Table 1. IC50 Values for Seven Reference Compounds Obtained by Four Different Techniques: RapidFire-MS, DESI-MS, Radioassay,3 and LC-MS3.

  IC50 (μM)
compound RapidFire-MS DESI-MS) radioassay LC-MS
glafenine N/Aa N/Aa 53.8 ± 14 28.1 ± 8.4
zafirlukast 7.30 ± 1.41 2.96 ± 1.99 8.57 ± 1.7 20.4 ± 10
glyburide 21.36 ± 6.94 3.78 ± 1.83 6.79 ± 2.3 4.26 ± 0.8
nefazodone 6.69 ± 1.36 4.16 ± 2.76 7.11 ± 4.8 3.61 ± 0.5
ritonavir 0.90 ± 0.35 0.61 ± 0.54 2.92 ± 0.7 2.49 ± 0.9
rosiglitazone 2.26 ± 0.22 7.59 ± 3.29 5.47 ± 0.5 2.60 ± 0.4
cyclosporin A 5.38 ± 1.46 4.49 ± 2.13 not tested not tested
a