Table 3.

Paired t-test results of the paired differences in the measurements of FVC and FEV1 between the prototype and the plethysmograph, as well their respective 95% confidence interval (CI) and p-values.

Paired Differences 95% CI of the Difference Significance
Metric Mean SD Lower Bound Upper Bound p-Value
FVC (L) 0.0063 0.1410 −0.1116 0.1241 0.9040
FEV1 (L) −0.0912 0.3180 −0.3571 0.1746 0.4440