Table 1.

MRI acquisition parameters

MRI parameter Standard CAIPIRINHA-VIBE DL CAIPIRINHA-VIBE HR-DL CAIPIRINHA-VIBE
Orientation Axial Axial Axial
Repetition time, ms 3.3 3.1 3.1
Echo time, ms 1.21 1.21 1.16
Flip angle, degree 11 11 11
Receiver bandwidth, Hertz/pixel 590 590 600
Field of view, mm2 380 × 313 380 × 313 380 × 313
Matrix 352 × 203 352 × 203 320 × 224
Slice thickness, mm 3.0 3.0 1.0
Number of slice 64 64 192
Acquired voxel size, mm3 1.1 × 1.5 × 6.0 1.1 × 1.5 × 6.0 1.2 × 1.4 × 2.0
Reconstructed voxel size, mm3 0.5 × 0.5 × 3.0 0.5 × 0.5 × 3.0 0.6 × 0.6 × 1.0
Number of excitation 1 1 1
Acceleration factor 4 4 6
Acquisition time, s 17 11 16

CAIPIRINHA controlled aliasing in parallel imaging results in higher acceleration, DL deep learning, HR high resolution, MRI magnetic resonance imaging, VIBE volumetric interpolated breath-hold examination

For precontrast image, standard CAIPIRINHA-VIBE and DL CAIPIRINHA-VIBE were performed

For HBP image, standard CAIPIRINHA-VIBE, DL CAIPIRINHA-VIBE, and HR-DL CAIPIRINHA-VIBE were performed