Skip to main content
. 2024 Sep 10;11:986. doi: 10.1038/s41597-024-03751-w

Table 8.

Precision (1-σ) and accuracy of the “check” standard for the L2130-i in IMAS (N = 400) and the the L2140-i in ANU (N = 94).

Instrument δ18O precision [‰] δ18O accuracy [‰] δD precision [‰] δD accuracy [‰]
L2130-i 0.025 ± 0.014 0.020 ± 0.011 0.230 ± 0.125 0.181 ± 0.097
L2140-i 0.049 ± 0.019 0.040 ± 0.018 0.467 ± 0.205 0.372 ± 0.150