Table 5.

Pearson correlation and intraclass correlation coefficients for the measurements of FVC and FEV1 between the prototype and the plethysmograph, as well as their respective p-values. The 95% CIs for ICC are also presented. The two-way random effects model (2,1) was employed.

Pearson Correlation Intraclass Correlation
Metric r p-Value ICC (2,1) (95% CI) p-Value
FVC 0.992 <0.001 0.987 (0.936 to 0.997) <0.001
FEV1 0.968 <0.001 0.907 (0.609 to 0.981) <0.001