller P. J., “A practical guide to adaptive light-sheet microscopy,” Nat. Protoc. 13, 2462–2500 (2018). 10.1038/s41596-018-0043-4 [DOI] [PubMed] [Google Scholar]
  • 61.Huisken J. and Stainier D. Y. R., “Even fluorescence excitation by multidirectional selective plane illumination microscopy (mSPIM),” Opt. Lett. 32, 2608–2610 (2007). 10.1364/OL.32.002608 [DOI] [PubMed] [Google Scholar]
  • 62.McGloin D. and Dholakia K., “Bessel beams: Diffraction in a new light,” Contemp. Phys. 46, 15–28 (2005). 10.1080/0010751042000275259 [DOI] [Google Scholar]
  • 63.Khonina S. N., Kazanskiy N. L., Karpeev S. V., and Butt M. A., “Bessel beam: Significance and applications—A progressive review,” Micromachines 11, 997 (2020). 10.3390/mi11110997 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 64.Tumkur T. U., Voisin T., Shi R., Depond P. J., Roehling T. T., Wu S., Crumb M. F., Roehling J. D., Guss G., Khairallah S. A., and Matthews M. J., “Nondiffractive beam shaping for enhanced optothermal control in metal additive manufacturing,” Sci. Adv. 7, eabg9358 (2021). 10.1126/sciadv.abg9358 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 65.Jiménez N., Romero-García V., Picó R., Cebrecos A., Sánchez-Morcillo V. J., Garcia-Raffi L. M., Sánchez-Pérez J. V., and Staliunas K., “Acoustic Bessel-like beam formation by an axisymmetric grating,” Europhys. Lett. 106, 24005 (2014). 10.1209/0295-5075/106/24005 [DOI] [Google Scholar]
  • 66.Sochacki J., Kołodziejczyk A., Jaroszewicz Z., and Bará S., “Nonparaxial design of generalized axicons,” Appl. Opt. 31, 5326–5330 (1992). 10.1364/AO.31.005326 [DOI] [PubMed] [Google Scholar]
  • 67.Herman R. M. and Wiggins T. A., “Production and uses of diffractionless beams,” J. Opt. Soc. Am. A 8, 932–942 (1991). 10.1364/JOSAA.8.000932 [DOI] [Google Scholar]
  • 68.Durnin J., Miceli J. J., and Eberly J. H., “Diffraction-free beams,” Phys. Rev. Lett. 58, 1499–1501 (1987). 10.1103/PhysRevLett.58.1499 [DOI] [PubMed] [Google Scholar]
  • 69.McLeod J. H., “The axicon: A new type of optical element,” J. Opt. Soc. Am. 44, 592–597 (1954). 10.1364/JOSA.44.000592 [DOI] [Google Scholar]
  • 70.Burvall A., Kołacz K., Jaroszewicz Z., and Friberg A. T., “Simple lens axicon,” Appl. Opt. 43, 4838–4844 (2004). 10.1364/AO.43.004838 [DOI] [PubMed] [Google Scholar]
  • 71.Fujiwara S., “Optical properties of conic surfaces. I. Reflecting cone,” J. Opt. Soc. Am. 52, 287–292 (1962). 10.1364/JOSA.52.000287 [DOI] [Google Scholar]
  • 72.Pérez M. V., Gómez-Reino C., and Cuadrado J. M., “Diffraction patterns and zone plates produced by thin linear axicons,” Opt. Acta: Int. J. Opt. 33, 1161–1176 (1986). 10.1080/713822071 [DOI] [Google Scholar]
  • 73.Brzobohatý O., Čižmár T., and Zemánek P., “High quality quasi-Bessel beam generated by round-tip axicon,” Opt. Express 16, 12688–12700 (2008). 10.1364/OE.16.012688 [DOI] [PubMed] [Google Scholar]
  • 74.Fahrbach F. O., Simon P., and Rohrbach A., “Microscopy with self-reconstructing beams,” Nat. Photon. 4, 780–785 (2010). 10.1038/nphoton.2010.204 [DOI] [Google Scholar]
  • 75.Purnapatra S. B., Bera S., and Mondal P. P., “Spatial filter based Bessel-like beam for improved penetration depth imaging in fluorescence microscopy,” Sci. Rep. 2, 692 (2012). 10.1038/srep00692 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 76.Deng S., Yuan L., Cheng P., Wang Y., and Liu M., “Enhancement of image quality in planar airy light-sheet microscopy via subtraction method,” J. Opt. 24, 025301 (2022). 10.1088/2040-8986/ac4437 [DOI] [Google Scholar]
  • 77.Remacha E., Friedrich L., Vermot J., and Fahrbach F. O., “How to define and optimize axial resolution in light-sheet microscopy: A simulation-based approach,” Biomed. Opt. Express 11, 8–26 (2020). 10.1364/BOE.11.000008 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 78.Morris J. E., Mazilu M., Baumgartl J., Čižmár T., and Dholakia K., “Propagation characteristics of airy beams: Dependence upon spatial coherence and wavelength,” Opt. Express 17, 13236–13245 (2009). 10.1364/OE.17.013236 [DOI] [PubMed] [Google Scholar]
  • 79.Latychevskaia T., Schachtler D., and Fink H.-W., “Creating airy beams employing a transmissive spatial light modulator,” Appl. Opt. 55, 6095–6101 (2016). 10.1364/AO.55.006095 [DOI] [PubMed] [Google Scholar]
  • 80.Ren Y.-X., He H., Tang H., and Wong K. K. Y., “Non-diffracting light wave: Fundamentals and biomedical applications,” Front. Phys. 9, 698343 (2021). 10.3389/fphy.2021.698343 [DOI] [Google Scholar]
  • 81.Papazoglou D. G., Suntsov S., Abdollahpour D., and Tzortzakis S., “Tunable intense airy beams and tailored femtosecond laser filaments,” Phys. Rev. A 81, 061807 (2010). 10.1103/PhysRevA.81.061807 [DOI] [Google Scholar]
  • 82.Taege Y., Borre A. L., Veettikazhy M., Schulz S. L., Marti D., Andersen P. E., Messerschmidt B., and Ataman C., “Design parameters for airy beams in light-sheet microscopy,” Appl. Opt. 61, 5315–5319 (2022). 10.1364/AO.458343 [DOI] [PubMed] [Google Scholar]
  • 83.Mazilu M., Stevenson D., Gunn-Moore F., and Dholakia K., “Light beats the spread: “Non-diffracting” beams,” Laser Photonics Rev. 4, 529–547 (2010). 10.1002/lpor.200910019 [DOI] [Google Scholar]
  • 84.Broky J., Siviloglou G. A., Dogariu A., and Christodoulides D. N., “Self-healing properties of optical airy beams,” Opt. Express 16, 12880–12891 (2008). 10.1364/OE.16.012880 [DOI] [PubMed] [Google Scholar]
  • 85.Nylk J., McCluskey K., Aggarwal S., Tello J. A., and Dholakia K., “Enhancement of image quality and imaging depth with airy light-sheet microscopy in cleared and non-cleared neural tissue,” Biomed. Opt. Express 7, 4021–4033 (2016). 10.1364/BOE.7.004021 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 86.Luna-Palacios Y. Y., Licea-Rodriguez J., Camacho-Lopez M. D., Teichert I., Riquelme M., and Rocha-Mendoza I., “Multicolor light-sheet microscopy for a large field of view imaging: A comparative study between Bessel and Gaussian light-sheets configurations,” J. BioPhoton. 15, e202100359 (2022). 10.1002/jbio.202100359 [DOI] [PubMed] [Google Scholar]
  • 87.Keller P. J., Schmidt A. D., Wittbrodt J., and Stelzer E. H., “Reconstruction of zebrafish early embryonic development by scanned light sheet microscopy,” Science 322, 1065–1069 (2008). 10.1126/science.1162493 [DOI] [PubMed] [Google Scholar]
  • 88.Keller P. J. and Stelzer E. H. K., “Digital scanned laser light sheet fluorescence microscopy,” Cold Spring Harb. Protoc. 2010, pdb.top78 (2010). 10.1101/pdb.top78 [DOI] [PubMed] [Google Scholar]
  • 89.Goldsmith P., “Gaussian beam transformation with cylindrical lenses,” IEEE Trans. Antennas Propag. 34, 603–607 (1986). 10.1109/TAP.1986.1143861 [DOI] [Google Scholar]
  • 90.Shi G.-Y., Li S., Huang K., Yi H., and Yang J.-L., “Gaussian beam shaping based on aspheric cylindrical lenses,” Optoelectron. Lett. 10, 439–442 (2014). 10.1007/s11801-014-4169-5 [DOI] [Google Scholar]
  • 91.Adams D., “Cylinder lenses for beam shaping,” Laser Tech. J. 15, 26–28 (2018). 10.1002/latj.201800002 [DOI] [Google Scholar]
  • 92.Chizhik A. M., Stein S., Dekaliuk M. O., Battle C., Li W., Huss A., Platen M., Schaap I. A. T., Gregor I., Demchenko A. P., Schmidt C. F., Enderlein J., and Chizhik A. I., “Super-resolution optical fluctuation bio-imaging with dual-color carbon nanodots,” Nano Lett. 16, 237–242 (2016). 10.1021/acs.nanolett.5b03609 [DOI] [PubMed] [Google Scholar]
  • 93.Geissbuehler S., Dellagiacoma C., and Lasser T., “Comparison between SOFI and storm,” Biomed. Opt. Express 2, 408–420 (2011). 10.1364/BOE.2.000408 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 94.Dertinger T., Colyer R., Vogel R., Enderlein J., and Weiss S., “Achieving increased resolution and more pixels with superresolution optical fluctuation imaging (SOFI),” Opt. Express 18, 18875–18885 (2010). 10.1364/OE.18.018875 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 95.Choi C., Song K.-D., Kang S., Park J.-S., and Choi W., “Optical imaging featuring both long working distance and high spatial resolution by correcting the aberration of a large aperture lens,” Sci. Rep. 8, 9165 (2018). 10.1038/s41598-018-27289-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 96.McConnell G., Trägårdh J., Amor R., Dempster J., Reid E., and Amos W. B., “A novel optical microscope for imaging large embryos and tissue volumes with sub-cellular resolution throughout,” eLife 5, e18659 (2016). 10.7554/eLife.18659 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 97.Birk U., Hase J. V., and Cremer C., “Super-resolution microscopy with very large working distance by means of distributed aperture illumination,” Sci. Rep. 7, 3685 (2017). 10.1038/s41598-017-03743-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 98.Galland R., Grenci G., Aravind A., Viasnoff V., Studer V., and Sibarita J.-B., “3D high- and super-resolution imaging using single-objective SPIM,” Nat. Methods 12, 641–644 (2015). 10.1038/nmeth.3402 [DOI] [PubMed] [Google Scholar]
  • 99.Engelbrecht C. J. and Stelzer E. H., “Resolution enhancement in a light-sheet-based microscope (SPIM),” Opt. Lett. 31, 1477–1479 (2006). 10.1364/OL.31.001477 [DOI] [PubMed] [Google Scholar]
  • 100.Becker K., Saghafi S., Pende M., Sabdyusheva-Litschauer I., Hahn C. M., Foroughipour M., Jährling N., and Dodt H.-U., “Deconvolution of light sheet microscopy recordings,” Sci. Rep. 9, 17625 (2019). 10.1038/s41598-019-53875-y [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 101.Diaspro A., Federici F., and Robello M., “Influence of refractive-index mismatch in high-resolution three-dimensional confocal microscopy,” Appl. Opt. 41, 685–690 (2002). 10.1364/AO.41.000685 [DOI] [PubMed] [Google Scholar]
  • 102.Hell S., Reiner G., Cremer C., and Stelzer E. H. K., “Aberrations in confocal fluorescence microscopy induced by mismatches in refractive index,” J. Microsc. 169, 391–405 (1993). 10.1111/j.1365-2818.1993.tb03315.x [DOI] [Google Scholar]
  • 103.Gibson S. F. and Lanni F., “Experimental test of an analytical model of aberration in an oil-immersion objective lens used in three-dimensional light microscopy,” J. Opt. Soc. Am. A 8, 1601–1613 (1991). 10.1364/JOSAA.8.001601 [DOI] [PubMed] [Google Scholar]
  • 104.Fouquet C., Gilles J.-F., Heck N., Santos M. D., Schwartzmann R., Cannaya V., Morel M.-P., Davidson R. S., Trembleau A., and Bolte S., “Improving axial resolution in confocal microscopy with new high refractive index mounting media,” PLoS One 10, e0121096 (2015). 10.1371/journal.pone.0121096 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 105.Besseling T., Jose J., and Blaaderen A. V., “Methods to calibrate and scale axial distances in confocal microscopy as a function of refractive index,” J. Microsc. 257, 142–150 (2015). 10.1111/jmi.12194 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 106.Monge Neria R. and Kisley L., “Single-molecule imaging in commercial stationary phase particles using highly inclined and laminated optical sheet microscopy,” Anal. Chem. 95, 2245–2252 (2023). 10.1021/acs.analchem.2c03753 [DOI] [PubMed] [Google Scholar]
  • 107.Theer P., Dragneva D., and Knop M., “ πSPIM: High NA high resolution isotropic light-sheet imaging in cell culture dishes,” Sci. Rep. 6, 32880 (2016). 10.1038/srep32880 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 108.Castranova D., Samasa B., Venero Galanternik M., Gore A. V., Goldstein A. E., Park J. S., and Weinstein B. M., “Long-term imaging of living adult zebrafish,” Development 149, dev199667 (2022). 10.1242/dev.199667 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 109.Keller P. J., “In vivo imaging of zebrafish embryogenesis,” Methods Zebrafish Methods 62, 268–278 (2013). 10.1016/j.ymeth.2013.03.015 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 110.Ding Y., Lee J., Ma J., Sung K., Yokota T., Singh N., Dooraghi M., Abiri P., Wang Y., Kulkarni R. P., Nakano A., Nguyen T. P., Fei P., and Hsiai T. K., “Light-sheet fluorescence imaging to localize cardiac lineage and protein distribution,” Sci. Rep. 7, 42209 (2017). 10.1038/srep42209 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 111.Shang L., Nienhaus K., Jiang X., Yang L., Landfester K., Mailänder V., Simmet T., and Nienhaus G. U., “Nanoparticle interactions with live cells: Quantitative fluorescence microscopy of nanoparticle size effects,” Beilstein J. Nanotechnol. 5, 2388–2397 (2014). 10.3762/bjnano.5.248 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 112.Ma Y., Benda A., Nicovich P. R., and Gaus K., “Measuring membrane association and protein diffusion within membranes with supercritical angle fluorescence microscopy,” Biomed. Opt. Express 7, 1561–1576 (2016). 10.1364/BOE.7.001561 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 113.Ritter J. G., Veith R., Veenendaal A., Siebrasse J. P., and Kubitscheck U., “Light sheet microscopy for single molecule tracking in living tissue,” PLoS One 5, e11639 (2010). 10.1371/journal.pone.0011639 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 114.Tomer R., Khairy K., Amat F., and Keller P. J., “Quantitative high-speed imaging of entire developing embryos with simultaneous multiview light-sheet microscopy,” Nat. Methods 9, 755–763 (2012). 10.1038/nmeth.2062 [DOI] [PubMed] [Google Scholar]
  • 115.Krieger J. W., Singh A. P., Garbe C. S., Wohland T., and Langowski J., “Dual-color fluorescence cross-correlation spectroscopy on a single plane illumination microscope (SPIM-FCCS),” Opt. Express 22, 2358–2375 (2014). 10.1364/OE.22.002358 [DOI] [PubMed] [Google Scholar]
  • 116.Migliori B., Datta M. S., Dupre C., Apak M. C., Asano S., Gao R., Boyden E. S., Hermanson O., Yuste R., and Tomer R., “Light sheet theta microscopy for rapid high-resolution imaging of large biological samples,” BMC Biol. 16, 57 (2018). 10.1186/s12915-018-0521-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 117.Zhao Z. W., Roy R., Gebhardt J. C. M., Suter D. M., Chapman A. R., and Xie X. S., “Spatial organization of RNA polymerase II inside a mammalian cell nucleus revealed by reflected light-sheet superresolution microscopy,” Proc. Natl. Acad. Sci. U.S.A. 111, 681–686 (2014). 10.1073/pnas.1318496111 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 118.Meddens M. B. M., Liu S., Finnegan P. S., Edwards T. L., James C. D., and Lidke K. A., “Single objective light-sheet microscopy for high-speed whole-cell 3D super-resolution,” Biomed. Opt. Express 7, 2219–2236 (2016). 10.1364/BOE.7.002219 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 119.Saliba N., Gagliano G., and Gustavsson A.-K., “Whole-cell multi-target single-molecule super-resolution imaging in 3D with microfluidics and a single-objective tilted light sheet,” bioRxiv (2023). 10.1101/2023.09.27.559876 [DOI] [Google Scholar]
  • 120.Krzic U., Gunther S., Saunders T. E., Streichan S. J., and Hufnagel L., “Multiview light-sheet microscope for rapid in toto imaging,” Nat. Methods 9, 730–733 (2012). 10.1038/nmeth.2064 [DOI] [PubMed] [Google Scholar]
  • 121.Santi P. A., Johnson S. B., Hillenbrand M., Grandpre P. Z., Glass T. J., and Leger J. R., “Thin-sheet laser imaging microscopy for optical sectioning of thick tissues,” BioTechniques 46, 287–294 (2009). 10.2144/000113087 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 122.Girkin J. M. and Carvalho M. T., “The light-sheet microscopy revolution,” J. Opt. 20, 053002 (2018). 10.1088/2040-8986/aab58a [DOI] [Google Scholar]
  • 123.Chatterjee K., Pratiwi F. W., Wu F. C. M., Chen P., and Chen B.-C., “Recent progress in light sheet microscopy for biological applications,” Appl. Spectrosc. 72, 1137–1169 (2018). 10.1177/0003702818778851 [DOI] [PubMed] [Google Scholar]
  • 124.Kafian H., Mozaffari-Jovin S., Bagheri M., and Shaegh S. A. M., “Light-sheet fluorescent microscopy: Fundamentals, developments and applications,” Phys. Scr. 98, 082001 (2023). 10.1088/1402-4896/acd7ae [DOI] [Google Scholar]
  • 125.Siedentopf H. and Zsigmondy R., “Uber sichtbarmachung und größenbestimmung ultramikoskopischer teilchen, mit besonderer anwendung auf goldrubingläser,” Ann. Phys. 315, 1–39 (1902). 10.1002/andp.19023150102 [DOI] [Google Scholar]
  • 126.Greger K., Swoger J., and Stelzer E. H. K., “Basic building units and properties of a fluorescence single plane illumination microscope,” RSI 78, 023705 (2007). 10.1063/1.2428277 [DOI] [PubMed] [Google Scholar]
  • 127.Wohland T., Shi X., Sankaran J., and Stelzer E. H. K., “Single plane illumination fluorescence correlation spectroscopy (SPIM-FCS) probes inhomogeneous three-dimensional environments,” Opt. Express 18, 10627–10641 (2010). 10.1364/OE.18.010627 [DOI] [PubMed] [Google Scholar]
  • 128.Singh A. P., Krieger J. W., Buchholz J., Charbon E., Langowski J., and Wohland T., “The performance of 2D array detectors for light sheet based fluorescence correlation spectroscopy,” Opt. Express 21, 8652–8668 (2013). 10.1364/OE.21.008652 [DOI] [PubMed] [Google Scholar]
  • 129.Oyler-Yaniv J., Oyler-Yaniv A., Maltz E., and Wollman R., “TNF controls a speed-accuracy tradeoff in the cell death decision to restrict viral spread,” Nat. Commun. 12, 2992 (2021). 10.1038/s41467-021-23195-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 130.Gualda E. J., “3D volume rendering of invertebrates using light-sheet fluorescence microscopy,” Photonics 9, 208 (2022). 10.3390/photonics9040208 [DOI] [Google Scholar]
  • 131.Wu Y., Ghitani A., Christensen R., Santella A., Du Z., Rondeau G., Bao Z., Colón-Ramos D., and Shroff H., “Inverted selective plane illumination microscopy (iSPIM) enables coupled cell identity lineaging and neurodevelopmental imaging in Caenorhabditis elegans,” Proc. Natl. Acad. Sci. U.S.A. 108, 17708–17713 (2011). 10.1073/pnas.1108494108 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 132.Wang Z., Qiao W., Jiang T., Chen S., Lu B., Ning K., Jin R., Gong H., and Yuan J., “Axial resolution and imaging contrast enhancement in inverted light-sheet microscopy by natural illumination modulation,” Front. Neurosci. 16, 1032195 (2022). 10.3389/fnins.2022.1032195 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 133.Bissardon C., Mermet X., Quintard C., Sanjuan F., Fouillet Y., Bottausci F., Carriere M., Rivera F., and Blandin P., “Selective plane illumination microscope dedicated to volumetric imaging in microfluidic chambers,” Biomed. Opt. Express 13, 5261–5274 (2022). 10.1364/BOE.455377 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 134.McGorty R., Liu H., Kamiyama D., Dong Z., Guo S., and Huang B., “Open-top selective plane illumination microscope for conventionally mounted specimens,” Opt. Express 23, 16142–16153 (2015). 10.1364/OE.23.016142 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 135.Zhu T., Nie J., Yu T., Zhu D., Huang Y., Chen Z., Gu Z., Tang J., Li D., and Fei P., “Large-scale high-throughput 3D culture, imaging, and analysis of cell spheroids using microchip-enhanced light-sheet microscopy,” Biomed. Opt. Express 14, 1659–1669 (2023). 10.1364/BOE.485217 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 136.Mcgorty R., Xie D., and Huang B., “High-NA open-top selective-plane illumination microscopy for biological imaging,” Opt. Express 25, 17798–17810 (2017). 10.1364/OE.25.017798 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 137.Kim B., Na M., Park S., Kim K., Park J.-H., Chung E., Chang S., and Kim K. H., “Open-top axially swept light-sheet microscopy,” Biomed. Opt. Express 12, 2328–2338 (2021). 10.1364/BOE.419030 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 138.Glaser A. K., Reder N. P., Chen Y., McCarty E. F., Yin C., Wei L., Wang Y., True L. D., and Liu J. T. C., “Light-sheet microscopy for slide-free non-destructive pathology of large clinical specimens,” Nat. Biomed. Eng. 1, 1–10 (2017). 10.1038/s41551-017-0084 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 139.Chen Y., Xie W., Glaser A. K., Reder N. P., Mao C., Dintzis S. M., Vaughan J. C., and Liu J. T. C., “Rapid pathology of lumpectomy margins with open-top light-sheet (OTLS) microscopy,” Biomed. Opt. Express 10, 1257–1272 (2019). 10.1364/BOE.10.001257 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 140.Barner L. A., Glaser A. K., True L. D., Reder N. P., and Liu J. T. C., “Solid immersion meniscus lens (SIMlens) for open-top light-sheet microscopy,” Opt. Lett. 44, 4451–4454 (2019). 10.1364/OL.44.004451 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 141.Barner L. A., Glaser A. K., Huang H., True L. D., and Liu J. T. C., “Multi-resolution open-top light-sheet microscopy to enable efficient 3D pathology workflows,” Biomed. Opt. Express 11, 6605–6619 (2020). 10.1364/BOE.408684 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 142.Mayer J., Robert-Moreno A., Sharpe J., and Swoger J., “Attenuation artifacts in light sheet fluorescence microscopy corrected by OPTiSPIM,” Light. Sci. Appl. 7, 70 (2018). 10.1038/s41377-018-0068-z [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 143.Ricci P., Sancataldo G., Gavryusev V., Franceschini A., Müllenbroich M. C., Silvestri L., and Pavone F. S., “Fast multi-directional DSLM for confocal detection without striping artifacts,” Biomed. Opt. Express 11, 3111–3124 (2020). 10.1364/BOE.390916 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 144.Wu Y., Wawrzusin P., Senseney J., Fischer R. S., Christensen R., Santella A., York A. G., Winter P. W., Waterman C. M., Bao Z., Colón-Ramos D. A., McAuliffe M., and Shroff H., “Spatially isotropic four-dimensional imaging with dual-view plane illumination microscopy,” Nat. Biotechnol. 31, 1032–1038 (2013). 10.1038/nbt.2713 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 145.Kumar A., Wu Y., Christensen R., Chandris P., Gandler W., McCreedy E., Bokinsky A., Colón-Ramos D. A., Bao Z., McAuliffe M., Rondeau G., and Shroff H., “Dual-view plane illumination microscopy for rapid and spatially isotropic imaging,” Nat. Protoc. 9, 2555–2573 (2014). 10.1038/nprot.2014.172 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 146.Schueth A., Hildebrand S., Samarska I., Sengupta S., Kiessling A., Herrler A., zur Hausen A., Capalbo M., and Roebroeck A., “Efficient 3D light-sheet imaging of very large-scale optically cleared human brain and prostate tissue samples,” Commun. Biol. 6, 1–15 (2023). 10.1038/s42003-023-04536-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 147.Vladimirov N., Preusser F., Wisniewski J., Yaniv Z., Desai R. A., Woehler A., and Preibisch S., “Dual-view light-sheet imaging through a tilted glass interface using a deformable mirror,” Biomed. Opt. Express 12, 2186–2203 (2021). 10.1364/BOE.416737 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 148.Kumar A., Christensen R., Guo M., Chandris P., Duncan W., Wu Y., Santella A., Moyle M., Winter P. W., Colón-Ramos D., Bao Z., and Shroff H., “Using stage- and slit-scanning to improve contrast and optical sectioning in dual-view inverted light sheet microscopy (diSPIM),” Biol. Bull. 231, 26–39 (2016). 10.1086/689589 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 149.Greiss F., Deligiannaki M., Jung C., Gaul U., and Braun D., “Single-molecule imaging in living Drosophila embryos with reflected light-sheet microscopy,” Biophys. J. 110, 939–946 (2016). 10.1016/j.bpj.2015.12.035 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 150.Wang Y., Dong D., Yang W., He R., Lei M., and Shi K., “Reflective ultrathin light-sheet microscopy with isotropic 3D resolutions,” Photon. Res. 12, 271–281 (2024). 10.1364/PRJ.500618 [DOI] [Google Scholar]
  • 151.Gebhardt J. C. M., Suter D. M., Roy R., Zhao Z. W., Chapman A. R., Basu S., Maniatis T., and Xie X. S., “Single-molecule imaging of transcription factor binding to DNA in live mammalian cells,” Nat. Methods 10, 421–426 (2013). 10.1038/nmeth.2411 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 152.Jannasch A., Szilagyi S. A., Burmeister M., Davis Q. T., Hermsdorf G. L., De S., and Schäffer E., “Fast 3D imaging of giant unilamellar vesicles using reflected light-sheet microscopy with single molecule sensitivity,” J. Microsc. 285, 40–51 (2022). 10.1111/jmi.13070 [DOI] [PubMed] [Google Scholar]
  • 153.Kuhn T., Landge A. N., Mörsdorf D., Coßmann J., Gerstenecker J., Čapek D., Müller P., and Gebhardt J. C. M., “Single-molecule tracking of nodal and lefty in live zebrafish embryos supports hindered diffusion model,” Nat. Commun. 13, 6101 (2022). 10.1038/s41467-022-33704-z [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 154.Nelsen E., Hobson C. M., Kern M. E., Hsiao J. P., O’Brien III E. T., Watanabe T., Condon B. M., Boyce M., Grinstein S., Hahn K. M., Falvo M. R., and Superfine R., “Combined atomic force microscope and volumetric light sheet system for correlative force and fluorescence mechanobiology studies,” Sci. Rep. 10, 8133 (2020). 10.1038/s41598-020-65205-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 155.Beicker K., O’Brien E. T., Falvo M. R., and Superfine R., “Vertical light sheet enhanced side-view imaging for AFM cell mechanics studies,” Sci. Rep. 8, 1504 (2018). 10.1038/s41598-018-19791-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 156.Zagato E., Brans T., Verstuyft S., Thourhout D. V., Missinne J., Steenberge G. V., Demeester J., Smedt S. D., Remaut K., Neyts K., and Braeckmans K., “Microfabricated devices for single objective single plane illumination microscopy (SoSPIM),” Opt. Express 25, 1732–1745 (2017). 10.1364/OE.25.001732 [DOI] [PubMed] [Google Scholar]
  • 157.Coelho S., Baek J., Walsh J., Justin Gooding J., and Gaus K., “Direct-laser writing for subnanometer focusing and single-molecule imaging,” Nat. Commun. 13, 647 (2022). 10.1038/s41467-022-28219-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 158.Beghin A., Grenci G., Sahni G., Guo S., Rajendiran H., Delaire T., Mohamad Raffi S. B., Blanc D., de Mets R., Ong H. T., Galindo X., Monet A., Acharya V., Racine V., Levet F., Galland R., Sibarita J.-B., and Viasnoff V., “Automated high-speed 3D imaging of organoid cultures with multi-scale phenotypic quantification,” Nat. Methods 19, 881–892 (2022). 10.1038/s41592-022-01508-0 [DOI] [PubMed] [Google Scholar]
  • 159.Konopka C. A. and Bednarek S. Y., “Variable-angle epifluorescence microscopy: A new way to look at protein dynamics in the plant cell cortex,” Plant J. 53, 186–196 (2008). 10.1111/j.1365-313X.2007.03306.x [DOI] [PubMed] [Google Scholar]
  • 160.Tokunaga M., Imamoto N., and Sakata-Sogawa K., “Highly inclined thin illumination enables clear single-molecule imaging in cells,” Nat. Methods 5, 159–161 (2008). 10.1038/nmeth1171 [DOI] [PubMed] [Google Scholar]
  • 161.Knight A. E., “Single-molecule fluorescence imaging by total internal reflection fluorescence microscopy (IUPAC technical report),” Pure Appl. Chem. 86, 1303–1320 (2014). 10.1515/pac-2012-0605 [DOI] [Google Scholar]
  • 162.Bergmann S., Wrede O., Huser T., and Hellweg T., “Super-resolution optical microscopy resolves network morphology of smart colloidal microgels,” Phys. Chem. Chem. Phys. 20, 5074–5083 (2018). 10.1039/C7CP07648G [DOI] [PubMed] [Google Scholar]
  • 163.Kozgunova E. and Goshima G., “A versatile microfluidic device for highly inclined thin illumination microscopy in the moss Physcomitrella patens,” Sci. Rep. 9, 15182 (2019). 10.1038/s41598-019-51624-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 164.Scalisi S., Pisignano D., and Cella Zanacchi F., “Single-molecule localization microscopy goes quantitative,” Microsc. Res. Tech. 86, 494–504 (2023). 10.1002/jemt.24281 [DOI] [PubMed] [Google Scholar]
  • 165.Tang J. and Han K. Y., “Extended field-of-view single-molecule imaging by highly inclined swept illumination,” Optica 5, 1063–1069 (2018). 10.1364/OPTICA.5.001063 [DOI] [Google Scholar]
  • 166.Huisken J. and Stainier D. Y. R., “Selective plane illumination microscopy techniques in developmental biology,” Development 136, 1963–1975 (2009). 10.1242/dev.022426 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 167.Fujiwara E., Cabral T. D., Sato M., Oku H., and Cordeiro C. M. B., “Agarose-based structured optical fibre,” Sci. Rep. 10, 7035 (2020). 10.1038/s41598-020-64103-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 168.Pampaloni F., Reynaud E. G., and Stelzer E. H. K., “The third dimension bridges the gap between cell culture and live tissue,” Nat. Rev. Mol. Cell Biol. 8, 839–845 (2007). 10.1038/nrm2236 [DOI] [PubMed] [Google Scholar]
  • 169.Keller P. J., Schmidt A. D., Wittbrodt J., and Stelzer E. H. K., “Digital scanned laser light-sheet fluorescence microscopy (DSLM) of zebrafish and drosophila embryonic development,” Cold Spring Harb. Protoc. 2011, pdb.prot065839 (2011). 10.1101/pdb.prot065839 [DOI] [PubMed] [Google Scholar]
  • 170.Kaufmann A., Mickoleit M., Weber M., and Huisken J., “Multilayer mounting enables long-term imaging of zebrafish development in a light sheet microscope,” Development 139, 3242–3247 (2012). 10.1242/dev.082586 [DOI] [PubMed] [Google Scholar]
  • 171.Yang G., Wang L., Qin X., Chen X., Liang Y., Jin X., Chen C., Zhang W., Pan W., and Li H., “Heterogeneities of zebrafish vasculature development studied by a high throughput light-sheet flow imaging system,” Biomed. Opt. Express 13, 5344–5357 (2022). 10.1364/BOE.470058 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 172.He B., Sridhar A., Streiff C., Deketelaere C., Zhang H., Gao Y., Hu Y., Pirotte S., Delrez N., Davison A. J., Donohoe O., and Vanderplasschen A. F. C., “In vivo imaging sheds light on the susceptibility and permissivity of Carassius auratus to cyprinid herpesvirus 2 according to developmental stage,” Viruses 15, 1746 (2023). 10.3390/v15081746 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 173.Pampaloni F., Berge U., Marmaras A., Horvath P., Kroschewski R., and Stelzer E. H. K., “Tissue-culture light sheet fluorescence microscopy (TC-LSFM) allows long-term imaging of three-dimensional cell cultures under controlled conditions,” Integr. Biol. 6, 988–998 (2014). 10.1039/C4IB00121D [DOI] [PubMed] [Google Scholar]
  • 174.Jahr W., Schmid B., Weber M., and Huisken J., “eduSPIM: Light sheet microscopy in the museum,” PLoS One 11, e0161402 (2016). 10.1371/journal.pone.0161402 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 175.Lichtenberg M., Trampe E. C., Vogelmann T. C., and Kühl M., “Light sheet microscopy imaging of light absorption and photosynthesis distribution in plant tissue,” Plant Physiol. 175, 721–733 (2017). 10.1104/pp.17.00820 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 176.Kafian H., Lalenejad M., Moradi-Mehr S., Birgani S. A., and Abdollahpour D., “Light-sheet fluorescence microscopy with scanning non-diffracting beams,” Sci. Rep. 10, 8501 (2020). 10.1038/s41598-020-63847-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 177.Gustavsson A.-K., Petrov P. N., Lee M. Y., Shechtman Y., and Moerner W. E., “3D single-molecule super-resolution microscopy with a tilted light sheet,” Nat. Commun. 9, 123 (2018). 10.1038/s41467-017-02563-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 178.Nussbaumer R. J., Halter M., Tervoort T., Caseri W. R., and Smith P., “A simple method for the determination of refractive indices of (rough) transparent solids,” J. Mater. Sci. 40, 575–582 (2005). 10.1007/s10853-005-6291-z [DOI] [Google Scholar]
  • 179.Murray E., Cho J. H., Goodwin D., Ku T., Swaney J., Kim S.-Y., Choi H., Park Y.-G., Park J.-Y., Hubbert A., McCue M., Vassallo S., Bakh N., Frosch M. P., Wedeen V. J., Seung H. S., and Chung K., “Simple, scalable proteomic imaging for high-dimensional profiling of intact systems,” Cell 163, 1500–1514 (2015). 10.1016/j.cell.2015.11.025 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 180.Williams R. M., Bloom J. C., Robertus C. M., Recknagel A. K., Putnam D., Schimenti J. C., and Zipfel W. R., “Practical strategies for robust and inexpensive imaging of aqueous-cleared tissues,” J. Microsc. 291, 237–247 (2023). 10.1111/jmi.13213 [DOI] [PubMed] [Google Scholar]
  • 181.Song D.-H., Woo Song C., Chung J., Jang E.-H., Kim H., Hur Y., Hur E.-M., Kim D., and Chang J.-B., “In situ silver nanoparticle development for molecular-specific biological imaging via highly accessible microscopies,” Nanoscale Adv. 5, 1636–1650 (2023). 10.1039/D2NA00449F [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 182.Verma N. C., Khan S., and Nandi C. K., “Single-molecule analysis of fluorescent carbon dots towards localization-based super-resolution microscopy,” Methods Appl. Fluoresc. 4, 044006 (2016). 10.1088/2050-6120/4/4/044006 [DOI] [PubMed] [Google Scholar]
  • 183.Blanquer G., van Dam B., Gulinatti A., Acconcia G., De Wilde Y., Izeddin I., and Krachmalnicoff V., “Relocating single molecules in super-resolved fluorescence lifetime images near a plasmonic nanostructure,” ACS Photon. 7, 393–400 (2020). 10.1021/acsphotonics.9b01317 [DOI] [Google Scholar]
  • 184.Jaimes R. F., Bramich D. M., Skottfelt J., Kains N., Jørgensen U. G., Horne K., Dominik M., Alsubai K. A., Bozza V., Novati S. C., Ciceri S., D’Ago G., Galianni P., Gu S.-H., Harpsøe K. B. W., Haugbølle T., Hinse T. C., Hundertmark M., Juncher D., Korhonen H., Mancini L., Popovas A., Rabus M., Rahvar S., Scarpetta G., Schmidt R. W., Snodgrass C., Southworth J., Starkey D., Street R. A., Surdej J., Wang X.-B., and Wertz O., “Exploring the crowded central region of ten galactic globular clusters using EMCCDs—Variable star searches and new discoveries,” A&A 588, A128 (2016). 10.1051/0004-6361/201527641 [DOI] [Google Scholar]
  • 185.Gili R., Prieur J.-L., Rivet J.-P., Scardia M., Pansecchi L., Argyle R. W., Ling J. F., Piccotti L., Aristidi E., Koechlin L., Bonneau D., and Maccarini L., “Measurements of visual binaries with EMCCD cameras and the nice 76-cm refractor in 2009–2010,” Astron. Nachr. 341, 441–452 (2020). 10.1002/asna.202013792 [DOI] [Google Scholar]
  • 186.Andor , see https://andor.oxinst.com/learning/view/article/comparing-scmos for information about “A comparison of EMCCD vs sCMOS cameras.”
  • 187.Nyquist H., “Certain factors affecting telegraph speed,” Trans. Am. Inst. Electr. Eng. 3, 324–346 (1924). 10.1002/j.1538-7305.1924.tb01361.x [DOI] [Google Scholar]
  • 188.Luo W., Zhang Y., Feizi A., Göröcs Z., and Ozcan A., “Pixel super-resolution using wavelength scanning,” Light Sci. Appl. 5, e16060 (2016). 10.1038/lsa.2016.60 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 189.Heidingsfelder P., Gao J., Wang K., and Ott P., “Subpixel edge localization with reduced uncertainty by violating the Nyquist criterion,” Opt. Eng. 53, 122410 (2014). 10.1117/1.OE.53.12.122410 [DOI] [Google Scholar]
  • 190.Fazel M. and Wester M. J., “Analysis of super-resolution single molecule localization microscopy data: A tutorial,” AIP Adv. 12, 010701 (2022). 10.1063/5.0069349 [DOI] [Google Scholar]
  • 191.Smal I., Loog M., Niessen W., and Meijering E., “Quantitative comparison of spot detection methods in fluorescence microscopy,” IEEE Trans. Med. Imaging 29, 282–301 (2010). 10.1109/TMI.2009.2025127 [DOI] [PubMed] [Google Scholar]
  • 192.Qiao L., Wang M., and Jin Z., “A method for EMCCD multiplication gain measurement with comprehensive correction,” Sci. Rep. 11, 6058 (2021). 10.1038/s41598-021-85511-z [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 193.Andor , see https://andor.oxinst.com/products/ixon-emccd-cameras-for-life-science for information about “iXon EMCCD cameras for microscopy & life science.”
  • 194.Mandracchia B., Hua X., Guo C., Son J., Urner T., and Jia S., “Fast and accurate sCMOS noise correction for fluorescence microscopy,” Nat. Commun. 11, 94 (2020). 10.1038/s41467-019-13841-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 195.Sadoine M., Ishikawa Y., Kleist T. J., Wudick M. M., Nakamura M., Grossmann G., Frommer W. B., and Ho C.-H., “Designs, applications, and limitations of genetically encoded fluorescent sensors to explore plant biology,” Plant Physiol. 187, 485–503 (2021). 10.1093/plphys/kiab353 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 196.Basumatary J., Baro N., Zanacchi F. C., and Mondal P. P., “Temporally resolved SMLM (with large PAR shift) enabled visualization of dynamic HA cluster formation and migration in a live cell,” Sci. Rep. 13, 12561 (2023). 10.1038/s41598-023-39096-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 197.Fujiwara T. K., Takeuchi S., Kalay Z., Nagai Y., Tsunoyama T. A., Kalkbrenner T., Iwasawa K., Ritchie K. P., Suzuki K. G., and Kusumi A., “Development of ultrafast camera-based single fluorescent-molecule imaging for cell biology,” J. Cell Biol. 222, e202110160 (2023). 10.1083/jcb.202110160 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 198.Liu S., Mlodzianoski M. J., Hu Z., Ren Y., McElmurry K., Suter D. M., and Huang F., “sCMOS noise-correction algorithm for microscopy images,” Nat. Methods 14, 760–761 (2017). 10.1038/nmeth.4379 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 199.Fowler B., Liu C., Mims S., Balicki J., Li W., Do H., Appelbaum J., and Vu P., “A 5.5mpixel 100 frames/sec wide dynamic range low noise CMOS image sensor for scientific applications,” in Sensors, Cameras, and Systems for Industrial/Scientific Applications XI (SPIE, 2010), Vol. 7536, pp. 58–69.
  • 200.Muschol M., Kosterin P., Ichikawa M., and Salzberg B. M., “Activity-dependent depression of excitability and calcium transients in the neurohypophysis suggests a model of “stuttering conduction”,” J. Neurosci. 23, 11352–11362 (2003). 10.1523/JNEUROSCI.23-36-11352.2003 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 201.Kisley L., Brunetti R., Tauzin L. J., Shuang B., Yi X., Kirkeminde A. W., Higgins D. A., Weiss S., and Landes C. F., “Characterization of porous materials by fluorescence correlation spectroscopy super-resolution optical fluctuation imaging,” ACS Nano 9, 9158–9166 (2015). 10.1021/acsnano.5b03430 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 202.Yoshida S., Schmid W., Vo N., Calabrase W., and Kisley L., “Computationally-efficient spatiotemporal correlation analysis super-resolves anomalous diffusion,” Opt. Express 29, 7616–7629 (2021). 10.1364/OE.416465 [DOI] [PubMed] [Google Scholar]
  • 203.Chatterjee S., Kramer S. N., Wellnitz B., Kim A., and Kisley L., “Spatially resolving size effects on diffusivity in nanoporous extracellular matrix-like materials with fluorescence correlation spectroscopy super-resolution optical fluctuation imaging,” J. Phys. Chem. B 127, 4430–4440 (2023). 10.1021/acs.jpcb.3c00941 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 204.Antarasen J., Wellnitz B., Kramer S. N., Chatterjee S., and Kisley L., “Cross-correlation increases sampling in diffusion-based super-resolution optical fluctuation imaging,” bioRxiv (2024). 10.1021/cbmi.4c00032 [DOI] [Google Scholar]
  • 205.Ricci P., Gavryusev V., Müllenbroich C., Turrini L., de Vito G., Silvestri L., Sancataldo G., and Saverio Pavone F., “Removing striping artifacts in light-sheet fluorescence microscopy: A review,” Prog. Biophys. Mol. Biol. 168, 52–65 (2022). 10.1016/j.pbiomolbio.2021.07.003 [DOI] [PubMed] [Google Scholar]
  • 206.R. H. Hill, Jr and Finster D. C., “Minimizing, controlling and managing hazards,” in Laboratory Safety for Chemistry Students, 1st ed. (Wiley, 2010), Chap. 7.
  • 207.Goncalves D. P. and Griffith D. J., “Estimating uncertainty in resolution tests,” Opt. Eng. 45, 053601 (2006). 10.1117/1.2202914 [DOI] [Google Scholar]
  • 208.Picazo-Bueno J. A., Trindade K., Sanz M., and Micó V., “Design, calibration, and application of a robust, cost-effective, and high-resolution lensless holographic microscope,” Sensors 22, 553 (2022). 10.3390/s22020553 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 209.K. R. G. CWRU, “SuperResKinetics/LRG_superres_particle_identify,” GitHub (2021).
  • 210.Zessin P. J., Krüger C. L., Malkusch S., Endesfelder U., and Heilemann M., “A hydrophilic gel matrix for single-molecule super-resolution microscopy,” Opt. Nano 2, 4 (2013). 10.1186/2192-2853-2-4 [DOI] [Google Scholar]
  • 211.Ma H., Xu J., Jin J., Huang Y., and Liu Y., “A simple marker-assisted 3D nanometer drift correction method for superresolution microscopy,” Biophys. J. 112, 2196–2208 (2017). 10.1016/j.bpj.2017.04.025 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 212.Damstra H. G. J., Passmore J. B., Serweta A. K., Koutlas I., Burute M., Meye F. J., Akhmanova A., and Kapitein L. C., “Gelmap: Intrinsic calibration and deformation mapping for expansion microscopy,” Nat. Methods 20, 1573–1580 (2023). 10.1038/s41592-023-02001-y [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 213.Cohen E. A. K. and Ober R. J., “Analysis of point based image registration errors with applications in single molecule microscopy,” IEEE Trans. Signal Process. 61, 6291–6306 (2013). 10.1109/TSP.2013.2284154 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 214.Hedde P. N. and Gratton E., “Active focus stabilization for upright selective plane illumination microscopy,” Opt. Express 23, 14707–14714 (2015). 10.1364/OE.23.014707 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 215.Coelho S., Baek J., Walsh J., Gooding J. J., and Gaus K., “3D active stabilization for single-molecule imaging,” Nat. Protoc. 16, 497–515 (2021). 10.1038/s41596-020-00426-9 [DOI] [PubMed] [Google Scholar]
  • 216.Kisley Research Group @ CWRU (2024). “fcsSOFI/LSM_3d_fcssofi_20240425.m,” GitHub. 10.5281/zenodo.12726411 [DOI]
  • 217.Edelstein A., Amodaj N., Hoover K., Vale R., and Stuurman N., “Computer control of microscopes using Manager,” Curr. Protoc. Mol. Biol. 92, 14.20.1–14.20.17 (2010). 10.1002/0471142727.mb1420s92 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 218.Edelstein A. D., Tsuchida M. A., Amodaj N., Pinkard H., Vale R. D., and Stuurman N., “Advanced methods of microscope control using μManager misc,” J. Biol. Methods 1, e10 (2014). 10.14440/jbm.2014.36 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 219.Toloui M. and Hong J., “High fidelity digital inline holographic method for 3D flow measurements,” Opt. Express 23, 27159–27173 (2015). 10.1364/OE.23.027159 [DOI] [PubMed] [Google Scholar]
  • 220.Sanders T., Hedges R., Schulz T., Abijaoude M., Peters J., Steinbock M., Arreola A., and Holmes T., “Real time deconvolution of adaptive optics ground based telescope imagery,” J. Astronaut. Sci. 69, 175–191 (2022). 10.1007/s40295-021-00285-w [DOI] [Google Scholar]
  • 221.Richardson W. H., “Bayesian-based iterative method of image restoration,” J. Opt. Soc. Am. 62, 55–59 (1972). 10.1364/JOSA.62.000055 [DOI] [Google Scholar]
  • 222.Zeng G. and Gullberg G., “Unmatched projector/backprojector pairs in an iterative reconstruction algorithm,” IEEE Trans. Med. Imag. 19, 548–555 (2000). 10.1109/42.870265 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 223.Ichioka Y. and Nakajima N., “Iterative image restoration considering visibility,” J. Opt. Soc. Am. 71, 983–988 (1981). 10.1364/JOSA.71.000983 [DOI] [Google Scholar]
  • 224.Guo M., Li Y., Su Y., Lambert T., Nogare D. D., Moyle M. W., Duncan L. H., Ikegami R., Santella A., Rey-Suarez I., Green D., Chen J., Vishwasrao H., Ganesan S., Waters J. C., Annunziata C. M., Hafner M., Mohler W. A., Chitnis A. B., Upadhyaya A., Usdin T. B., Bao Z., Colón-Ramos D., Riviere P. L., Liu H., Wu Y., and Shroff H., “Accelerating iterative deconvolution and multiview fusion by orders of magnitude,” bioRxiv (2019).
  • 225.Kawata S. and Ichioka Y., “Iterative image restoration for linearly degraded images. I. Basis,” J. Opt. Soc. Am. 70, 762–768 (1980). 10.1364/JOSA.70.000762 [DOI] [Google Scholar]
  • 226.Latychevskaia T., Gehri F., and Fink H.-W., “Depth-resolved holographic reconstructions by three-dimensional deconvolution,” Opt. Express 18, 22527–22544 (2010). 10.1364/OE.18.022527 [DOI] [PubMed] [Google Scholar]
  • 227.Shuang B., Chen J., Kisley L., and Landes C. F., “Troika of single particle tracking programing: SNR enhancement, particle identification, and mapping,” Phys. Chem. Chem. Phys. 16, 624–634 (2013). 10.1039/C3CP53968G [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 228.Woodford O., “vol3d v2,” MATLAB Central File Exchange (2018).
  • 229.Pluen A., Netti P. A., Jain R. K., and Berk D. A., “Diffusion of macromolecules in agarose gels: Comparison of linear and globular configurations,” Biophys. J. 77, 542–552 (1999). 10.1016/S0006-3495(99)76911-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 230.Khodadadi Yazdi M., Taghizadeh A., Taghizadeh M., Stadler F. J., Farokhi M., Mottaghitalab F., Zarrintaj P., Ramsey J. D., Seidi F., Saeb M. R., and Mozafari M., “Agarose-based biomaterials for advanced drug delivery,” J. Control. Release 326, 523–543 (2020). 10.1016/j.jconrel.2020.07.028 [DOI] [PubMed] [Google Scholar]
  • 231.Kalwarczyk T., Kwapiszewska K., Szczepanski K., Sozanski K., Szymanski J., Michalska B., Patalas-Krawczyk P., Duszynski J., and Holyst R., “Apparent anomalous diffusion in the cytoplasm of human cells: The effect of probes’ polydispersity,” J. Phys. Chem. B 121, 9831–9837 (2017). 10.1021/acs.jpcb.7b07158 [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The data that support the findings of this study are available from the corresponding author upon reasonable request.


Articles from Journal of Applied Physics are provided here courtesy of American Institute of Physics

RESOURCES